1、PN 686953 March 1998 Rev.1, 9/01 1998-2001 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies. T5-600/T5-1000 Electrical Tester Service Information Introduction This service information sheet provides the following information for the
2、 T5-600 and T5-1000 Electrical Testers (hereafter referred to as the tester). Safety information Parts and service information Specifications Cleaning procedure Required equipment Performance tests Parts and accessories list Battery replacement procedure For operating instructions, refer to the T5-6
3、00/T5-1000 Electrical Tester Instruction Sheet. Safety Information Warning To avoid possible electric shock or personal injury, follow these guidelines: Do not use the tester if it is damaged or operating abnormally. Protection may be impaired. Before each use: Make sure the battery door is closed a
4、nd latched. Inspect the tester and test leads. Look for cracks, missing plastic, exposed metal, or damaged insulation. Replace damaged test leads before using the tester. Verify the testers operation by measuring a known voltage. Replace the batteries as soon as the low battery indicator (?) appears
5、. Do not use the tester around explosive gas, vapor or dust. Do not apply more than the rated voltage, as marked on the tester, between terminals or between any terminal and earth ground. Refer servicing to qualified personnel. Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc. Wh
6、en using the probes, keep your fingers behind the finger guards on the probes. Connect the common test lead before you connect the live test lead. Disconnect the live test lead first. The display will not display hazardous voltages in display hold. The voltage indicator LED () continues to operate.
7、99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentD RadioFans.CN 收音机爱 好者资料库 T5-600/T5-1000 Electrical Tester 2 Parts and Service The tester is warranted to be free from defects in material and workmanship for two years, while under normal
8、 use. Parts and repairs are warranted for 90 days. For the complete warranty statement, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet. To order parts, or for warranty service, contact Fluke as follows: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Eur
9、ope: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 Or, visit Flukes Web site at . Specifications Accuracy is specified for one year after calibration, at 18 C to 28 C (64 F to 82 F) with relative humidity to 90 %. AC conversions are ac-coupled,
10、 average responding, and calibrated to the rms value of a sine wave input. Accuracy specifications are given as follows: ( % of reading + number of least significant digits) Temperature coefficient of 0.1 x (specified accuracy)/ C for 28 C (82.4 F) Serial Numbers 79560000 and Above General Specifica
11、tions CalibrationOne-year calibration cycle. Maximum Voltage Between any Terminal and Earth Ground T5-600: 600 V rms, Overvoltage Category III, Pollution Degree 2 T5-1000: 1000 V rms, Overvoltage Category III, 600 V RMS, OverVoltage Category IV, Pollution Degree 2 Maximum Voltage Between Current For
12、k and Earth Ground T5-600 and T5-1000: 1000 V rms, Overvoltage Catergory III, also 600 V rms, Overvoltage Category IV, Pollution Degree 2, (This dual rating applies to the current fork only.) TemperatureOperating: -10 C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F) AltitudeOper
13、ating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft) Relative Humidity0 % to 95 %, 5 C to 30 C (41 F to 86F); 0 % to 75 %, 30 C to 40 C (86 F to 104 F); 0 % to 45 %, 40 C to 50 C (104 F to 122 F) Battery Type and LifeAA (2); 360 hours continuous with alkaline; 125 hours continuous with zinc chlorid
14、e Shock, Vibration1 m drop at -10 C to + 50 C (14 F to 122 F) per ANSI/ISA-S82.01-1994 and EN 61010-1 1995. Random vibration per MIL-PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g maximum) Surge ProtectionT5-600: 6 kV per IEC 61010, T5-1000: 8 kV per IEC 61010 Enclosure RatingIP 52 per IEC
15、60529, no vacuum applied RadioFans.CN 收音机爱 好者资料库 Test Equipment Depot? 99 Washington Street? Melrose, MA 02176-6024? ? ? 800-517-8431? 781-665-0780 FAX Service Information Serial Numbers 79559999 and Below 3 General Specifications (continued) RF Field Specification0.5 % full scale + (specified accur
