1、INSTRUCTION MANUAL MODEL 5105B 150MHz (1GS/s) Analog/Digital Oscilloscope RadioFans.CN 收音机爱 好者资料库 2 Subject to change without notice TEST INSTRUMENT SAFETY TEST INSTRUMENT SAFETY WARNING Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing must
2、 some- times be performed where exposed high voltage is present. An electrical shock causing 10 milliamps of current to pass through the heart will stop most human heartbeats.Voltage as low as 35 volts dc or ac rms should be considered dangerous and hazardous since it can produce a lethal current un
3、der certain conditions. Higher voltages are even more dangerous. Your normal work habits should include all accepted practices to prevent contact with exposed high voltage, and to steer cur- rent away from your heart in case of accidental contact with a high voltage. Observe the following safety pre
4、cautions: 1 There is little danger of electrical shock from the dc output of this power supply. However, there are several other possible test conditions using this power supply that can create a high voltage shock hazard: 1a If the equipment under test is the hot chassis“ type, a seri- ous shock ha
5、zard exists unless the equipment is unplugged (just turning off the equipment does not remove the hazard), or an isolation transformer is used. 1b If the equipment under test is powered up“ (and that equipment uses high voltage in any of its circuits), the power supply outputs may be fl oated to the
6、 potential at the point of connection. Remember that high voltage may appear at unexpected points in defective equipment. Do not fl oat the power supply output to more than 100 volts peak with respect to chassis or earth ground. 1c If the equipment under test is off“ (and that equipment uses high vo
7、ltage in any of its circuits under normal operation), discharge high-voltage capacitors before making connec- tions or tests. Some circuits retain high voltage long after the equipment is turned off. 2 Use only a polarized 3-wire ac outlet.This assures that the power supply chassis, case, and ground
8、 terminal are con- nected to a good earth ground and reduces danger from electrical shock. 3 Dont expose high voltage needlessly. Remove housings and covers only when necessary.Turn off equipment while making test connections in high-voltage circuits. Discharge high-voltage capacitors after removing
9、 power. 4 If possible, familiarize yourself with the equipment being tested and the location of its high voltage points. However, remember that high voltage may appear at unexpected points in defective equipment. 5 Use an insulated fl oor material or a large, insulated fl oor mat to stand on, and an
10、 insulated work surface on which to place equipment; and make certain such surfaces are not damp or wet. 6 When testing ac powered equipment, the ac line voltage is usually present on some power input circuits such as the on-off switch, fuses, power transformer, etc. any time“ the equipment is conne
11、cted to an ac outlet. 7 B+K Precision products are not authorized for use in any application involving direct contact between our product and the human body, or for use as a critical component in a life support device or system. Here, direct contact“ refers to any connection from or to our equipment
12、 via any cabling or switching means.A critical component“ is any component of a life support device or system whose failure to perform can be reasonably expected to cause failure of that device or system, or to affect its safety or effectiveness. 8 Never work alone. Someone should be nearby to rende
13、r aid if necessary. Training in CPR (cardio-pulmonary resuscitation) fi rst aid is highly recommended. RadioFans.CN 收音机爱 好者资料库 3 Subject to change without notice Contents 150 MHz Analog-/Digital Oscilloscope 4 Specifi cations 5 Important hints 6 List of symbols used: 6 Positioning the instrument 6 S
