1、G Safety Warnings and Precautions These must be read and understood before the instrument is used. They must be observed during use. IThe circuit under test must be switched off, de-energised and isolated before test connections are made when carrying out insulation and continuity tests. ICircuit co
2、nnections and exposed metalwork of an installation or equipment under test must not be touched. Remember remote conductors may be charged to the test voltage. IThe live circuit warning and automatic discharge are additional safety features which may fail and therefore safe working procedures/practic
3、es must be observed. IThe voltage function will only work if the instrument is functional and switched on. IAfter insulation tests, capacitive circuits must be allowed to discharge before disconnecting test leads. IThe instrument, test-leads, probes and crocodile clips must be in good order, clean a
4、nd with no broken or cracked insulation. IEnsure that hands remain behind guards of probes/clips when testing. INational Safety Authorities may recommend the use of fused test leads when measuring voltage on high-energy systems. IReplacement fuses must be of the correct type and rating. Failure to o
5、bserve this will result in a safety hazard. IThe rear cover must be in place whilst conducting tests. NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. M MIT200 Series Insulation and Continuity Testers User guide Users of this equipment and/or their employers are remin
6、ded that National Health and Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits. Where the assessments show that the risk is significant
7、then the use of fused test leads may be appropriate. Symbols used on the instrument: G Caution: refer to accompanying notes t Equipment protected throughout by Double Insulation (Class II) c Equipment complies with relevant EU Directives Equipment complies with C tick requirements Do not dispose of
8、in the normal waste stream Maximum input voltage 600 V rms Application BS EN 61010 defines measurement categories from I to IV relating transient over voltages and the location within the electrical installation. This instrument is designed for use at Category III (Building installation/distribution
9、 board level) up to 600 V (maximum) rms to earth and 600 V (maximum) rms between terminals. It may also be used at Category IV (Primary supply level) on 300 V rms phase to earth systems. Display layout Operation All circuits must be isolated before an insulation or continuity test. Automatic Voltage
10、 detection The instrument will automatically display voltage between 25 V and 600 Vac or dc when connected to the circuit under test. Remove the source of any voltage before continuing to test. Test lead check 1.Before each use of the instrument, inspect the test leads, prods and crocodile clips to
11、confirm that their condition is good, with no damaged or broken insulation. 2.Check continuity of the test leads by firmly shorting the leads together and read the test lead resistance measurement directly from the display. Backlight operation 1.Turn the instrument backlight on by selecting the posi
12、tion. Backlight stays on for 1 minute. 2.When the backlight is activated, select the desired test position. 3.Change range or press TEST to re-initialise backlight for another minute. Test lead null To remove test lead resistance before continuity tests: 1.Switch the instrument to 2.Short test leads
13、 together, wait for a stable reading and press the test button on the instrument. The display should read 0.00 and display the z symbol. Lead null value is automatically stored until reset. To remove, press the TEST button again with leads open circuit. Continuity Measurement 1.Set range switch to .
14、 2.Connect the red and black test leads to the circuit to be tested. The instrument will automatically display continuity resistance when connected (up to 99.9 ). NOTE: If “100.0 “ is displayed the resistance of the circuit is over-range. Continuity Buzzer 1.Set the instrument to range. 2.Connect th
15、e test leads to the circuit to be tested. 3.The buzzer will sound automatically when a circuit is made (2 V is detected the test will be inhibited. Insulation testing M all instruments 1.Set range switch to 250 V, 500 V or 1 kV as available. 2.Press and hold down the TEST button to start test. The d
16、isplay will settle, and show the insulation test resistance in M. 3.If a circuit voltage of 25 V is detected the instrument will give a warning Beep but allow testing to continue. If a circuit voltage 50 V is detected, further testing will be prevented. Warning: After testing ensure the display retu
17、rns to Zero (circuit is discharged) before removing test leads. Error messages: 100.0 Continuity over-range 1000 M Insulation test result is over-range UNC Uncalibrated E17Instrument not configured to a specific type. Battery replacement Low battery voltage is indicated by the symbol in the display.