16、acy) at 3 V/m SafetyComplies with ANSI/ISA-S82.01-94, UL Classified to IEC 61010, CSA/CAN C22.2 No.1010.1- 92, and EN61010-1 1995. EMCEN 61326 Certifications In accordance with IEC 61010-1 54CJ Resolution and Accuracy FunctionT5-600 RangeT5-1000 RangeResolutionAccuracy ?600 V rms1000 V rms1 V(1.5 %
17、+ 2 digits) ?600 V1000 V1 V(1 % + 1 digit) ?100.0 A100.0 A0.1 A(3 % + 3 digits) ?1000 1000 1 (1 % + 2 digits) Input Characteristics Input Protection FunctionT5-600T5-1000 Input impedance (nomial ? 600 V rms1000 V rms1 M, 100 pF ac-coupled ? 600 V rms1000 V rms1 M, 100 pF ? 600 V rms1000 V rms Open C
18、ircuit Test VoltageShort Circuit Current ?1.65 V dc (nominal)600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in large buildings. CAT IV protects against transients from a primary supply such as an electricity meter or an
19、overhead or underground utility service. Serial Numbers 79559999 and Below General Specifications CalibrationOne-year calibration cycle. Maximum Voltage Between any Terminal and Earth Ground T5-600: 600 V rms, Overvoltage Category III T5-1000: 1000 V rms, Overvoltage Category III TemperatureOperatin
20、g: -10 C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F) AltitudeOperating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft) Relative Humidity0 % to 95 %, 5 C to 30 C (41 F to 86F); 0 % to 75 %, 30 C to 40 C (86 F to 104 F); 0 % to 45 %, 40 C to 50 C (104 F to 122 F) Battery Type a
21、nd LifeAA (2); 400 hours continuous with alkaline; 200 hours continuous with zinc chloride RadioFans.CN 收音机爱 好者资料库 T5-600/T5-1000 Electrical Tester 4 General Specifications (continued) Shock, Vibration1 m drop at 15 C to 35 C (59 F to 95 F) per ANSI/ISA-S82.01-1994 and EN 61010-1 1993. Sinusoidal vi
22、bration per MIL-PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g maximum) Surge ProtectionT5-600: 6 kV per IEC 1010-1, 1990-09: T5-1000: 8 kV per IEC 1010-1, 1990-09 Enclosure RatingIP 52 per IEC 60529, no vacuum applied RF Field Specification0.5 % full scale + (specified accuracy) at 3 V/m S
23、afetyComplies with ANSI/ISA-S82.01-94 for use in overvoltage category III (CAT III) locations, UL3111, CSA/CAN C22.2 No.1010.1-92, and EN61010-1 1993. EMCEN 50081-1, EN 50082-1 Certifications Resolution and Accuracy FunctionT5-600 RangeT5-1000 RangeResolutionAccuracy ?600 V rms1000 V rms1 V(1.5 % +
24、2 digits) ?600 V1000 V1 V(1 % + 1 digit) ?100.0 A100.0 A0.1 A(3 % + 3 digits) ?1000 1000 1 (1 % + 2 digits) Input Characteristics Input Protection FunctionT5-600T5-1000 Input impedance (nomial ? 600 V rms1000 V rms1 M, 100 pF ac-coupled ? 600 V rms1000 V rms1 M, 100 pF ? 600 V rms1000 V rms Open Cir
25、cuit Test VoltageShort Circuit Current ?2.4 V dc (nominal)600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in large buildings. CAT IV protects against transients from a primary supply such as an electricity meter or an ove
26、rhead or underground utility service. Cleaning the Tester ?Warning To avoid electric shock or damage to the tester, never allow water inside the case. To avoid damaging the testers case, never use solvents on the tester. If the tester requires cleaning, wipe it down with a cloth that is lightly damp
27、ened with water or a mild detergent. Do not use aromatic hydrocarbons, chlorinated solvents, or methanol-based fluids when wiping down the tester. Service Information Equipment Required for Performance Tests 5 Equipment Required for Performance Tests The following equipment is required for performan
28、ce tests: Fluke 5500A Multi-Product Calibrator, or equivalent 0 V to 5 V adjustable dc power supply Approximately 8 m (26 ft) of #14 single-conductor magnet wire wound into a butterfly-shaped coil. Figure 1 shows how to make the coil. Performance Tests Use the following procedures to verify the test
29、ers performance. If the tester fails any of the tests, return it to Fluke for calibration or repair. Testing the AC Current Function The tests in this section require the butterfly coil shown in Figure 1. Note that the 10 conductors in the middle of the coil cause the tester to read 10 times more cu
30、rrent than the calibrator supplies. 1. Put one edge of the butterfly coil in a vice to hold the coil during testing. 2. Set the calibrator to the current and frequency given in Step 1 of Table 1. 3. Connect the butterfly coil to the calibrator. 4. Set the tester to the amperage function. 5. Place th
