1、M S1-1052 10kV Digital Insulation Tester USER MANUAL RadioFans.CN 收音机爱 好者资料库 GSAFETY WARNINGS Safety Warning must be observed during use. IThe circuit under test must be switched off, de-energised, isolated and checked to be safe before insulation test connections are made. Make sure the circuit is
2、not re-energised whilst the instrument is connected. ICircuit connections must not be touched during an insulation test. IAfter completing a test, capacitive circuits must be completely discharged before disconnecting the test leads. Capacitive charges can be lethal. ITested items should be firmly s
3、horted out with a shorting link, after discharge, until required for use. This is to guard against any stored dielectric absorption charge subsequently being released thereby raising the voltage to potentially dangerous levels. IThe voltage indicator and automatic discharge features should be regard
4、ed as additional safety features and not a substitute for normal safe working practice. IIt is rare, but in certain circumstances, breakdown of the circuit under test may cause the instrument to terminate the test in an uncontrolled manner, possibly causing a loss of display while the circuit remain
5、s energised. In this event, the unit must be turned off and the circuit discharged manually. ITest leads, including crocodile clips, must be in good order, clean and with no broken or cracked insulation. IThe instrument should not be used if any part of it is damaged. IWater on the surface of the in
6、strument may lead to erroneous results. It is recommended that the instrument is not used in wet weather conditions. IThis instrument is not intrinsically safe and must not be used in hazardous atmospheres. 2 NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. RadioFans.
7、CN 收音机爱 好者资料库 Breakdown mode / burn mode15 Alarm limit mode15 Step voltage SV test16 Polarisation index PI test, and dielectric absorption ratio DAR test16 Dielectric discharge DD test16 Measurements above 100 G17 Circuit block diagram18 Specifications19 Accessories20 Repair and warranty21 Symbols u
8、sed on the instrument are: FCaution: risk of electric shock GCaution: refer to accompanying notes tEquipment protected throughout by Double Insulation (Class II) cEquipment complies with current EU directives. Terms used in this manual The word must is used to indicate that the instructions followin
9、g should be followed under all circumstances. Failure to follow these instructions could result in damage to the instrument and / or a hazard to the operator. The word should is used to indicate that the instructions indicate best practice. CONTENTS 3 Safety warnings2 Introduction4 General descripti
10、on4 Features4 Cleaning4 Power lead and battery charging5 Instrument controls and indicators6 Power On/Off button7 Test voltage L and M buttons7 Test start / stop button7 /I button7 Bbutton10 Fn button10 Mode button10 Timer L and M buttons11 Record button11 Recording to on-board memory11 Downloading
11、results11 Deleting results11 Recording to a PC12 High voltage warning LED12 Line input present LED12 Test terminals12 Guard terminal12 RS232 / USB connections13 Battery bar graph13 Voltage at terminals13 Timer indicator13 Digital display13 Analogue display14 Secondary display14 Pre-Test / during tes
12、t key action table14 Test mode summary15 Insulation resistance IR test15 RadioFans.CN 收音机爱 好者资料库 Features IContinuous resistance or current measurement on main display. IOptional resistance, current, or figure of merit measurements on secondary display. IStandard test voltages 250 V , 500 V , 1 kV ,
13、 2.5 kV , 5 kV , 10 kV INon-standard test voltages selectable between 50 V and 1 kV in 10 V steps, and selectable between 1 kV and 10 kV in 25 V steps. ITest modes include insulation resistance, polarisation index, step voltage and dielectric discharge. IThe dielectric absorption ratio is automatica
14、lly calculated if the corresponding timers are set. IEither burn or breakdown selectable in insulation resistance mode. IInsulation Alarm Limit available in insulation resistance mode. IProgrammable timers include a main test duration timer, plus T1 and T2 timers for time resistance method type test
15、s. ILoad capacitance and time constant measurements displayed at end of test. ILCD backlight. IData storage, data retrieval and real time data output. IUSB or RS232 communications. IBattery level / charge level indicator. Cleaning Disconnect the instrument and wipe it with a clean cloth slightly dam