18、 To maintain performance the batteries should be replaced at this point. Uses 6 x 1.5 V AA (LR6) alkaline dry cells or NiMH rechargeable batteries. To replace batteries switch off instrument and disconnect test leads from the circuit under test. Remove the rear battery cover and replace the batterie
19、s. Do not mix old and new batteries. Warning: OBSERVE CORRECT POLARITY. Incorrect battery polarity could cause batteries to leak and damage the instrument. Fuse replacement A blown fuse is indicated by the symbol in the display. To replace the fuse remove battery cover and replace fuse with type: 50
20、0 mA (F) HBC 50 kA 600 V . Cleaning: Wipe disconnected instrument with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA). Specification Insulation ranges Test voltage accuracy:-0% +25% over full operating temperature Nominal Test Voltage:1000 V , 500 V , 250 V (d.c.) Measuring Range
21、:10 k - 1000 M on all ranges Short Circuit Current: 2 mA +0 mA -50% Test Current on Load:1 mA at min. pass values of insulation (as specified in BS 7671, EN61557, HD 384 and IEC 364) Accuracy (at 20 C)3%, 2 digits (or 30% of reading 200 k to 10 M) Continuity ranges Measuring Range:0,01 - 100 (0 -50
22、on analogue scale) Open Circuit Voltage: 5 V 1 V Short Circuit Current:205 mA, 5 mA (0 to 10 ) (20 mA up to 100 ) Accuracy (at 20 C):3% of reading 2 digits up to 10 5% of reading 2 digits up to 100 M 30 of reading 2 digits up to 1000 M Null Offset Adjust: 0 9.99 Continuity Buzzer:Operates at 5 Respo
23、nse time 25 V a.c. or d.c. applied display will operate as a voltmeter Test inhibit:50 V AutoPower Down: Operates after 10 minutes if left in standby mode. Environmental Operating Range:-10C to +55C Operating Humidity:93% R.H. at +40C max. Storage Range:-25C to +65C Dimensions:195 mm x 98 mm x 40 mm
24、 Weight:550 gms Battery:6 x 1,5 V cells IEC LR6 type (AA alkaine) or NiMH rechargeable Battery life:3000 consecutive tests (5 sec/test) on any test using 2Ah batteries Fuses Terminals:500 mA (F) 600 V , 32 x 6 ceramic HBC 50 kA minimum Safety Meets the requirements of BS EN61010-1, Cat III 600 V pha
25、se to earth. Refer to safety warnings supplied. Automatic discharge After an insulation test, the circuit under test will be discharged automatically. Any voltage present will be indicated on the display so that the discharge can be monitored. EMC In accordance with IEC 61326 including amendment No.
26、1 Included accessories Test lead set (red and black) with crocodile clips6220-779 Hand carry case5410-419 Optional accessories 2 wire fused probe and clip set6220-789 Rubber holster (Boot)5410-346 Repair and Warranty The instrument contains static sensitive devices, and care must be taken in handlin
27、g the printed circuit board. If an instruments protection has been impaired it should not be used, but sent for repair by suitably trained and qualified personnel. The protection is likely to be impaired if for example, it shows visible damage, fails to perform the intended measurements, has been su
28、bjected to prolonged storage under unfavourable conditions, or has been subjected to severe transport stresses. NEW INSTRUMENTS ARE GUARANTEED FOR 3 YEARS FROM THE DATE OF PURCHASE BY THE USER. Note: Any unauthorized prior repair or adjustment will automatically invalidate the Warranty. CALIBRATION,
29、 REPAIR AND SPARE PARTS For service requirements for Megger Instruments contact: Megger LimitedorMegger Archcliffe Road Valley Forge Corporate Centre Dover 2621 Van Buren Avenue Kent CT17 9ENNorristown PA 19403 England.U.S.A. Tel: +44 (0) 1304 502 243Tel: +1 610 676 8579 Fax: +44 (0) 1304 207 342 Fa
30、x: +1 610 676 8625 Megger operate fully traceable calibration and repair facilities, ensuring your instrument continues to provide the high standard of performance and workmanship you expect. These facilities are complemented by a worldwide network of approved repair and calibration companies to off
31、er excellent in-service care for your Megger products. Returning your product to Megger - UK and USA service centres 1.When an instrument requires recalibration, or in the event of a repair being necessary, a Returns Authorisation (RA) number must first be obtained from one of the addresses shown ab
32、ove. You will be asked to provide the following information to enable the Service Department to prepare in advance for receipt of your instrument, and to provide the best possible service to you. IModel, e.g. MIT200. ISerial number, to be found on the underside of the case or on the calibration cert
33、ificate (e.g. 6111-357/050305/1234) IReason for return, e.g. calibration required, or repair. IDetails of the fault if the instrument is to be repaired. 2.Make a note of the RA number. A returns label can be emailed or faxed to you if you wish. 3.Pack the instrument carefully to prevent damage in tr
