1、Optical Power Meter MEGGER MPM2000/2000H User Guide Guide de lutilisateur Benutzerhandbuch Guida per lutente Gua del Usuario Gebruikersgids Anvndarmanual Brugervejledning Brukerhndbok Kyttohjeet MEGGER RadioFans.CN 收音机爱 好者资料库 2 Contents Introduction4 User Controls and Display5 Operation6 Batteries7
2、Specifi cation8 Accessories9 Care and Maintenance11 Repair and Warranty12 Symbols used on the Instrument are: Caution: refer to accompanying notes. Equipment complies with relevant EU Directives. RadioFans.CN 收音机爱 好者资料库 SAFETY WARNINGS This instrument is designed for testing fi bre optic installatio
3、ns and so meets the safety requirements of IEC 60825-1: 1994 (including amendment A11: 1996) and IEC 60825-2: 1995 (including amendment A1: 1998). This products output levels are within the Accessible Emission Limits (AEL) for a Class 1 Laser Product and so does not generate any optical hazards inte
4、rnally. However it can be used to test optical fi bres in environments where light sources that would cause an optical hazard could be present. This corresponds to working within the AEL for Class 1, Class 2, Class 3A and Class 3B lasers, therefore, please take note of the following safety warnings:
5、 Under no circumstances should a connector end-face, prepared optical fi bre or fractured optical fi bre be viewed directly unless the power received from the fi bre is known to be safe (as defi ned by IEC 60825) and under local control. AVOID EXPOSURE TO BEAM if the accessible radiation is within t
6、he AEL for Class 3B lasers. DO NOT STARE INTO BEAM OR VIEW DIRECTLY WITH OPTICAL INSTRUMENTS if the accessible radiation is within the AEL for Class 3A lasers. DO NOT STARE INTO BEAM if the accessible radiation is within the AEL for Class 2 lasers. The instrument should not be used if any part of it
7、 is damaged. Refer to operating instructions for further explanation and precautions. Safety Warnings and Precautions must be read and understood before the instrument is used. They must be observed during use. NOTE: THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. RadioFa
8、ns.CN 收音机爱 好者资料库 4 Introduction Thank you for purchasing this quality AVO product. Before attempting use of your new instrument please take the time to read this user guide, ultimately this will save you time, advise you of any precautions you need to take and could prevent damage to yourself and th
9、e instrument. The MPM2000 and MPM2000H are advanced optical power meters that can be used for optical loss testing of fi bre optic cables. These have been pre-calibrated for absolute power levels with reference to 1mW (dBm) for 850nm, 1300nm, 1310nm and 1550nm laser wavelengths. A further ten wavele
10、ngths can be programmed in at the factory. With its high-resolution reading (to 0.01 dBm) and wide dynamic range (+10dBm to 70dBm for MPM 2000, +20dBm to 60dBm for MPM 2000H) this instrument favours single mode operation and is supplied with a single- mode patch lead. However it can also be used wit
11、h multi-mode fi bre optic cables and corresponding multi-mode patch leads are available in the Accessories. With such a wide dynamic resolution, this product is ideal for CATV and other long distance transmission applications. 5 1.0 User Controls and Display 1.1 ON/OFF Switch To turn the unit on, de
12、press the ON/OFF switch. To disable the Auto-Power-Off during switch on, hold the ON/OFF switch down until the Auto-Off annunciator on the LCD disappears. In Auto-Power-Off mode, the unit will power off after 15 minutes unless a switch is pressed. To turn the unit off, press the ON/OFF switch. 1.2 Z
13、ero Switch The zero switch is used to automatically zero the optical power meter before making power level measurements in the -45 to 60dBm (H series) or -55 to -70 dBm range. To zero the meter, fi t a blanking adapter to the sensor holder and then depress the zero switch. On completion the LCD indi
14、cates -62.00 dBm (H series) or -72.00 dBm and UR (under range) is displayed. NOTE. When making low level power measurements in the last 10dB of dynamic range, allow more time for the meter indication to settle due to extra auto averaging. 1.3 dBm / Watts Switch The dBm /Watts switch selects the meas
15、urement mode and returns the unit to dBm measurement mode if dB REL is selected. 1.4 Wavelength Switch () The wavelength switch allows the correct wavelength to be selected prior to making a measurement as indicated on the small 4-digit display. NOTE: The wavelength switch is disabled during dB REL
16、measurements 1.5 dB REL Switch The dB REL switch allows measurements to be made relative to the previously stored reference power level, indicated on the small 4-digit display. All further measurements will be referenced against this previously stored power level. To store a new reference power leve
17、l, press and hold the dB REL switch for 3 seconds. After the 3 second delay the main 4-digit display will indicate 0.00 and the new stored reference level will be indicated on the small 4- digit display. To return to dBm measurement mode, press the dBm/Watts switch. 6 2.0 Operation This instrument c
