FLUKE 5700A, 5720A Getting Started 电路图.pdf

上传人:cc518 文档编号:200533 上传时间:2025-03-08 格式:PDF 页数:68 大小:808.67KB
下载 相关 举报
FLUKE 5700A, 5720A Getting Started 电路图.pdf_第1页
第1页 / 共68页
FLUKE 5700A, 5720A Getting Started 电路图.pdf_第2页
第2页 / 共68页
FLUKE 5700A, 5720A Getting Started 电路图.pdf_第3页
第3页 / 共68页
亲,该文档总共68页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

《FLUKE 5700A, 5720A Getting Started 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 5700A, 5720A Getting Started 电路图.pdf(68页珍藏版)》请在收音机爱好者资料库上搜索。

1、 5700A/5720A Series II Multi-Function Calibrator Getting Started PN 1668111 March 2002 2002 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 LIMITED WARRANTY the display shows linearity, offset, and scale e

2、rrors Keys that multiply and divide the output value by 10 to simplify work on meters with calibration points at decade multiples of a fraction of full-scale Programmable entry limits used for restricting the levels that can be keyed into the calibrator, preventing access to levels that may be harmf

3、ul to equipment or personnel A s key that provides the capability of displaying the instruments specification at the selected operating point, calibration interval, and specification confidence level 5700A/5720A Series II Getting Started 2 An auxiliary current binding post that allows you to calibra

4、te meters with separate current inputs without moving cables Real-time clock and calendar for date stamping reports Offset and scaling modes that simplify linearity testing of multimeters Variable phase reference signal output and phase-lock input Interface for the Fluke 5725A Amplifier Standard IEE

5、E-488 (GPIB) interface, complying with ANSI/IEEE Standards 488.1-1987 and 488.2-1987 Selectable normal remote mode or emulation of the Fluke 5100B and 5200A Series calibrators in functions and response to system controller software EIA Standard RS-232C serial data interface for printing, displaying,

6、 or transferring internally-stored calibration constants, and for remote control of the calibrator Extensive internal self-testing and diagnostics of analog and digital functions A traceable calibration procedure for all modes and ranges that requires only 10 V, 1 , and 10 k external standards, with

7、 only occasional independent verification Fast, simple, automated calibration check providing added confidence between calibration recalls, and data that can be used to document and characterize the calibrators performance between calibration recalls Instruction Manuals The 5700A/5720A Series II Cal

8、ibrators ship with a complete manual set that contains information for the operator and service or maintenance technicians. The set includes: 5700A/5720A Series II Getting Started Manual (PN 1668111) 5700A/5720A Series II Operator Reference Guide (PN 601648) 5700A/5720A Series II Remote Programming

9、Reference Guide (PN 601655) 5700A/5720A Series II Operator Manual (provided on CD-ROM, PN 1668127, or a printed copy is available for purchase through the Fluke Service Department under PN 601622) 5700A/5720A Series II Service Manual (provided on CD-ROM, PN 1668127, or a printed copy is available fo

10、r purchase through the Fluke Service Department under PN 105798) Order additional copies of these instruction manuals separately using the part numbers provided. For ordering instructions, refer to the Fluke Catalog or contact a Fluke sales representative. Getting Started Wideband AC Voltage Module

11、(Option 5700A-03) 3 Wideband AC Voltage Module (Option 5700A-03) The Wideband AC Voltage Module (Option 5700A-03) can be installed in both the 5700A and 5720A Series II Calibrators. The module is a high-accuracy, low-noise, extremely flat ac voltage source for calibrating rf voltmeters, with a frequ

12、ency range of 10 Hz to 30 MHz. Output is in seven ranges from 300 V (-57 dBm) to 3.5 V (+24 dBm) through a Type-N coaxial connector into a 50 load. The output level is selected in volts or dBm through either the front panel controls or under remote control. The wideband module also functions with th

13、e calibrators output adjust controls that let display the error of a wideband meter in either percentage of output or in decibels. Included with the wideband module is a Type-N output cable and a 50 terminator. The wideband module is calibrated to the end of its standard-equipment output cable. Auxi