14、afety 6 Proper operation 6 CAT I 6 Environment of use. 6 Environmental conditions 7 Maintenance 7 Line voltage 7 Description of the controls 8 Basic signal measurement 10 Signals which can be measured 10 Amplitude of signals 10 Values of a sine wave signal 10 DC and AC components of an input signal
15、11 Timing relationships 11 Connection of signals 11 First time operation and initial adjustments 12 Trace rotation TR 12 Probe adjustment and use 12 1 kHz adjustment 12 1 MHz adjustment 13 Operating modes of the vertical amplifi er 13 XY operation 14 Phase measurements with Lissajous fi gures 14 Mea
16、surement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering (MODE menu) 15 Normal trigger mode (See menu MODE) 16 Slope selection (Menu FILTER) 16 Trigger coupling (Menu: FILTER) 16 Video (tv triggering) 16 Fra
17、me sync pulse triggering 17 Line sync pulse triggering 17 LINE trigger 17 Alternate trigger 17 External triggering 17 Indication of triggered operation (TRIGD LED) 17 Hold-off time adjustment 17 Time base B (2nd time base). Delaying, Delayed Sweep. Analog mode. 18 Alternate sweep 18 AUTOSET 19 Compo
18、nent tester 19 CombiScope 21 DSO Operation 22 DSO operating modes 22 Memory resolution 22 Memory depth 23 Horizontal resolution with X magnifi er 23 Maximum signal frequency in DSO mode 23 Display of aliases 23 Vertical amplifi er operating modes 23 Data transfer 23 RS-232 Interface, Remote control
19、24 Selection of Baud rate 24 Data transmission 24 General information concerning MENU 25 Controls and Readout 26 RadioFans.CN 收音机爱 好者资料库 4 Subject to change without notice 5105B 1 GSa/s Real Time Sampling, 10 GSa/s Random Sampling 8-Bit Low Noise Flash A/D Converters Pre-/Post-Trigger -100% to +400%
20、 Time Base 50 s/cm 5 ns/cm 1 MPts memory per channel allows zoom up to 40,000:1 Acquisition modes: Single Event, Refresh, Average, Envelope, Roll, Peak-Detect RS-232 Interface Signal display: Yt and XY; Interpolation: Sinx/x, Pulse, Dot Join (linear) 1 5 0 M H z ( 1 G G / s ) A n a l o g - / D i g i
21、 t a l O s c i l l o s c o p e Cursor measurement choices in digital mode Digital Mode: TV field and zoomed display of one selected line 5 Subject to change without notice Important hints 150 MHz Analog/Digital Oscilloscope 5105B Technical description Vertical Deflection Channels: Analog: 2 Digital:
22、2 Operating Modes: Analog:CH 1 or CH 2 separate, DUAL (CH 1 and CH 2 alternate or chopped), Addition Digital:CH 1 or CH 2 separate, DUAL (CH 1 and CH 2), Addition Y in XY-Mode:CH 1 Invert:CH 1, CH 2 Bandwidth (-3dB):2 x 0 - 150MHz Rise time: 3.5ns Overshoot:max. 1% Deflection Coefficient(CH 1, 2):14
23、 calibrated steps 1mV 2mV/cm (10MHz) 5% (0 - 10MHz (-3dB) 5mV 20V/cm3% (1-2-5 sequence) variable (uncalibrated): 2.5:1 to 50V/cm Inputs CH 1, 2: Impedance:1M / 15pF Coupling:DC, AC, GND (ground) Max. Input Voltage:400V (DC + peak AC) Y Delay Line (analog):70ns Measuring Circuits:Measuring Category I
24、 Analog mode only: Auxiliary input: Function (selectable): Extern Trigger, Z (unblank) Coupling:AC, DC Max. input voltage:100V DC +peak AC Triggering Analog and Digital Mode Automatic (Peak to Peak): Min. signal height:5mm Frequency range:10Hz - 250MHz Level control range:from Peak- to Peak+ Normal
25、(without peak):Slope/Video Min. signal height:5mm Frequency range:0 - 250MHz Level control range: 10cm to +10cm Operating modes:Slope/Video Slope:positive, negative, both Sources:CH 1, CH 2, alt.1/2 ( 8mm), Line, Ext. Coupling:AC:(10Hz-250MHz) DC:(0-250MHz) HF:(30kHz 250MHz) LF:(0-5kHz) Noise Rej. s
