FLUKE 80T-IR 80T-E Instruction 电路图.pdf

上传人:cc518 文档编号:200429 上传时间:2025-03-08 格式:PDF 页数:10 大小:354.60KB
下载 相关 举报
FLUKE 80T-IR 80T-E Instruction 电路图.pdf_第1页
第1页 / 共10页
FLUKE 80T-IR 80T-E Instruction 电路图.pdf_第2页
第2页 / 共10页
FLUKE 80T-IR 80T-E Instruction 电路图.pdf_第3页
第3页 / 共10页
亲,该文档总共10页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

《FLUKE 80T-IR 80T-E Instruction 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 80T-IR 80T-E Instruction 电路图.pdf(10页珍藏版)》请在收音机爱好者资料库上搜索。

1、 80T-IR/E Extended Range Infrared Temperature Probe Instruction Sheet Introduction The Fluke 80T-IR/E Extended Range Infrared Temperature Probe (the probe) is a noncontact temperature measurement accessory for use with a test instrument capable of measuring DC volts in the millivolt range such as a

2、digital multimeter (DMM). The probe has a temperature range of 32F to 1000F, with a basic accuracy of 3% of reading, and an output of 1 mV dc per F. Temperature is measured by pointing the probe at the surface to be measured, and reading the temperature on the test instrument display. Box Contents T

3、emperature Probe, Battery (installed), Instruction Sheet, and Warranty Card. RadioFans.CN 收音机爱 好者资料库 Safety Information The 80T-IR/E complies with IEC Publication 1010-1- 1990 including Amendment 1, CSA C22.2 No. 231, ANSI/ISA-S82.01 and .03 Safety Standards. W WARNING IF TARGET EMMISSIVITY IS LESS

4、THAN 0.95, THE PROBE CAN INDICATE A TEMPERATURE LOWER THAN THE ACTUAL TARGET TEMPERATURE. AVOID TOUCHING THE TARGET; THERMAL BURNS COULD RESULT. CAUTION Do not place the probe on or around hot objects (70C / 158F). It will damage the probe case. If the probe is exposed to significant changes in ambi

5、ent temperature (hot to cold or cold to hot), allow 20 minutes for temperature stabilization, before taking measurements. Do not operate the probe near large electrical or magnetic fields such as arc welders and induction heaters. These fields can cause measurement errors. Condensation may form on t

6、he lens when going from a cold to hot environment - wait 10 minutes for condensation to dissipate before taking measurements. Connectors must only be plugged into voltage measurement input jacks of the test instruments. Do not touch or hold by the front cone. Temperature readings can be affected by

7、heat from the hand. Equipment use not specified by manufacturer may impair safety. RadioFans.CN 收音机爱 好者资料库 Compatibility The probe is compatible with all DC millivolt measuring instruments that have a minimum of 1 MW input impedance and accept safety shrouded, standard diameter 0.16 in. (4 mm) banan

8、a plugs. Operation To take a measurement, perform the following steps: 1. Plug the red connector into the VW dc input jack and the black connector into the common or ground input jack on the test instrument. 2. Select mV dc on the test instrument. 3. Slide the probe switch forward to the ON position

9、. 4. Point the tip of the probe as close as possible to the object being measured without touching the object. 5. Read the test instrument display. Additional considerations are: After 10 minutes of use the probe will automatically shift to Sleep mode (the display will show 0F). It can be restarted

10、by sliding the switch to OFF and then to ON (see Table 1). Sleep mode extends battery life. However, for maximum battery life, switch the probe to the OFF position. If the test instrument displays an overload condition, switch the DMM range from mV dc to V dc. Increasing the range to V dc moves the

11、decimal position three places to the left (1000F displays as 1.000V dc). RadioFans.CN 收音机爱 好者资料库 300mV V V OFF 0102030 300mV V V OFF A A ! 1000V 750V FUSED 10A RANGE PRESS AUTORANGE 300 mA COM V 1 SEC TOUCH HOLD Avoid touching or holding cone ON/OFF switch Battery clip Battery 9V DC Battery door Coi

12、led cable Banana plugs 3 2 1 4 C F C/F switch Figure 1. Display Codes Under the conditions shown in Table 1, the meter will alternate between displaying a reading and Display Code. Emissivity All objects emit invisible infrared energy. This ability, called emissivity, is based upon the material that

13、 the object is made of and its surface finish. Emissivity values typically range from 0.10 for a very reflective object to 0.98 for a near perfect black body. The probe senses this energy assuming that the target has an emissivity value of 0.95. This value is factory set in the probe. If the actual

14、target emissivity is less than 0.95, the indicated temperature could be less than the actual target surface temperature. To correct for this, apply masking tape or a coat of matte paint to the target. The resulting target will have an emissivity of approximately 0.95. Table 1. Display CodesCondition

15、Action 1041FTarget temperature is over range. Select target within probes specified temperature range. -135FThe temperature of the probe is near the low ambient operating range limit.* Ensure that the probe is within the specified ambient operating range. -192FBattery power is low.Replace the batter