14、liary Amplifier The Fluke Model 5725A Amplifier is available to extend the high voltage performance and current range of the calibrator: Interface connectors on the calibrators rear panel accept cables to directly operate a 5725A. Three amplifiers can be connected to the calibrator at the same time,

15、 but only one output can be active at a time. Once you have connected the amplifiers and configured the calibrator in a setup menu, amplifier operation is controlled by the calibrator. Operating instructions for the 5725A are provided in Chapter 4 of the Operators Manual. The General Specifications

16、include specifications for operating the calibrator with the 5725A. For other amplifier specifications, refer to their instruction manuals. Table summarizes the extended capabilities offered by the 5725A. Brief descriptions of the extended capabilities follow. Table 1. Auxiliary Amplifier Data Model

17、ModeRange 5725A AmplifierAC V20 to 1100 V rms up to 70 mA, 40 Hz to 30 kHz (50 mA 1 V/m but 3 V/m, add 0.01% of range. Relative Uncertainty 1 C 24 Hours90 Days180 Days1 Year24 Hours90 Days (ppm output + V) (ppm output + V) 220 mV10 nV0 5 + 0.57 + 0.50 8 + 0.50 9 + 0.50 2 + 0.42.5 + 0.4 2.2 V100 nV3.

18、5 + 0.84 + 0.84.5 + 0.80 6 + 0.80 2 + 0.82.5 + 0.8 11 V1 V2.5 + 33 + 33.5 + 30 4 + 30 1 + 31.5 + 3 22 V1 V2.5 + 53 + 53.5 + 50 4 + 50 1 + 51.5 + 5 220 V10 V3.5 + 504 + 500 5 + 506 + 502 + 502.5 + 50 1100 V100 V0 5 + 5006 + 5000 7 + 5008 + 5002.5 + 4003 + 400 5720A 95% 95% Confidence Level RangeResol

19、ution Absolute Uncertainty 5 C from calibration temperature For fields strengths 1 V/m but 3 V/m, add 0.01 % of range. Relative Uncertainty 1 C 24 Hours90 Days180 Days1 Year24 Hours90 Days (ppm output + V) (ppm output + V) 220 mV10 nV4 + 0.40 6 + 0.46.5 + 0.47.5 + 0.41.6 + 0.402 + 0.4 2.2 V100 nV3 +

20、 0.73.5 + 0.70 4 + 0.705 + 0.71.6 + 0.702 + 0.7 11 V1 V2 + 2.52.5 + 2.503 + 2.53.5 + 2.50.8 + 2.51.2 + 2.5 22 V1 V2 + 42.5 + 403 + 43.5 + 40.8 + 41.2 + 4 220 V10 V3 + 403.5 + 4004 + 400 5 + 401.6 + 4002 + 40 1100 V100 V4 + 4004.5 + 40006 + 4006.5 + 40002 + 4002.4 + 400 5700A/5720A Series II Getting

21、Started 12 Table 3. 5700A Series II DC Voltage Specifications: 99 % and 95 % Confidence Levels 5700A 99% 99 % Confidence Level RangeResolution Absolute Uncertainty 5 C from calibration temperature For fields strengths 1 V/m but 3 V/m, add 0.01 % of range. Relative Uncertainty 1 C 24 Hours90 Days180

22、Days1 Year24 Hours90 Days (ppm output + V) (ppm output + V) 220 mV10 nV6.5 + .757 + .7508 + .7509 + .82.5 + .54 + .5 2.2 V100 nV3.5 + 1.26 + 1.207 + 1.208 + 1.22.5 + 1.24 + 1.2 11 V1 V3.5 + 35 + 407 + 408 + 41.5 + 33.5 + 4 22 V1 V3.5 + 65 + 807 + 808 + 81.5 + 63.5 + 8 220 V10 V5 + 1006 + 10008 + 100