18、an be used for multiple applications including Optical Loss Testing, Optical Continuity Testing, Acceptance Testing of Fibre Transmitters and Received Power Testing of optical receivers. There are multiple techniques to achieve these types of tests, but a typical procedure is detailed below: 2.1 Opt
19、ical Loss Testing Take an optical light source and power meter with appropriate patch cords and mating adapter (uniter) to match the optical fi bre under test. Clean all the connectors and connect the light source via the fi rst patch cord, the uniter and the second patch cord to the power meter so
20、the attenuation of both cords can be eliminated. The link to be tested is then inserted between them so the optical loss associated only with this link can be found. This test is normally repeated multiple times with the interconnections between the patch cord and the fi bre under test disconnected
21、and reconnected before each time and an average result taken to ensure repeatability and stability. 2.2 Optical Continuity Testing Use an optical light source (of a known output power) and power meter with patch leads of a known optical loss and a uniter matched to the optical fi bre under test. Cle
22、an all the connections and insert the link to be tested between the light source and the power meter. Measure the output of the light source through the fi bre under test and check that it is above a pre- determined power level. This is a more simple go, no go test. 2.3 Acceptance Testing of Fibre T
23、ransmitters Use a power meter with a suitable dynamic range for the transmitter under test and patch leads of a known optical loss. Clean all the connections and connect the power meter to the transmitter via the patch cord. Turn on the transmitter and allow the output to stabilise before measuring
24、the output and comparing the result to an acceptance standard or versus historical data. This test is normally repeated multiple times with the interconnections disconnected and reconnected before each time and an average result taken to ensure repeatability and stability. 7 2.4 Received Power Testi
25、ng Use a power meter with a suitable dynamic range for the receiver under test and an adapter to match the connections on the receiver. Clean all the connections and connect the power meter to the fi bre supplying the receiver under test so that the power meter is now where the receiver would normal
26、ly go. Turn on the optical transmitter of the fi bre system and allow the output to stabilise before measuring the output at the receiver and comparing the result to the input specifi cations of the receiver. This test is normally repeated multiple times with the interconnections disconnected and re
27、connected before each time and an average result taken to ensure repeatability and stability. The MPM2000/MPM2000H are supplied and confi gured for use with a 6LR61 (PP3) dry cell. A 12V DC charger (See Optional Accessories) can be used as a dual power source; this will not apply a charge to the dry
28、 cells while fi tted. Alternatively, within the battery compartment there is a switch to change the battery type from alkaline to NiCd/NiMh. If the DC charger is now connected to the instrument with rechargeable batteries in it (the switch is set to NiCd/NiMh) then the batteries will start to rechar
29、ge. Allow 14 16 hours to fully charge a rechargeable battery. Do NOT attempt to recharge alkaline cells, over heating would occur and the instrument may be damaged. When the switch is set to NiCd/NiMh, only fi t rechargeable cells. 3.0 Batteries 8 4.0 Specifi cation 4.1 Optical Specifi cation Sensor
30、 Type:Germanium Dynamic Range:+10 to -70 dBm (MPM2000), +20 to -60 dBm (MPM2000H) dBrel Mode:Yes Wavelength:850nm, 1300nm, 1310nm and 1550nm standard, 10 other wavelengths can be factory programmed Measurement Accuracy:5% -23 dBm Resolution:0.01 dBm Display:Digit Custom LCD, displays dBm, dB, Watts,
31、 wavelength, OR, UR, LO Batt, Auto off 4.2 General Specifi cation Safety:This instrument is designed for testing fi bre optic installations and so meets the safety requirements of IEC 60825-1: 1994 (including amendment A11: 1996) and IEC 60825-2: 1995 (including amendment A1: 1998). Though this prod
32、uct does not generate any optical hazards internally, it can be used to test light sources that are of sufficient power and wavelength that an optical hazard may exist. EMC: Complies with Electromagnetic Compatibility Specifi cations BS/EN61326-1-1997 - Performance Criteria A (Light industrial - Cla
33、ss B) Batteries: 1 6LR61 (PP3), Manganese-alkali or nickel-cadmium or nickel-metal-hydride cells (For dry cells, the switch in the battery compartment must be set to Alkaline.) Battery Consumption:20mA nominal Power Down:Automatic after 20 minutes with no key-press in dBm mode Case Dimensions:160mm
