1、500 V. Analogue Insulation clean, and with no broken or cracked insulation. Replacement fuses must be of the correct type and rating. BM 100/4 Only U.K. Safety Authorities recommend the use of fused test leads when measuring voltage on high energy systems. When making a voltage measurement, the Test

2、 push must not be pressed. NOTE T H E I N S T R U M E N T S M U S T O N Y B E U S E D B Y S U I TA B LY T R A I N E D A N D C O M P E T E N T P E R S O N S. Symbols used on the instrument Caution: Refer to accompanying notes. Equipment protected throughout by Double Insulation (Class II). Equipment

3、complies with relevant EU Directives RadioFans.CN 收音机爱 好者资料库 OPERATION Preliminary Test lead check 1.Before each use of the instrument, inspect the test leads to confirm that their condition is good, with no damaged or broken insulation. 2. Insert the red and black test leads, terminated with the ap

4、propriate prods or clips, to their respective + and - terminal sockets. 3.Short the test lead prods or clips together. Set the selector switch to 2 . 4. Press the Test push. Check that the instrument reads zero. If a higher resistance reading is obtained check all connections. If the reading remains

5、 high, suspect that the test leads may be at fault, or that the 2A non replaceable fuse has ruptured. Continuity Testing 1. Set the selector switch to 2 , or 200, as appropriate. 2. Connect the test leads across the isolated item / circuit under test. 3. Press and hold the Test push button; allow th

6、e needle to settle. There is a separate scale for each range so each reading is direct. Take the reading from the appropriate scale of the instrument. 4. Release the Test push button and disconnect the test leads. Insulation Testing 1. Set the selector switch to 200 M. 2. Connect the R e d test lead

7、 to earth, equipment frame, or cable sheath etc. 3. Connect the Black test lead to the circuit under test or cable core. 4. Confirm that Voltage is not indicated on the instrument. (BM100/4 only) 5. Press and hold the Test push; allow needle to settle, and take the reading from the top scale of the

8、instrument. 6. Release the Test push button. 7. Wait a few moments before disconnecting the test leads, to allow any circuit capacitance to discharge. The BM100/4 will automatically monitor the discharge on its Voltage range, indicating when it is safe to remove the test leads. Resistance Testing (B

9、M101/4 only) 1. Set the selector switch to the 1M resistance range. 2.Connect the test leads across the isolated item / circuit under test. 3. Press and hold the Test push button, allow the needle to settle. Take the reading from the 1M range scale. 4.Release the Test push button and disconnect the

10、test leads. Refer to Safety Warnings before using the instrument RadioFans.CN 收音机爱 好者资料库 SPECIFICATION Range:BM100/4BM101/4 Insulation0,01 - 200 M0,01 - 200 M Resistance-0,01 - 1 M Continuity ()0,01 - 200 0,01 - 200 () 0,01 - 2 0,01 - 2 Insulation Test Voltage:500V d.c.500V d.c. Open Cct. Terminal V

11、oltage: Insulation Range600V d.c.500V at 0,5M)500V at 0,5M) Resistance Range-53V d.c.(nom.) Continuity Ranges ()0,76V d.c.(nom.)0,76V d.c.(nom) ()4,7V d.c.(nom.)4,7V d.c.(nom) Typical Terminal CharacteristicsLimiting value curve Short Cct.Terminal Current: Insulation Range1,6 mA1,6 mA Resistance Ran

12、ge-325 A Continuity Ranges()12,3 mA12,3 mA ()217 mA217 mA Movement:Robust Taut Band type. 250 A F.S.D. A.C. Voltage Measurement (BM100/4 only): The Voltage range is automatically selected when the switch is in any position and the Test push button is not pressed. When connected across the circuit un

13、der test, the BM100/4 will indicate if the circuit is energised (a.c. or d.c.), and measure a.c.voltage up to 600 V. Voltmeter Input Impedance:330 k Automatic Discharge:Capacitive circuits are automatically discharged via an internal (500k) resistor, when the Test push button is released following a

14、n insulation test. Accuracy (at 20C): 2,5% of scale length on all ranges (e.g. 1,9 mm on Insulation range). Insulation range is marked with a white band (Class 1,5 as VDE 0413). Other ranges are Class 3. Humidity:Operation:90% R.H.max. at 20C 80% R.H. max.at 35C Storage95% R.H. max. at 35C Temperatu

