1、M 210600 Insulation clean, and with no broken or cracked insulation. ,Replacement fuses must be of the correct size, type and rating. Symbols used on the instrument Caution: risk of electric shock Caution: refer to accompanying notes Equipment protected throughout by Double Insulation (Class II) NOT

2、E THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. RadioFans.CN 收音机爱 好者资料库 OPERATION Live Circuit Warning The live circuit warning indicator will illuminate if 60 V or more is detected at the terminals. Testing should be aborted if lit. Default Voltage measurement The 2106

3、00 will act as a direct reading default voltmeter (0 to 600 V a.c./d.c.) in the Standby mode (when switched On, and any of the Insulation test positions is selected). The Test push must not be pressed when making a voltage test. Note:- The 210600 is internally fuse protected to 500 V. For fuse prote

4、ction with supplies above 500 V, use Fuse Probe Kit FPK5. 1.Turn the instrument On by pressing the On/Off switch-button once. 2.Select any Insulation test position with the rotary selector switch. V is displayed. 3.Carefully connect the test leads to the circuit under test. Do not press the Test pus

5、h. 4.Read the voltage measurement directly from the display. 5.Carefully disconnect the test leads and turn the instrument Off by pressing the On/Off switch-button once. Low Resistance measurement (k) 1.Turn the instrument On by pressing the On /Off switch-button once. 2.Select k with the rotary sel

6、ector switch. 3.Ensure that all test leads are clean and in good condition, and connect them to the isolated circuit under test. 4.Press the Test push and read the measurement directly from the display. 5.Disconnect the test leads and turn the instrument Off by pressing the On/Off switch-button once

7、. Insulation Testing (M) Automatic circuit discharge is effective when the Test push is released, and decaying voltage is displayed until the circuit is discharged. 1.Turn the instrument On by pressing the On/Off switch-button once. 2.Ensure that all test leads are clean and in good condition, and c

8、onnect them to the isolated circuit under test. 3.Select the required test voltage with the rotary selector switch. 4.Press the Test push and read the measurement directly from the display. Overrange is indicated by . 5.Release the Test push and monitor any displayed voltage to confirm when any disc

9、harging voltage decays to zero. 6.When the circuit has discharged, disconnect the test leads and turn the instrument Off by pressing the On/Off switch-button once. Low Battery Refer to Safety Warnings before using the instrument DISPLAY FEATURES Live Circuit Warning RadioFans.CN 收音机爱 好者资料库 Using the

10、 Guard Terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement, it is unnecessary to use the guard terminal. In cable testing, there may be surface leakage paths across the insulation between the bare cable and the external sheathing due

11、 to the presence of moisture or dirt. Where it is required to remove the effect of this leakage, particularly at high testing voltages, a bare wire may be bound tightly around the insulation and connected via the third test lead to the guard terminal G. The guard terminal is at the same potential as

12、 the negative terminal. Since the leakage resistance is effectively in parallel with the resistance to be measured, the use of the guard causes the current flowing through surface leakage to be diverted from the measuring circuit. The instrument therefore reads the leakage of the insulator, ignoring

13、 leakage across its surface. SPECIFICATION Insulation Ranges Measuring Ranges:0 - 1.999M at all test voltages 19.99 M at all test voltages 199.9 M at all test voltages 1999 M at 500 V and 1000 V d.c. test voltages Fully auto - ranging at all voltages Test Voltages (d.c.):100 V; 250 V; 500 V; 1kV Tes

14、t V. Accuracy:5% at open circuit Short Cct. Current:1.2 mA nominal Accuracy (73.4F 3.6F): 500 M: 3% 2 digits 500 M to 1999 M: 5% 2 digits Temp. Co-efficient:86F 0.14% per F 500 M to 1999 M: 86F 0.11% per F, 86F, 0.28% per F, Hum Rejection:Within twice the normal accuracy if the noise current does no

15、t exceed 1 mA for the 1 kV range, or 0.5 mA for the other ranges and is within 45 Hz to 65 Hz. Low resistance Range Measuring Range:0 - 1.999 k 19.99 k Open Cct. Voltage:4.3 V nominal Accuracy (73.4F 3.6F): 2% of reading 1digit 1100 1000 900 800 700 600 500 400 300 200 100 0 0.010.111010010001000010

16、0000 Terminal Characteristics Leakage path to +ve terminal to G terminal to -ve terminal Tightly bound bare wire RadioFans.CN 收音机爱 好者资料库 M Archcliffe Road Dover Kent CT17 9ENTel: +44 (0) 1304 502100 EnglandFax: +44 (0) 1304 207342 PO Box 9007 Valley Forge PA 19484-9007Tel: +1 (610) 676-8500 U.S.A.Fa

17、x: +1 (610) 676-8610 4271 Bronze Way Dallas TX 75237-1017Tel: +1 (800) 723-2861 (U.S.A. only) U.S.A.Tel: +1 (214) 330-3203 (International) Fax: +1 (214) 337-3038 110 Milner Avenue, Unit 1 Scarborough ON M1S 3R2Tel: +1 (800) 567-0286 CanadaFax: +1 (416) 298-0848 This instrument is manufactured in the