34、x 83mm x 30mm Instrument Weight:230g Display:3 Digit LCD, displays dBm, dB Operational Temperature: -10 C to +50 C (14 F to 122 F) Operational Humidity: 95% at 40 C (104 F) Storage Temperature: -20 C to 70 C (-4 F to158 F) 9 Optional Accessories:Part No. Single Mode Patch Cord with FC / SC Connector
35、s (1 metre)22440-007 Single Mode Patch Cord with FC / FC Connectors (1 metre)22440-008 Single Mode Patch Cord with FC / ST Connectors (1 metre)22440-009 Single Mode Patch Cord with ST / ST Connectors (1 metre)22440-015 Single Mode Patch Cord with SC / SC Connectors (1 metre)22440-016 Multimode Patch
36、 Cord with FC / FC Connectors (1 metre)22440-010 Multimode Patch Cord with FC / ST Connectors (1 metre)22440-011 Multimode Patch Cord with FC / SC Connectors (1 metre)22440-012 Multimode Patch Cord with ST / ST Connectors (1 metre)22440-013 Multimode Patch Cord with ST / SC Connectors (1 metre)22440
37、-014 Multimode Patch Cord with SC / SC Connectors (1 metre)22440-017 Multimode Patch Cord with ST / SMA 9mm Connectors (1 metre)22440-018 Multimode Patch Cord with FC / SMA 9mm Connectors (1 metre)22440-023 Multimode Patch Cord with SC / SMA 9mm Connectors (1 metre)22440-024 MPM Power Meter Adaptor
38、with FC connector6220-715 MPM Power Meter Adaptor with ST connector6220-716 MPM Power Meter Adaptor with SC connector6220-717 In-line gender changing adaptor with Multimode ST connectors22440-019 5.0 Accessories 5.0 Accessories (cont). In-line gender changing adaptor with Single mode SC connectors22
39、440-020 In-line gender changing adaptor with Multimode SC connectors22440-021 In-line gender changing adaptor with Single mode ST connectors22440-022 In-line gender changing adaptor with Single mode FC connectors22440-023 Bare Fibre adaptor SC connector25970-034 Bare Fibre adaptor FC connector25970-
40、035 Bare Fibre adaptor ST connector25970-036 *Bare Fibre adaptor Single mode / Multimode body25970-037 UK DC Power Supply27900-050 USA DC Power Supply27900-051 EU DC Power Supply27900-052 Rechargeable Battery 25985-022 Included Accessories:Part No. User Guide6172-629 Test ceci ne chargera pas les pi
41、les sches lorsquelles sont utilises. Alternativement, dans le compartiment des piles se trouve un commutateur pour modifi er le type de piles dalcalines en NiCd/NiMh. Si le chargeur en courant continu est maintenant connect linstrument avec des piles rechargeables (le commutateur est plac sur NiCd/N
42、iMh), les piles commenceront alors se recharger. Il faut entre 14 et 16 heures pour entirement recharger une pile rechargeable. Nessayez pas de recharger des piles alcalines, il se produirait une surchauffe et linstrument pourrait tre endommag. Lorsque linterrupteur est plac sur NiCd/NiMh, nemployer
43、 que des piles rechargeables. 22 4.0 Specifi cations 4.1 Spcifi cations optiques Type de capteur :Germanium Gamme dynamique : +10 -70 dBm (MPM2000), +20 -60 dBm (MPM2000H) Mode dBrel Mode :Oui Longueur donde :850nm, 1300nm, 1310nm et 1550nm classiques, 10 autres longueurs donde peuvent tre programme
44、s en usine. Prcision de la mesure: 5% -23 dBm Rsolution :0,01 dBm Affichage :cran cristaux liquides personnalis, affichages dBm, dB, Watts, longueur donde, OR, UR, Batteries faibles, hors tension automatique 4.2 Spcifi cations gnrales Scurit :Cet instrument est conu pour mesurer les installations fi
45、 bres optiques et rpond ainsi aux exigences en matire de scurit de la CIE n 60825- 1 de 1994 (y compris lamendement y A11 de 1996) et n60825-2 de 1995 (y compris lamendement A1 de 1998). Bien que ce produit ne cr aucun danger optique en interne, il peut tre employ pour mesurer les sources lumineuses
46、 qui ont une puissance et des longueurs donde suffisantes pour quun danger optique puisse exister. CEM : Conforme aux Prescriptions de Compatibilit Electromagntique NF/EN61326-1-1997 Critres de Performance A (Industriel lger - Classe B) Piles : 1 6LR61 (PP3). Piles alcalines au manganse ou au nickel
47、-cadmium ou aux hydrures mtalliques de nickel (pour les piles sches, le commutateur dans le compartiment des piles doit tre positionn sur alcaline.) Consommation des piles : nominal 20mA Mise hors tension : automatique aprs 20 minutes sans slection du mode dBm Dimensions du botier : 160mm x 83mm x 3
48、0mm Poids de linstrument : 230g Ecran:3 crans cristaux liquides avec affichage dBm, Db Temprature de fonctionnement : -10 C +50 C (14 F 122 F) Humidit de fonctionnement : 95% 40C (104F) Temprature de stockage : -20 C 70 C (-4 F to158 F) 23 5.0 Accessoires Accessoires en option:Numro de pice. Cordon
49、de branchement Mode simple avec connecteurs FC / SC (1 mtre)22440-007 Cordon de branchement Mode simple avec connecteurs FC / FC (1 mtre)22440-008 Cordon de branchement Mode simple avec connecteurs FC / ST (1 mtre)22440-009 Cordon de branchement Mode simple avec connecteurs ST/ ST (1 mtre)22440-015 Cordon de branchement Mode simple avec connecteurs SC / SC (1 mtre)22440-016 Cordon de branchement Multimode avec connecteurs FC / FC (1 mtre)22440-010 Cordon de branchement Multimode avec connecteurs FC / ST (1 mtre)22440-011 Cordon de branche