15、re Range: Operation: -5C to + 40C RadioFans.CN 收音机爱 好者资料库 Storage-20C to + 70C Temperature Coefficient: 0,1% per C Power Supply:Single 9V Alkaline battery I.E.C. 6 LR61 type Current consumption:110 mA max on Insulation 40mA max on Resistance 220 mA max on 2 range Battery life: Not less than 1800, 5

16、second operations on 2 range. Battery Condition Test:The battery condition test position checks the battery voltage on a simulated load, when the Test push button is pressed. The instrument pointer should deflect clockwise to within the Battery Check arc on the scale. Remove the battery if the instr

17、ument is not to be used for an extended period of time. Battery Replacement: For safety reasons, the battery compartment must not be opened if the test leads are connected. To remove the battery cover, release the screw in the centre of the cover and lever the cover upwards. Replaceable Fuse:500 mA,

18、 250V. H.B.C. type (F) (20mm. x 5mm) To check this fuse, open circuit the test leads, switch to the 200range and press the Test push button. A reading of 200 indicates a ruptured fuse. Located in the battery compartment, this fuse can be replaced by the user. The replacement fuse must be of the corr

19、ect type and rating.To replace the fuse, disconnect the test leads, remove the battery cover by releasing the screw in the centre of the cover and levering the cover upwards. A spare fuse is clipped onto the battery cover. Non Replaceable Fuse:2A - 500 V H.B.C. type (F) 50 kA. Fitted for overall saf

20、ety protection, this fuse is not replaceable by the user. No response from the instrument to an insulation test, with the leads shorted together is likely to mean that this fuse has ruptured. The instrument must be returned to the manufacturer, or an approved agent for repair. Safety:Meets the requi

21、rements for double insulation to IEC1010-1 (1995) EN61010-1 (1995) to installation Category III*, 300 V phase to earth (ground) and 500 V phase to phase, without the need for separately fused test leads. If required, fused test leads are available as an optional accessory. Indicates circuit discharg

22、e after insulation test, 600 V Category II *. * Relates to transient overvoltage likely to be found in fixed installation wiring. * Relates to transient overvoltage likely to be found in portable equipment and appliances. E.M.C.In accordance with IEC61326 including amendment No.1 Standards: BS 7430

23、(1992)-VDE 0413 (1982) Weight:485g Dimensions: 175 mm x 95 mm x 57 mm. Cleaning: Wipe disconnected instrument with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA). ACCESSORIES SuppliedPart Number Test lead set User Guide6172-048 Optional Synthetic Test el bastidor del equipo o rev

24、estimiento del cable, etc. 4. Conecte el cable de prueba Negro al circuito bajo prueba o alma del cable. 5. Confirme que no se indica la Tensin en el instrumento (slo BM100/4). 6. Pulse y mantenga apretado el botn pulsador Test . Deje que a aguja se pare y tome la lectura en la escala superior del i

25、nstrumento. 7. Suelte el botn pulsador Test. 8. Espere unos segundos antes de desconectar los cables de prueba para permitir la descarga de cualquier circuito de capacitancia. El BM100/4 supervisar automticamente la descarga en su gama de Tensin, mostrando por tanto el momento en el que no existe pe

26、ligro para sacar los cables de prueba. Comprobacin de continuidad 1. Regule el interruptor selector a 2 200 , segn proceda. 2. Conecte los cables de prueba por el circuito bajo prueba. 3. Pulse y mantenga apretado el botn pulsador Test. Deje que la aguja se pare. Hay una escala distinta para cada ra

27、ngo, de manera que cada lectura es directa. Tome nota de la lectura en la escala adecuada al instrumento. 4. Suelte el botn pulsador Test y desconecte los cables de prueba. Comprobacin de la resistencia (slo BM101/4) 1. Regule el interruptor selector al rango de resistencia 1M . 2. Conecte los cable

28、s de prueba por el circuito bajo prueba. 3. Pulse y mantenga apretado el botn pulsador Test. Deje que la aguja se pare y tome la lectura en la escala del rango 1M . 4. Suelte el botn pulsador Test y desconecte los cables de prueba. Comprobacin de la tensin (slo BM100/4) La gama de tensin se seleccio

29、na automticamente cuando el interruptor se encuentra en cualquier posicin y NO se ha pulsado el botn pulsador Test. Las autoridades sobre seguridad del Reino Unido han recomendado el uso de cables de prueba protegidos por fusibles para este uso. 1. Conecte los cables de prueba por el circuito bajo prueba. 2. El instrumento indicar si el circuito est activado y el nivel de tensin existente en ese momento (mximo de 600V C.A.) aparecer en la escala inferior. Consultar los avisos de seguridad antes de usar el instrumento