23、09 + 1002.5 + 1004 + 100 1100 V100 V7 + 6008 + 60010 + 60011 + 6003 + 6004.5 + 600 5700A 95% 95 % Confidence Level RangeResolution Absolute Uncertainty 5 C from calibration temperature For fields strengths 1 V/m but 3 V/m, add 0.01 % of range. Relative Uncertainty 1 C 24 Hours90 Days180 Days1 Year24

24、 Hours90 Days (ppm output + V) (ppm output + V) 220 mV10 nV5.5 + 0.66 + 0.67 + 0.68 + 0.62 + 0.43.5 + 0.4 2.2 V100 nV3.5 + 15 + 16 + 17 + 12 + 13.5 + 1 11 V1 mV3 + 3.54 + 3.56 + 3.57 + 3.51.2 + 33 + 3.5 22 V1 mV3 + 6.54 + 6.56 + 6.57 + 6.51.2 + 63 + 7 220 V10 mV4 + 805 + 807 + 808 + 802 + 803.5 + 80

25、 1100 V100 mV6 + 5007 + 5008 + 5009 + 5002.4 + 5004 + 500 Getting Started DC Voltage Specifications 13 Table 4. DC Voltage Secondary Performance Specifications and Operating Characteristics Temperature Coefficient Adder Note 2 Noise RangeStability Note 1 1 C 10 -40 C0 -10 C and Linearity 1 C Bandwid

26、th 0.1-10 Hz Bandwidth 10 Hz-10 kHz 24 Hours40-50 Cpk-pkRMS (ppm output + V) (ppm output + V)/C (ppm output + V)V 220 mV 2.2 V 11 V 22 V 220 V 1100 V 0.3 + 0.3 0.3 + 1 0.3 + 2.5 0.4 + 5 0.5 + 40 0.5 + 200 00.4 + 0.1 00.3 + 0.1 0.15 + 0.2 00.2 + 0.4 00.3 + 5 00.5 + 10 1.5 + 0.5 1.5 + 2 0 1 + 1.5 1.5

27、+ 3 1.5 + 40 0 3 + 200 0 1 + 0.2 0 1 + 0.6 0.3 + 2 0.3 + 4 0 1 + 40 0 1 + 200 0.15 + 0.1 0.15 + 0.4 0.15 + 2 0.15 + 4 0.15 + 60 0.15 + 300 5 15 50 50 150 500 Notes: 1. Stability specifications are included in the Absolute Uncertainty values in the primary specification tables. 2. Temperature coeffic

28、ient is an adder to uncertainty specifications that does not apply unless operating more than 5 C from calibration temperature. Minimum output: 0 V for all ranges, except 100 V for 1100 V range Maximum load: 50 mA for 2.2 V through 220 V ranges; 20 mA for 1100 V range; 50 output impedance on 220 mV

29、range; all ranges 25 Load regulation: (0.2 ppm of output + 0.1ppm of range), full load to no load Line regulation: 0.1 ppm change, 10 % of selected nominal line Settling time: 3 seconds to full accuracy; + 1 second for range or polarity change; + 1 second for 1100 V range Overshoot: 50 , 1000 pF Min

30、imum output: 10 % on each range External sense: Selectable for 2.2 V, 22 V, 220 V, and 1100 V ranges; 5700A/5720A 100 kHz, 5725A 30 kHz 1100 V6 mA600 pF Settling time to full accuracy: 5725A Amplifier:1000 pF Note 2 Frequency (Hz)Settling time (seconds) 040 Hz-5 kHz50 mA 120 k2 Notes: 1 Stability sp

31、ecifications are included in Absolute Uncertainty values for the primary specifications. 2. The 5725A will drive up to 1000 pF of load capacitance. Uncertainty specifications include loads to 300 pF and 150 pF as shown under Load Limits. For capacitances up to the maximum of 1000 pF, add Load Regula

32、tion. 3. 2.2 V Range, 100 kHz-1.2 MHz only: uncertainty specifications cover loads to 10 mA or 1000 pF. For higher loads, load regulation is added. 4. Applies from 0 C to 40 C 1000 100 10 0 3.5 V 10000 50 Hz10 Hz1 kHz100 kHz1 MHz30 MHz Voltage Frequency 40 Hz30 Hz 5725A 5700A-03 2.2 x 107 V-Hz 15 Hz