16、y. 0FSleep mode or battery is dead. Restart the probe by sliding switch to OFF and then to ON, or replace battery. * All values are nominal. * Although a display code may be present, the displayed reading is valid if the probe is within the specified ambient operating range. Distance to Spot Size Ra

17、tio Distance to Spot Size Ratio (or Field of View) refers to the diameter of the spot that the probe is sensing at a given distance. The closer you are to the object (or target), the smaller the area (or spot) the probe is sensing. For example when the probe is held at a 200 mm (8 in.) distance from

18、 the target, the spot size is approximately 50 mm (2 in.); at 100 mm (4 in.) the spot size is approximately 25 mm (1 in.), and with the probe held at a 50 mm (2 in.) distance from the target, the spot size is approximately 13 mm (1/2 in.). Hot spots can be missed if too large an area is included in

19、the field of view, so get as close as possible! (See Figure 2.) 2.5 0.1 14 0.6 7.5 0.3 21 0.8 33 1.3 mm Inch 0 0 25 1 50 2 76 3 130 5 mm Inch Distance to object Diameter of spot Spot size increases with distance from the probe tip, as shown Figure 2. Measurement Considerations If the surface to be m

20、easured is small (13 mm (1/2 in.) or less), hold the probe as close as possible to the surface (no more than 50 mm (2 in.) away). If the surface to be measured is covered by frost or other material, clean it to expose the surface. If the surface to be measured is highly reflective such as polished m

21、etal, apply masking tape or a matte finish paint to the surface. If the probe seems to be giving incorrect readings check the front of the probe. There may be condensation or debris obstructing the sensor; clean per instructions in the maintenance section. Quick Check For a quick check of the probe,

22、 point it directly at ice immersed in water (slush), the meter should read, within specifications limits, 32F (see Accuracy specifications). Specifications Temperature Range:32F to 1000F Operating Temperature:32F to 122F Accuracy for 1 year: (Operating temperature 64F to 82F) 32F to 180F: 5.4F 180F

23、: 3% of reading Temperature Coefficient:0.3% of reading or 0.7F, whichever is greater, change in accuracy per 1.8F change in ambient operating temperature above 82F or below 64F. Response Time:1 second Spectral Response:8 to 14 microns nominal Emissivity:pre-set 0.95 Output:1 mV/degree F Relative Hu

24、midity:95% RH or less 86F noncondensing, Temp. Coef. applies Storage Temperature:-13 to 158F without battery Power:9V battery; (NEDA 1604A, 6F22, 006P) Battery life (Alkaline):50 hours typical, 73F 33% duty cycle Dimensions:(L x W x H) 180 mm x 30 mm x 50 mm (7.1 in.) x (1.2 in.) x (2 in.) Weight:18

25、0 gm (6.4 oz) Maintenance Battery Replacement Remove battery door (See Figure 1.) and replace with a 9V Alkaline battery (ANSI/NEDA 1604A, IEC 6LR61). Front-Window Cleaning (as necessary) 1. Blow off loose particles using clean compressed air. 2. Gently brush remaining debris away with a camel hair

26、brush or Q-tip. 3. Carefully wipe the surface with a moist Q-tip. The swab may be moistened with water or a water- based glass cleaner. Allow to air dry. (Do not use solvents to clean the window.) Case Cleaning To clean the exterior housing, simply use soap and water or a mild commercial cleaner. Wi

27、pe with a damp sponge or soft rag. LIMITED WARRANTY & LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are

28、 warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Flukes opinion, has been misused, altered, neglected, contaminated, or damaged by accident or a

29、bnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without

30、interruption. Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales ou

31、tlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country. Flukes warranty obligation is limited, at Flukes option, to refund

32、of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period. To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then sen

33、d the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines th

34、at failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the products specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repai

35、r costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point). THIS WARRANTY IS BUYER?S SOLE AND EXCLUSIVE REMEDY AND I

36、S IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM A

37、NY CAUSE OR THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held in

38、valid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision Fluke CorporationFluke Europe B.V. P.O. Box 9090P.O. Box 1186 Everett, WA 98206-90905602 BD Eindhoven U.S.A.The Netherlands 11/99 Se

39、rvice For service information in the U.S.A., call 1-800-526- 4711. Outside the U.S.A., contact the nearest Fluke Service Center. To locate an authorized service center, visit us on the World Wide Web: or call Fluke using the phone numbers listed below: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-

40、800-36-FLUKE (1-800-363-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 Calibration Fluke recommends that the user return the probe annually to a Fluke Service Center for calibration, starting one year after purchase. Replacement Part Battery (Alkaline) - PN 614487 Calibration Procedures - PN 650900 PN 601788 February 1996 Rev. 3, 4/00 1993,1994,1996,1997,2000 Fluke Corporation. All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.

展开阅读全文
相关资源
猜你喜欢
相关搜索
资源标签

当前位置:首页 > 仪器仪表 > Fluke

copyright@ 2008-2025 收音机爱好者资料库 版权所有
备案编号:鄂ICP备16009402-5号