《FLUKE 80K-6 Instruction 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 80K-6 Instruction 电路图.pdf(2页珍藏版)》请在收音机爱好者资料库上搜索。
1、LIMITED WARRANTY & LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warrant
2、y extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Flukes opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation or handling. Fluke w
3、arrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non- defective media. Fluke does not warrant that software will be error free or operate without interruption. Fluke authorized resellers shall exte
4、nd this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international pric
5、e. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country. Flukes warranty obligation is limited, at Flukes option, to refund of the purchase price, free of charge repair, or replace
6、ment of a defective product which is returned to a Fluke authorized service center within the warranty period. To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insurance prepaid (FOB Destination),
7、to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormal conditi
8、on of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipp
9、ing Point). THIS WARRANTY IS BUYERS SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSE
10、QUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential d
11、amages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision. Fluke CorporationFluke Eu
12、rope B.V. P.O. Box 9090P.O. Box 1186 Everett, WA 98206-90905602 BD Eindhoven U.S.A.The Netherlands 5/94 80K-6 High Voltage Probe Instruction Sheet Description The 80K-6 is a high voltage probe designed to extend the voltage measuring capability of an ac/dc voltmeter to 6000 volts peak AC or DC Overv
13、oltage Category I. This means the probe can only be used to make measurements on energy limited circuits within equipment. Examples include high voltage within televisions or photo copy machines. DO NOT use this probe to measure high voltages on power distribution systems. A 1000:1 voltage divider p
14、rovides the probe with a high input impedance. The divider also provides high accuracy when used with a voltmeter having a 10 megohm input impedance. A molded plastic body houses the divider and protects the user from the voltage being measured. Specifications The 80K-6 will achieve rated accuracy w
15、hen it is used with an ac/dc voltmeter having 0.25% accuracy, or better, and an input impedance of 10 megohms 10%. Specifications for the probe are as follows: Voltage Range: 0 to 6 kV, dc or peak ac Input Impedance: 75 megohms nominal Division Ratio: 1000:1 Accuracy: DC to 500 Hz:1% 500 Hz to 1 kHz
16、:2% Above 1 kHz:Output reading falIs. Typically, -30% at 10 kHz. Temperature: Storage: -20 to 60C Operating:0 to 50C Temperature Coefficient: 100 ppm/C Relative Humidity: 20 - 80% Altitude:2000 m Shock and Vibration: Shock:1 m Drop Vibration:5-55 Hz swept sine 3 g max. Dimensions:24.89 cm (9.8 in) (
17、probe body), 5.10 cm (2.0 in) max. width Weight:12.7 g (45 oz) Safety: CAN/CSA C22.2 No. 1010.2.031-94 & IEC 1010-2-031:1993, Type B, 6 kVDC or peak AC, 4.24 kV rms AC, Overvoltage Cat. I (Voltage derived from limited energy transformers). Measurement Considerations Before attempting to use the 80K-
18、6, the following paragraphs should be read and understood. Particular attention should be given to Operator Safety. Operator Safety Indicates the operator must refer to an explanation in this manual. Indicates terminals at which lethal voltages may exist. P/N 613224 April 1981 Rev.6, 11/01 1981-2001
19、 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 WWarning To avoid damage or electric shock, use within ratings and under dry (no condensation) conditions. To avoid electrical shock, the 80K-6 user shou
20、ld be familiar with, and exercise, all possible high- voltage safety practices. When handling the probe the following additional cautions must be taken: When making a measurement, never make body contact with the probe tip or the red portion of the probe. Always hold the probe by its black handle. B
21、efore making a measurement, make sure that the ground side of the output connector is connected to the voltmeters common terminal. The clip lead must be attached to earth ground. Do not use the probe if it is damaged. Before using the probe, inspect it for cracks or missing plastic. Pay close attent
22、ion to the probe body, tip and insulation surrounding the connectors. Make sure the tip is firmly attached to the body. When servicing, use only specified replacement parts. Do not operate probe around explosive gas, vapor or dust. Do not use the probe if it operates abnormally. Protection may be im
23、paired. When in doubt have the probe serviced. Voltmeter Compatibility The 80K-6 is compatible with any ac or dc voltmeter that has an input impedance of 10 megohms 10%. Voltmeters with higher input impedances require the use of an external shunt to obtain an accurate measurement. Use the following
24、formula to determine the value of an external shunt resistor: Rs Rm x 10 Rm10 = Where: Rs= Shunt Resistance in Megohms Rm= Voltmeter Input Impedance in Megohms ( 10 megohms) Circuit Loading The 80K-6 represents a load of approximately 75 megohms to the circuit being measured (13 A per 1 kV). See Fig
25、ure 1. To more accurately determine the load resistance, measure the resistance of the 80K-6 at its voltmeter connector and multiply the measured value by 1000. Probe Tip (Dual Banana Plug) Ground Lead 75M 75k 10M 80K-6 (voltmeter) HI LO 80k6.eps Figure 1. 80K-6 Schematic Diagram Operation Use the f
26、ollowing procedure to operate the 80K-6: 1. Inspect the probe. (Refer to the preceding Warning.) 2. Connect the probe cable to a compatible voltmeter. (Refer to the preceding Warning.) Use a shunt resistor if required. 3. Select an appropriate voltage range (1 volt reading per 1000 volt probe input)
27、. 4. Connect the ground lead of the probe to earth ground. (Refer to the preceding Warning.) 5. Hold the probe by the black handle and connect the probe tip to the circuit being measured. Observe the voltmeter reading. Maintenance Performance Test Verify the probe accuracy by measuring a 5 kV dc 0.2
28、5% voltage source. When used with a compatible dc voltmeter, the probe should accurately measure the source voltage to within 1%. No calibration adjustments are provided on the probe. Cleaning Use a soft cloth dampened with distilled water to clean the 80K-6. Never use solvents or abrasive cleaners.
29、 Disassembly Use the following procedure to disassemble the probe: 1. Unscrew the black handle from the probe, and slide the handle onto the cable. 2. Unscrew the metal tip one turn, and push in on the tip until the internal assembly snaps free of the housing. Remove the tip. 3. Withdraw the interna
30、l assembly from the probe by pulling the metal ring over the threads on the probe body. 4. Logically reverse this procedure to reassemble the probe. Caution To avoid probe damage after repair, measure and, if necessary, adjust the outside distance between the ring and the probe coupling on the inter
31、nal assembly before reassembling the probe. The distance must be 4.9100.020 inches. Make the adjustment using the solder connection at the rear of the metal probe coupling. Replaceable Parts DescriptionFluke Part No. Tip, High Voltage536946 Body, Probe580621 Coupling, Probe536961 Divider, High Volta
32、ge632091 Sleeve580605 Bus Wire, Probe617290 Instruction Sheet613224 Decal587063 Alligator Clip306753 Boot217703 To order replacement parts, call 1-800-526-4731. Outside U.S.A. contact your nearest service center. For application and operation assistance or information on Fluke products, visit the Fluke Website at: or call: USA: 1-888-99 FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Sinapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 RadioFans.CN 收音机爱 好者资料库