16、ped with soapy water or Isopropyl alcohol (IPA). INTRODUCTION 4 General Description The S1-1052 is a microprocessor controlled 10 kV insulation tester offering measurement capability of up to 35 T. The instrument performs automatic tests and has data storage and data retrieval facilities. Polarizati
17、on index and dielectric discharge tests are performed automatically, and test duration and voltages can be adjusted according to user preference for these tests. A step voltage test can be performed automatically, with a default voltage of 1 kV , and test duration of 5 minutes. It is possible for th
18、e user to set different voltages and test durations for the step voltage test. The S1-1052 can be powered from the mains supply or by its own internal rechargeable battery, which provides for at least 4 hours of continuous testing. A battery level indicator on the LCD display indicates battery capac
19、ity. Connecting power to the mains supply connector will automatically charge the battery whether the instrument is switched on or off, except during testing. A high level of internal isolation allows the instrument to be used while the unit is supplied from the mains. An internal battery management
20、 system switches the instrument off after ten minutes of inactivity. If the battery approaches a very low state of charge the instrument turns itself off, and mains power must be applied before the instrument can be used again. Recorded test results and settings will not be lost when the instrument
21、is switched off. A comprehensive LCD display shows resistance, current, capacitance, time constant, voltage, timer measurements, and figure of merit measurements such as polarisation index. POWER LEAD AND BATTERY CHARGING 5 If the power lead supplied is not suitable for your mains connection, do not
22、 use an adaptor. Always use a power lead fitted with the correct plug. The instrument is fitted with a two-pin IEC60320 power inlet. Most power leads are made with three-core cable, so the ground connection will not be used. Power lead connection table ConnectionUK/InternationalUSA Earth/Ground Yell
23、ow/Green Green Neutral Blue White Live (Line) Brown Black If using a fused plug, ensure that it is fitted with a 3 Amp fuse. The instrument can be powered from 85 - 265 Vrms a.c. at 50/60 Hz. The battery will charge as long as the mains supply is connected, except when a test is in progress. The pow
24、er On/Off button has a green LED, which illuminates when mains power is present. For optimum battery life, charge the battery after each use. A completely exhausted battery will take 14 hours to recharge. INSTRUMENT CONTROLS AND INDICATORS 6 Measurement terminals -veguard+ve RS232USBLine input Time
25、L and M selectors Voltage L M selectors Line input Present indicator Power on/off /I selector Burn indicatorAlarmData recordingTest modes Voltage at terminals Battery level Breakdown indicator High voltage warning indicator High voltage warning indicator Analogue display Timers Main, T1, T2 Dielectr
26、ic absorption ratio/T1 timer Digital display Polarisation index/timer T2 Time constant Capacitance display Data record High voltage warning indicator Test start/stopModeFunction Backlight on/off 7 When the test has stopped, the display continues to show the voltage present on the test leads. Pressin
27、g either test voltage L or M button will then display the test voltage immediately before the end of the test. Test start / stop button A test will only start if this button is pressed, held and then released as soon as the red high voltage warning indicator LED lights. The LCD and red LED high volt
28、age warning indicators flash when the test starts. A test will not start if the button is released before the red LED shows, or pressed continuously for longer than 5 seconds. This is a safety feature to prevent a test being started inadvertently. The presence of a voltage greater than 50 V on the t
29、est leads is indicated with flashing high voltage warning indicators. Testing is disabled if the external voltage exceeds 80 V . Testing will stop if the test start / stop button is pressed again, the preset test time is reached, or, if the unit is not in burn mode, insulation breakdown is detected.