31、e testers current fork around the middle of the coil so that the alignment marks are centered and perpendicular to the wire bundle, as shown in Figure 1. 6. Verify that the tester reads within the display limits shown for Step 1 in Table 1. 7. Apply the currents and verify the testers readings for t
32、he remaining steps in Table 1. Table 1. AC Current Tests StepCalibrator outputT5-600/T5-1000 Display Limits 110 A, 60 Hz96.7 to 103.3 210 A, 45 Hz96.7 to 103.3 30.5 A, 45 Hz4.5 to 5.5 T5-600/T5-1000 Electrical Tester 6 21 cm (8.25) 10 turns of #14 single-conductor magnet wireSeparate into two 5-turn
33、 coils. Use cable ties to hold coils together Connect to Calibrator Mount Banana plugs on wire end qp7f.eps Figure 1. Making the Butterfly Coil Service Information Performance Tests 7 Testing the Voltage Functions The tests in this section check the following voltage functions: Automatic selection o
34、f ac or dc voltage mode Correct operation of the hazardous voltage indicator Correct operation of the HOLD function Perform the tests as follows: 1. Set the calibrator to the T5-600 or T5-1000 test voltage given in Table 2. 2. Apply the voltage to the tester. Verify that the tester reads within the
35、display limits shown. 3. Apply the voltages and verify the testers readings for the steps in Tables 3 and 4. Table 2. AC Voltage Test T5-600T5-1000 Calibrator Output T5-600 Display Limits (AC annunciator ON)Calibrator Output T5-1000 Display Limits (AC annunciator ON) 600 V, 60 Hz589 to 6111000 V, 60
36、 Hz983 to 1017 Table 3. DC Voltage Tests T5-600T5-1000 StepCalibrator Output T5-600 Display Limits (DC annunciator ON)Calibrator Output T5-1000 Display Limits (DC annunciator ON) 1600 V dc593 to 6071000 V dc989 to 1011 2-60 V dc-58 to -62-100 V dc-98 to -102 31.0 V dc1 1*1.0 V dc1 1* 4-1.0 V dc-1 1*
37、-1.0 V dc-1 1* * Verify that the ac annunciator is OFF. Table 4. Hazardous Voltage LED and HOLD Function Tests StepCalibrator OutputT5-600/T5-1000 Display Indicators 112 V, 60 HzReading: 12 1 Hazardous voltage LED OFF 212 V, 60 Hz Tap testers HOLD button Reading: 12 1 Hazardous voltage LED OFF HOLD
38、indicator ON 3+38 V dcReading: 12 1 Hazardous voltage LED ON HOLD indicator ON T5-600/T5-1000 Electrical Tester 8 Resistance and Continuity Function Tests 1. Set the calibrator to the resistance given in Step 1 of Table 5. 2. Apply the resistance to the tester. Verify that the tester reads within th
39、e display limits shown. 3. Apply the resistances and verify the testers responses for the remaining steps in Table 5. Table 5. Resistance and Continuity Tests StepCalibrator OutputT5-600/T5-1000 Display Limits and Beeper Responses 11 k988 to 1012 226 Beeper ON 3300 Beeper OFF 40 -2 to 2, Beeper ON L
40、ow Battery Indicator Test The following procedure verifies correct operation of the low battery indicator. 1. Remove the testers batteries. 2. Set the dc power supply to 3.0 V. Apply this voltage to the testers battery terminals. 3. Set the calibrator to 1 k. Apply this resistance to the testers pro
41、bes. 4. Slowly decrease the dc voltage just until the testers low battery indicator (?) turns on. 5. Verify that the tester reads 1000 12 counts (988 to 1012). 6. Disconnect the calibrator and the dc power supply. Reinstall the testers batteries. Parts and Accessories Tables 6 shows the replacement
42、parts and accessories available from Fluke for the T5-600 and T5-1000 Electrical Testers. Table 6. Replacement Parts and Accessories DescriptionFluke Part Number Test lead assembly Replace only with Fluke double-insulated leads. 648029 Battery door1626588 Battery door screw1618578 AA battery, 1.5 V,
43、 carbon-zinc (2 required)650181 AA battery, 1.5 V, alkaline (2 required)376756 T5-600/T5-1000 Electrical Tester Instruction Sheet (English)1629509 Service Information Battery Replacement 9 Table 6. Replacement Parts and Accessories (continued) DescriptionFluke Part Number T5-600/T5-1000 Electrical T
44、ester Instruction Sheet (International)1621978 H5 Belt HolsterAccessory TP1 Probe Set, Flat BladeAccessory TP4 Probe Set, 4 mm RoundAccessory Battery Replacement Figure 2 shows how to replace the batteries. Observe the polarity markings inside the battery compartment. qp1.eps Figure 2. Replacing the Batteries Back to the Fluke T5 Series Product Page Visit us at www.TestEquipmentD Test Equipment Depot? 99 Washington Street? Melrose, MA 02176-6024? ? ? 800-517-8431? 781-665-0780 FAX