18、 United Kingdom. The company reserves the right to change the specification or design without prior notice. Megger is a registered trademark. Part No. 6172-373 - Edition 4 - Printed in England - 11HH Short Cct. Current:2.75 mA nominal Temp. Coefficient:0.06% per F Default Voltage measurement Range:0

19、 - 600 V a.c. / d.c. (effective on Insulation ranges with Test push not pressed) Input impedance:300 k to 1 M depending on input voltage Discharge Rate:A 1 F capacitor will discharge to a safe level in 2 s Accuracy:2% of reading 1 digit d.c. or a.c. at 45 Hz to 65 Hz Temp. Coefficient:0.06% per F Ge

20、neral Specifications Overload rating:The 210600 is protected for connection to Power distribution systems up to 300 V Line - Ground, and 500 V Line - Line for Installation Category III*. Temp. Range: Operating: 23F to 122F Storage:4F to 158F Humidity: Operating: 70% RH at 68F, 60% RH at 95F, 50% RH

21、at 104F for measurements 200 M Storage:95% RH at 95F Display:312digit L.C.D. Maximum reading 1999 Automatic Discharge: Capacitive circuits are automatically discharged when the Test push is released following an insulation test Power Supply:6 x 1.5 V AA ( Alkaline) type cells Battery life:Typically

22、2100 x 5 second operations (worst case) Note:- Battery cells should not be left in an instrument which may remain unused for extended periods of time. Low Battery Indicator: The low battery indicator symbolwill appear when the battery cells are exhausted. Battery Replacement: The rear cover must not

23、 be opened if the test leads are connected. To remove the rear cover, release the screw at the bottom of the cover and lift the cover upwards.To avoid the possibility of shock, do not press the Test push when changing batteries. Fuses:2 x 500 mA 500 V (F) H.B.C.10 kA min 114“ x 14“ To check the +ve

24、fuse, short the +ve and -ve test leads and press the Test push. An reading indicates that the +ve fuse has ruptured. Fuse Replacement: Held in screw type holders located in the base of the instrument. Use a flat blade screwdriver to release the center part of each fuse holder. Remove the fuses and t

25、est each in turn. Replace with fuse(s) of the correct type, size and rating. Weight:2.2 lb (including batteries) Dimensions:7” x 5” x 5” Cleaning:Wipe disconnected instrument with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA) *Relates to transient overvoltage likely to be found

26、in fixed installation wiring. 0.5A(F) 500V HBC(10 kA) Guard fusePositive fuse ACCESSORIES SuppliedPart Number User Guide6172-373 Battery - 6 x 1.5 V AA (Alkaline) type Test Record Card (5 supplied) Optional Carrying case 217740 Fuses 500 mA, 500 V (F) H.B.C. (pack of 5)6121-289 Fused probe kit FPK46

27、111-287 Test lead set comprising:- 1 black, 1 red, 1 green test lead, with alligator clips 6220-436 Test Record cards (Pack of 20)6111-216 Publication - A Stitch in TimeAVTM21-P8B REPAIR AND WARRANTY The instrument contains static sensitive devices, and care must be taken in handling the printed cir

28、cuit board. If an instruments protection has been impaired it should not be used, but sent for repair by suitably trained and qualified personnel. The protection is likely to be impaired if for example; it shows visible damage; fails to perform the intended measurements; has been subjected to prolon

29、ged storage under unfavourable conditions, or has been subjected to severe transport stresses. NEW INSTRUMENTS ARE GUARANTEED FOR 1 YEAR FROM THE DATE OF PURCHASE BY THE USER. Note: Any unauthorized prior repair or adjustment will automatically invalidate the Warranty. INSTRUMENT REPAIR AND SPARE PA

30、RTS For service requirements for Megger Instruments contact :- Megger Limited orMegger Valley Forge Corporate CenterArchcliffe Road 2621 Van Buren AvenueDover Norristown, PA 19403Kent, CT17 9EN U.S.A.England Tel: +1 (610) 676-8579Tel: +44 (0) 1304 502243 Fax: +1 (610) 676-8625Fax: +44 (0) 1304 20734

31、2 or an approved repair company. Approved Repair Companies A number of independent instrument repair companies have been authorised for repair work on most Megger instruments, using genuine Megger spare parts. Consult the Appointed Distributor / Agent regarding spare parts, repair facilities, and ad

32、vice on the best course of action to take. Returning an Instrument for Repair If returning an instrument to the manufacturer for repair, it should be sent freight pre-paid to the appropriate address. A copy of the invoice and of the packing note should be sent simultaneously by airmail to expedite clearance through Customs. A repair estimate showing freight return and other charges will be submitted to the sender, if required, before work on the instrument commences. MeterCenter 2046 West Peninsula Circle Chandler, AZ 85248 Telephone (480) 659-8351 Toll Free (800) 230-6008 Fax (480) 659-8361