33、 5700A/5720A Volt-Hertz Capability ahp160f.eps Figure 2. Volt-Hertz Capability + 1 second for amplitude or frequency range change; + 2 seconds for 5700A/5720A 1100 V range; + 4 seconds for 5725A 1100 V range Overshoot: 30 Hz: 1 + 0.05/kHz), 30 Hz: 3 Input voltage: 1 V to 10 V rms sine wave (do not e

34、xceed 1 V for mV ranges) Frequency range: 10 Hz to 1.1999 MHz Lock range: 2 % of frequency Lock-in time: Larger of 10/frequency or 10 msec Phase reference: Selectable, rear panel BNC output Range: 180 Phase Uncertainty (except 1100 V range): 1 at quadrature points (0, 90, 180) elsewhere 2 Stability:

35、 0.1 Resolution: 1 Output level: 2.5 V rms 0.2 V Frequency range: 50 kHz to 1 kHz, usable 10 Hz to 1.1999 MHz 5700A/5720A Series II Getting Started 20 Resistance Specifications Table 10. 5720A Series II Resistance Specifications: 99 % and 95 % Confidence Levels 5720A 99% 99% Confidence Level Nominal

36、 Value Absolute Uncertainty of Characterized Value 5 C from calibration temperature Note 1 Relative Uncertainty 1 C 24 Hours90 Days180 Days1 Year24 Hours90 Days ppmppm 0 1 1.9 10 19 100 190 1 k 1.9 k 10 k 19 k 100 k 190 k 1 M 1.9 M 10 M 19 M 100 M 50 85 85 23 23 10 10 8 8 8 9 9 9 16 17 33 43 100 50

37、95 95 25 25 11 11 9 9 9 9 11 11 18 19 37 47 110 50 100 100 26 26 11.5 11.5 9.5 9.5 9.5 9.5 12 12 20 21 40 50 115 50 110 110 27 27 12 12 10 10 10 10 13 13 23 24 46 55 120 50 32 25 5 4 2 2 2 2 2 2 2 2 2.5 3 10 20 50 50 40 33 8 7 4 4 3 3 3 3 3 3 5 6 14 24 60 5720A 95% 95 % Confidence Level Nominal Valu

38、e Absolute Uncertainty of Characterized Value 5 C from calibration temperature Note 1 Relative Uncertainty 1 C 24 Hours90 Days180 Days1 Year24 Hours90 Days ppmppm 0 1 1.9 10 19 100 190 1 k 1.9 k 10 k 19 k 100 k 190 k 1 M 1.9 M 10 M 19 M 100 M 40 70 70 20 20 8 8 6.5 6.5 6.5 7.5 7.5 7.5 13 14 27 35 85

39、 40 80 80 21 21 9 9 7.5 7.5 7.5 7.5 9 9 15 16 31 39 95 40 85 85 22 22 9.5 9.5 8 8 8 8 10 10 17 18 34 42 100 40 95 95 23 23 10 10 8.5 8.5 8.5 8.5 11 11 20 21 40 57 100 40 27 20 4 3.5 1.6 1.6 1.6 1.6 1.6 1.6 1.6 1.6 2 2.5 8 16 40 40 35 26 7 6 3.5 3.5 2.5 2.5 2.5 2.5 2.5 2.5 4 4 12 20 50 Note: 1. Speci

40、fications apply to displayed value. 4-wire connections, except 100 M. Getting Started Resistance Specifications 21 Table 11. 5700A Series II Resistance Specifications: 99 % and 95 % Confidence Levels 5700A 99% 99 % Confidence Level Nominal Value Absolute Uncertainty of Characterized Value 5 C from c

