1、T.O. 33K1-4-1968-1 15 OCTOBER 1991 TECHNICAL MANUAL CALIBRATION PROCEDURE FOR HIGH POTENTIAL TESTER 4045AI (ASSOCIATED RESEARCH) Distribution Statement - Distribution authorized to U.S. Government agencies and their contractors for official use or for administrative or operational purposes only, 15

2、October 1991. Other requests for this document shall be referred to AFMETCAL Detachment 1/MLLW, 813 Irving-Wick Dr W, Suite 4M, Heath, OH 43056-6116. Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent disclosure of the contents or reconstruction of the

3、document. Published under Authority of the Secretary of the Air Force RadioFans.CN 收音机爱 好者资料库 T.O. 33K1-4-1986-1 1 HIGH POTENTIAL TESTER 4045AI (ASSOCIATED RESEARCH) 1 CALIBRATION DESCRIPTION: Table 1. Test Instrument (TI)PerformanceTest CharacteristicsSpecificationsMethod DC VoltageRange: 0 to 5 kV

4、Measured with a DMM and Voltage Divider Accuracy: 2.0% FS AC VoltageRange: 0 to 5 kVMeasured with a DMM Accuracy: 2.0% FS 2 EQUIPMENT REQUIREMENTS: Minimum UseCalibrationSub- NounSpecificationsEquipmentItem 2.1 DC VOLTMETERRange: 0 to 10 VDCFluke 8506A Accuracy: 0.1% 2.2 DIGITAL MULTIMETERRange: 0 t

5、o 1100 VACFluke 8600A Accuracy: 0.1% 2.3 VOLTAGE DIVIDERRange: 0 to 5 kVDCFluke 80E-10 Accuracy: 0.1% 2.4 RESISTORRange: 3 M, 1/2 wattAs Avail. Accuracy: N/A 3 PRELIMINARY OPERATIONS: 3.1 Review and become familiar with entire procedure before beginning calibration process. RadioFans.CN 收音机爱 好者资料库 T

6、.O. 33K1-4-1968-1 2 WARNING Unless otherwise designated, and prior to beginning the Calibration Process, ensure that all test equipment voltage and/or current outputs are set to zero (0) or turned off, where applicable. Ensure that all equipment switches are set to the proper position before making

7、connections or applying power. 3.2 Ensure that TI is properly grounded, connect all case grounds to facility ground. WARNING This procedure involves voltages that are hazardous to personnel. All necessary precautions during the conduct of these tests must be observed. 3.3 Connect TI and test equipme

8、nt to 115 V/60 Hz power, set POWER switches to ON and allow a 15 minute warm-up period. 3.4 Zero TI meter if required. 3.5 Connect the 3 M Resistor to TI DC output. 3.6 Set TI output control for VDC and adjust for 900 VDC as monitored on the Digital Multimeter. 3.7 The TI alarm must sound and LEAKAG

9、E lamp must be illuminated. 3.8 Set TI output voltage to minimum. 3.9 Repeat steps 3.5 through 3.8 except set TI and Digital Multimeter to AC and use TI AC output terminals. NOTE If a maximum Leakage current value is required, calculate resistance needed and perform steps 3.5 through 3.8 using the c

10、alculated value of resistance in step 3.5. 3.10 Remove Resistor from TI and adjust TI voltmeter for an indication of 500 volts. 3.11 Carefully bring TI output lead tips close enough for arcing to occur. 3.12 The TI BREAKDOWN and LEAKAGE lamps must be illuminated. 3.13 Set TI output voltage to minimu

11、m. 4 CALIBRATION PROCESS: NOTE Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding. RadioFans.CN 收音机爱 好者资料库 T.O. 33K1-4-1986-1 3 4.1 DC VOLTAGE CALIBRATION: 4.1.1 Connect the Voltage Divider input to the T

12、I DC output and the DC Voltmeter (2.1) INPUT to the Voltage Divider 1000:1 OUTPUT terminals. 4.1.2 Set DC Voltmeter FUNCTION and RANGE to 10 VDC. 4.1.3 Set TI output for VDC. 4.1.4 Adjust TI output control until TI indicates the first value listed in the Applied column of Table 2. 4.1.5 The DC Voltm

13、eter must indicate within the corresponding limits listed in the Limits column of Table 2. 4.1.6 Repeat steps 4.1.4 through 4.1.5 for each remaining applied value in Table 2. Table 2. Applied (kVDC)Limits (kVDC) 10.90 to 1.10 21.90 to 2.10 32.90 to 3.10 43.90 to 4.10 54.90 to 5.10 4.1.7 Set TI outpu

14、t to minimum and disconnect test set up. 4.2 AC VOLTAGE CALIBRATION: 4.2.1 Set TI outp;ut for AC. 4.2.2 Connect TI AC output to Digital Multimeter voltage INPUT. 4.2.3 Set Digital Multimeter FUNCTION and RANGE to 1200 VAC. 4.2.4 Carefully adjust TI output control until TI meter indicates 1 kVAC. 4.2

15、.5 The Digital Multimeter must indicate within the lijits of 900 to 1100 VAC. 4.2.6 Set TI output to minimum. 4.2.7 Disconnect and secure all equipment. T.O. 33K1-4-1968-1 4 CALIBRATION PERFORMANCE TABLE DC VOLTAGE CALIBRATION: Applied (kVDC)Limits (kVDC) 10.90 to 1.10 21.90 to 2.10 32.90 to 3.10 43.90 to 4.10 54.90 to 5.10 AC VOLTAGE CALIBRATION: Applied Limits 1 kVAC900 to 1100 VAC