34、ansit. 4.Ensure the returns label is attached, or that the RA number is clearly marked on the outside of the package and on any correspondence, before sending the instrument, freight paid, to Megger. Copies of the original purchase invoice and packing note should be sent simultaneously by airmail to
35、 expedite clearance through customs. In the case of instruments requiring repair outside the warranty period, an immediate quotation can be provided when obtaining the RA number. 5.You may track the progress of your return on line at Approved Service Centres A list of Approved Service Centres may be
36、 obtained from the UK address above, or from Meggers website at Megger Limited Archcliffe Road, Dover Kent CT17 9EN England T +44 (0)1 304 502101 F +44 (0)1 304 207342 E Megger 4271 Bronze Way, Dallas, Texas 75237-1019 USA T +1 800 723 2861 (USA ONLY) T +1 214 333 3201 F +1 214 331 7399 E Megger Z.A
37、. Du Buisson de la Couldre 23 rue Eugne Henaff 78190 TRAPPES France T +33 (0)1 30.16.08.90 F +33 (0)1 34.61.23.77 E Megger Pty Limited Unit 26 9 Hudson Avenue Castle Hill Sydney NSW 2125 Australia T +61 (0)2 9659 2005 F +61 (0)2 9659 2201 E Megger Limited 110 Milner Avenue Unit 1 Scarborough Ontario
38、 M1S 3R2 Canada T +1 416 298 9688 (Canada only) T +1 416 298 6770 This instrument is manufactured in the United Kingdom. The company reserves the right to change the specification or design without prior notice. Megger is a registered trademark Part No. 6172-953 V02 Printed in England 0106 600V G An
39、alogue arc display Battery level Digital displayLead null indicator Fuse blown f G CONSIGNES DE SCURIT Lire attentivement les consignes de securit avant dutiliser lappareil. Ces consignes doivent tre respectes lors de toute utilisation. ILe circuit test doit tre coup, mis hors-tension et isol avant
40、deffectuer les raccordements de test lors des tests disolation et de continuit. INe pas toucher les raccordements des circuits et les parties mtalliques exposes de linstallation ou de lquipement test. ILavertissement indiquant un circuit sous-tension ainsi que la dcharge automatique sont des disposi
41、tifs de scurit complmentaires qui peuvent tomber en panne. Il est donc ncessaire de respecter les consignes de scurit. ILa fonction tension ne fonctionne que si lappareil est sous- tension et en bon tat de marche. IAprs chaque test disolation, les condensateurs doivent tre dchargs avant de dconnecte
42、r les cbles de test. ILes cbles de test, les sondes et les pinces crocodile doivent tre propres et en bon tat. Lisolant ne doit prsenter ni cassures, ni criques. ILors des tests, garder les mains derrire les protections des sondes ou des pinces. ILes autorits locales en charge de la scurit peuvent r
43、ecommander lutilisation de cbles de test protgs par fusibles lors de la mesure de la tension sur des circuits de haute nergie. ILes fusibles de rechange doivent tre adapts (amprage, type). La pose de fusibles inadapts peut prsenter un danger et endommager lappareil en cas de surcharge lectrique. ILe
44、 couvercle des piles doit tre pos lors de la conduite des tests. NOTA CET APPAREIL NE DOIT ETRE UTILISE QUE PAR DES PERSONNES COMPETENTES ET SPECIALEMENT FORMEES. M Sries MIT200 Testeurs numriques et analogiques de continuit et disolation User guide La lgislation en vigueur sur la scurit et la sant
45、impose aux utilisateurs de ces quipements deffectuer pralablement une valuation des risques sur tous les travaux lectriques, afin didentifier les sources potentielles de danger et les risques daccidents comme un court-circuit. Si cette valuation rvle un niveau de risque significatif, lutilisation de
46、 cbles de test protgs par fusibles peut tre ncessaire. Symboles affichs: G Attention: Consulter le manuel dutilisateur t Equipement entirement proteg par une double isolation (Classe II) c Equipement conforme aux directives europennes applicables Conformit C tick (Australie) Ne pas jeter la poubelle
47、 Tension dentre maximale 600 V RMS Application BS EN 61010 dfinit des catgories de mesure de I IV reliant la coupure au-dessus des tensions et lendroit dans linstallation lectrique. Cet appareil est conu pour une usage de catgorie III (installation et distribution dans la construction) jusqu 600 V (
48、maximum) RMS de mettre la terre et 600 V (maximum) RMS entre les bornes. Il peut galement tre utilis la catgorie IV (niveau primaire dapprovisionnement) la phase de 300 V RMS pour mettre des systmes la terre. Configuration de lcran Utilisation Tous les circuits doivent tre isol avant deffectuer un t
49、est disolation ou de continuit. Dtction automatique de tension Lappareil affichera automatiquement la tension entre 25 V et 600 V CA ou CC lorsquil sera connect au cble test. Mettre a lcart toute source de tension avant de continuer le test. Contrle de fil test 1. Avant chaque utilisation de lappareil, vrifier les fils test, pinces crocodiles, et sondes afin de sassurer que tous sont en bon tat. 2. Contrler la continuit des cbles tests en maintenant fermement les cbles ensemble et lire la mesure de la rsistance directement sur lcran. Rtro-clairage