41、alibration temperature Note 1 Relative Uncertainty 1 C 24 Hours90 Days180 Days1 Year24 Hours90 Days ppmppm 0 1 1.9 10 19 100 190 1 k 1.9 k 10 k 19 k 100 k 190 k 1 M 1.9 M 10 M 19 M 100 M 50 85 85 26 24 15 15 11 11 9 9 11 11 16 17 33 43 110 50 95 95 28 26 17 17 12 12 11 11 13 13 18 19 37 47 120 50 10

42、0 100 30 28 18 18 13 13 12 12 14 14 20 21 40 50 125 50 110 110 33 31 20 20 15 15 14 14 16 16 23 24 46 55 130 50 32 25 5 4 2 2 2 2 2 2 2 2 2.5 3.5 10 20 50 50 40 33 8 7 4 4 3.5 3.5 3.5 3.5 3.5 3.5 5 6 14 24 60 5700A 95% 95 % Confidence Level Nominal Value Absolute Uncertainty of Characterized Value 5

43、 C from calibration temperature Note 1 Relative Uncertainty 1 C 24 Hours90 Days180 Days1 Year24 Hours90 Days ppmppm 0 1 1.9 10 19 100 190 1 k 1.9 k 10 k 19 k 100 k 190 k 1 M 1.9 M 10 M 19 M 100 M 50 70 70 21 20 13 13 9 9 7.5 7.5 9 9 13 14 27 35 90 50 80 80 23 22 14 14 10 10 9.5 9.5 11 11 15 16 31 39

44、 100 50 85 85 27 24 15 15 11 11 10.5 10.5 12 12 17 18 34 42 105 50 95 95 28 27 17 17 13 13 12 12 14 14 20 21 40 47 110 50 32 25 5 4 2 2 2 2 2 2 2 2 2.5 3 10 20 50 50 40 33 8 7 4 4 3.5 3.5 3.5 3.5 3.5 3.5 5 6 14 24 60 Note: 1. Specifications apply to displayed value. 4-wire connections, except 100 M.

45、 5700A/5720A Series II Getting Started 22 Table 12. Resistance Secondary Performance Specifications and Operating Characteristics Nominal Value Stability 1 C Note 1 Temperature Coefficient Adder Note 2 Full Spec Load Range Maximum Peak Current Maximum Difference of Characterized Two-Wire Adder activ

46、e compensation Note 4 10 -40 C0 -10 C and Note 3to Nominal Value Lead Resistance 24 Hours40 -50 C 0.1 1 ppmppm/CmAmAppmm 0 00 8 - 500500 2 + Im 4 V 4 + Im 4 V 13245 00 8 - 100700 5002 + Im 4 V 4 + Im 4 V 1.92567 00 8 - 1005005002 + Im 4 V 4 + Im 4 V 10523 00 8 - 11 2203002 + Im 4 V 4 + Im 4 V 19423

47、00 8 - 111603002 + Im 4 V 4 + Im 4 V 100223 00 8 - 11701502 + Im 4 V 4 + Im 4 V 190223 00 8 - 11501502 + Im 4 V 4 + Im 4 V 1 k223 00 1 - 2221501015 1.9 k223 00 1 - 1.5161501015 10 k223100 - 500 A71505060 19 k223 0 50 - 250 A5150100120 100 k22310 - 100 A1150Im= Current produced by Ohmmeter 190 k22300

48、 5 - 50 A500 A150 1 M2.52.5600 5 - 20 A100 A200 1.9 M3.531002.5 - 10 A50 A200 10 M1052000.5 - 2 A10 A300 19 M208400.25 - 1 A5 A300 100 M50121000 50 - 200 nA 1 A 500 Notes: 1. Stability specifications are included in the Absolute Uncertainty values in the primary specification tables. 2. Temperature

49、coefficient is an adder to uncertainty specifications that does not apply unless operated more than 5 C from calibration temperature, or calibrated outside the range 19 C to 24 C. Two examples: Calibrate at 20 C: Temperature coefficient adder is not required unless operated below 15 C or above 25 C. Calibrate at 26 C: Add 2 C temperature coefficient adder. Additional temperature coefficient adder is not required unless operated below 21 C or above 31 C. 3. Refer to current derating factors