26、witchable Video:pos./neg. Sync. Impulse Standards:525 Line/60Hz Systems 625 Line/50Hz Systems Field:even/odd/both Line:all/line number selectable Source:CH 1, CH 2, Ext. Indicator for trigger action:LED External Trigger via:Auxiliary Input (0.3Vpp, 150MHz) Coupling:AC, DC Max. input voltage:100V DC
27、+peak AC Digital mode Pre/Post Trigger:-100% to +400% related to complete memory Analog mode 2nd Trigger Min. signal height:5mm Frequency range:0 - 250MHz Coupling:DC Level control range: 10cm to +10cm Horizontal Deflection Analog mode Operating modes:A, ALT (alternating A/B), B Time base A (Sequenc
28、e): 0.5s/cm - 50ns/cm (1-2-5 sequence) Time base B (Sequence): 20ms/cm 50ns/cm (1-2-5 sequence) Accuracy A and B: 3% X-Mag. x10:to 5 ns/cm Accuracy X x10:5% Variable time base A/B: cont. 1:2.5 Hold Off time:var. 1:10 LED-Indication Bandwidth X-Amplifier: 0 - 3MHz (-3dB) X-Y phase shift 3: 220kHz Dig
29、ital mode Time base range (sequence) Refresh Mode:20ms/cm - 5ns/cm (1-2-5 sequence) with Peak Detect:20ms/cm 50ns/cm (1-2-5 sequence) Roll Mode:50s/cm 50ms/cm (1-2-5 sequence) Accuracy time base Time base:50ppm Display:1% MEMORY ZOOM: max. 40,000:1 Bandwidth X-Amplifier:0 - 150MHz (-3dB) X-Y phase s
30、hift 3: 100MHz Digital Storage Acquisition (real time):2x 500MSa/s, 1GSa/s interleaved Acquisition (random sampling):10GSa/s Bandwidth:2 x 0 - 100MHz (random) Memory:1M-Samples per channel Operating modes:Refresh, Average, Envelope/ Roll: Free Run/Triggered, Peak-Detect Resolution (vertical):8 Bit (
31、25Pts/cm) Resolution (horizontal): Yt:11 Bit (200Pts/cm) XY:8 Bit (25 Pts /cm) Interpolation:Sinx/x, Dot Join (linear) Delay:1 Million * 1/Sampling Rate to 4 Million * 1/Sampling Rate Display refresh rate:max.170/s at 1MPts Display:Yt, XY (acquired points only), Interpolation, Dot Join Reference Mem
32、ories:9 with 2kPts each (for recorded signals) Display:2 signals of 9 (free selectable) Operation/Measuring/Interfaces Operation:Menu (multilingual), Autoset, help functions (multilingual) Save/Recall (instrument parameter settings):9 Signal display:max. 4 traces analog:CH 1, 2 (Time Base A) in comb
33、ination with CH 1, 2 (Time Base B) digital:CH1,2 and ZOOM or Reference or Mathematics) Frequency counter: 6 digit resolution: 1MHz 200MHz 5 digit resolution:0.5Hz 1MHz Accuracy:50ppm Auto Measurements: Analog mode:Frequency/Period/Vdc/Vpp/Vp+/Vp- add. in digital mode:Vrms/Vavg Cursor Measurements: A
34、nalog mode:V, t, 1/t (f), V to GND, ratio X, ratio Y add. in digital mode:Pulse count, Vt to Trigger, Peak to Peak, Peak+, Peak- Resolution Readout/Cursor: 1000 x 2000Pts, Signals: 250 x 2000 Interfaces (plug-in):RS-232 B Mathematic functions Number of Formula Sets:5 with 5 formulas each Sources:CH
35、1, CH 2, Math 1-Math 5 Targets:5 math. memories, Math 1-5 Functions:ADD, SUB, 1/X, ABS, MUL, DIV, SQ, POS, NEG, INV Display:max. 2 math. memories (Math 1-5) Display CRT:D14-375GH Display area (with graticule): 8cm x 10cm Acceleration voltage:approx. 14kV General Information Component tester Test vol
36、tage:approx. 7Vrms(open circuit), approx. 50Hz Test current:max. 7mArms(short circuit) Reference Potential :Ground (safety earth) Probe ADJ Output:1kHz/1MHz square wave signal 0.2Vpp (tr 4ns) Trace rotation:electronic Line voltage:105 253V, 50/60Hz 10%, CAT II Power consumption:42 Watt at 230V, 50Hz
37、 Protective system:Safety class I (EN61010-1) Weight:5.6kg Cabinet (W x H x D):285 x 125 x 380mm Ambient temperature:0C .+40C Accessories supplied: Line cord, Operating manual, 2 Probes 10:1 with attenuation ID, Windows Software for control and data transfer 150MHz (1GS/s) Analog/Digital Oscilloscop