30、 When a test has finished, the instrument will discharge the load, which may take some time. The operator must always check the load has been discharged before touching the test leads. /I button Pressing this button toggles the digital and secondary displays. The details available when scrolling the
31、 secondary display will depend upon the test mode selected, whether timers T1 and T2 have been set, and the duration of the test. The digital display toggles between insulation resistance and current. Power On/Off button The instrument will only turn on if this button is pressed, held and then relea
32、sed when the display responds. The instrument will not turn on if the button is released before the display responds, or if the button is held down for too long. This is a safety feature to prevent the instrument being inadvertently turned on. The instrument is turned off either by pressing the butt
33、on again, or if the instrument is running on the battery, by timing out after 10 minutes of inactivity. Upon switching the instrument on the display will first show Ini while it undergoes a self-checking routine. When Ini disappears, the instrument is ready for use. Test voltage L and M buttons Usin
34、g these buttons one of six test voltages can be selected: 250 V , 500 V , 1 kV , 2.5 kV , 5 kV , and 10 kV . The selected voltage is shown on the display. A non standard voltage between 50 V and 10 kV can be selected by holding down the Fn function button whilst operating the buttons. The selectable
35、 voltage is adjustable in 10 V steps between 50 V and 1 kV , and adjustable in 25 V steps between 1 kV and 10 kV . An auto repeat facility is enabled when the button is held down, allowing faster travel through the range. If there is an external voltage greater than 50 V on the test leads, the high
36、voltage warning indicators are flashed, and the display shows this voltage instead. The instrument will not perform a test if this voltage is greater than 80 V . During a test the display shows the actual voltage on the test leads. If the test voltage is changed during a test, the new test voltage w
37、ill be displayed briefly. 8 Figure 1 - The result of an insulation resistance test (IR) Test settings: T1 and T2 times set in order to measure the DAR Test conditions: test runs for longer than 10 minutes as this is required for a PI reading. In the IR, PI, and DD modes the secondary display initial
38、ly shows the PI (polarity index), DAR (dielectric absorption ratio), and, on completion of the test, the TC (time constant) and capacitance measurements. Toggling the display shows insulation resistances and currents. In the SV (step voltage) mode, the secondary display toggles between insulation re
39、sistances and currents. Figures 1 to 4 illustrate the display status for sequential /I key presses in the different modes. Press /I Press /I Press /I Figure 2 - The result of a polarity index test (PI) Test settings: T1 and T2 times set in order to measure the DAR Test conditions: timer defaults to
40、10 minutes, as this is required for a PI reading Figure 3 - The result of a dielectric discharge test (DD) Test settings: T1 and T2 times set in order to measure the DAR Test condition: test runs for longer than 10 minutes as this is required for a PI reading. 9 Press /I Press /I Press /I Press /I P
41、ress /I Press /I 10 test voltage buttons. TL and TM represent the timer buttons. Button PressFunctionComment Fn + VLIncrement in 10 V stepsVoltage range between 50 and 1000 V Fn + VMDecrement in 10 V stepsVoltage range between 50 and 1000 V Fn + VLIncrement in 25 V stepsVoltage range between 1000 an
42、d 10000 V Fn + VMDecrement in 25 V stepsVoltage range between 1000 and 10000 V Fn + TL or TMCycle through timersSelect main timer, T1 or T2 Fn + ModeCycle through IR modesSelect breakdown, burn or alarm limit for IR test Fn +Download mode See Downloading results Fn + Clear stored dataThe record butt
43、on is pressed twice Mode button Press down on the MODE button to cycle through and select the test mode. Modes of test to be chosen from include an insulation resistance IR test, a polarization index PI test, a step voltage SV test, and a dielectric discharge DD test. The insulation resistance IR te
44、st operates in either burn or breakdown mode, with an option of setting a resistance threshold alarm limit. These modes can be cycled through and selected by holding down the Fn button and pressing the Mode button. Figure 4 - The result of a step voltage (SV) test Test condition: timer defaults to 5
45、 minutes and test voltage defaults to 1000 V . Bbutton Pressing this button toggles the display backlight on and off. Fn button The function button when pressed allows other keys to perform another function. This is summarized in the table below. VL and VM represent the Press /I Press /I 11 Recordin
46、g to on-board memory Press the record button to start and stop recording. When data recording is enabled the record symbol will flash on and off repeatedly. Recording can only be activated before testing commences. Results are stored at 15, 30, 45 and 60 seconds. After 60 seconds, at minute interval
47、s up to 10 minutes. After 10 minutes, results are recorded at 5 minute intervals until the test terminates. At each interval the recorded data will contain selected voltage, test time elapsed, voltage applied, leakage current, and insulation resistance. Downloading results Connect the instrument to
48、the RS232 / USB port of a PC running Download Manager. Refer to section RS232 / USB connection for set up details. Start Download Manager on your PC, select the S1-1052 driver and right click the icon. Select Download. Switch the instrument on and wait until initialisation is complete. Press the fun
49、ction key along with the record key. The instrument now displays dld to indicate download mode. Press and hold down test button until download begins, shown by analogue display lighting clockwise. Results will not be erased during this operation and so may be downloaded repeatedly. Deleting test results Switch the instrument on and wait until initialisation is complete. Hold down the function key and press the record key twice. The instrument now displays clr to indicate clear mode. Press and hold down test button until clear process begins, shown by analogue display