38、e 5150B Specifi cations 6 Subject to change without notice connecting any signals. It is prohibited to separate the safety ground connection. Most electron tubes generate X rays; the ion dose rate of this in- strument remains well below the 36 pA/kg permitted by law. In case safe operation may not b
39、e guaranteed do not use the instrument any more and lock it away in a secure place. Safe operation may be endangered if any of the following was noticed: in case of visible damage. in case loose parts were noticed if it does not function any more. after prolonged storage under unfavourable condition
40、s (e.g. like in the open or in moist atmosphere). after any improper transport (e.g. insuffi cient packing not conforming to the minimum standards of post, rail or trans- port fi rm) Proper operation Please note: This instrument is only destined for use by person- nel well instructed and familiar wi
41、th the dangers of electrical measurements. For safety reasons the oscilloscope may only be operated from mains outlets with safety ground connector. It is prohibited to separate the safety ground connection. The plug must be inserted prior to connecting any signals. CAT I This oscilloscope is destin
42、ed for measurements in circuits not connected to the mains or only indirectly. Direct measurements, i.e. with a galvanic connection to circuits corresponding to the categories II, III, or IV are prohibited! The measuring circuits are considered not connected to the mains if a suitable isolation tran
43、sformer fulfi lling safety class II is used. Measurements on the mains are also possible if suitable probes like current probes are used which fulfi l the safety class II. The measurement category of such probes must be checked and observed. Measurement categories The measurement categories were der
44、ived corresponding to the distance from the power station and the transients to be expected hence. Transients are short, very fast voltage or cur- rent excursions which may be periodic or not. Measurement cat. IV: Measurements close to the power station, e.g. on electricity meters Measurement cat. I
45、II: Measurements in the interior of buildings (power distribution installations, mains outlets, motors which are permanently installed). Measurement cat. II: Measurements in circuits directly connected to the mains (household appliances, power tools etc). Environment of use The oscilloscope is desti
46、ned for operation in industrial, business, manufacturing, and living sites. Important hints Please check the instrument for mechanical damage or loose parts immediately after unpacking. In case of damage we advise to contact the sender. Do not operate. List of symbols used: Consult the manual High v
47、oltage Important note Ground Positioning the instrument For selection of the optimum position in use the instrument may be set up in three different positions (see pictures C,D,E). The handle will remain locked in the carrying position if the instrument is positioned on its rear feet. Move the handl
48、e to the instrument top if the horizontal operating position is preferred (See picture C). If a position corresponding the picture D (10 degrees inclination) is desired move the handle from the carrying position A towards the bottom until it engages and locks. In order to reach a position with still
49、 greater incli- nation (E shows 20 degrees) unlock the handle by pulling and move it further into the next locking position. For carrying the instrument in the horizontal position the handle can be locked horizontally by moving it upwards as shown in picture B. The instrument must be lifted while doing this, otherwise the handle will unlock again. Safety The instrument fulfi ls the VDE 0411 part 1 regulations for electrical measuring, control and laboratory instruments and was manufactured and tested accordingly.