1、 PN 601655 May 1996 1996 Fluke Corporation, Inc. All rights reserved. Printed in U.S.A. 5700A/5720A Series II Multi-Function Calibrator Remote Programming Reference Guide RadioFans.CN 收音机爱 好者资料库 RadioFans.CN 收音机爱 好者资料库 1 Contents Syntax Rules.3 Command Summary by Function.5 Commands.10 Operating Sta

2、te Transitions.20 Status Register Summary.21 RadioFans.CN 收音机爱 好者资料库 2 3 Syntax Rules Spaces Use at least one space between a command and a parameter. Do not use spaces within a parameter, or within a number. Multiple Parameters When a command has more than one parameter, separate the parameters wit

3、h commas. For example: OUT 1 V, 100 HZ. Numeric Parameters Parameters may have up to 255 significant figures. Exponents may range from -32000 to +32000. The useful range for 5720A programming is 2.2 E-308 to 1.8 E308. Case Sensitivity All lower case letters are converted to upper case except in quot

4、ed strings. Extra Parameters Avoid using extra parameters. Too many parameters will cause a command error. Null Parameters Null parameters will cause a command error (e.g., adjacent commas in CLOCK 133700, ,071787). Expressions Do not use expressions as parameters. An example of an expression is (4+

5、2*13). Units Accepted units in command parameters are: A amperes DB decibels DBM decibels references to 1 milliwatt for ac voltage amplitude HZ hertz OHM ohms PCT percent PPM parts per million V volts 4 Syntax Rules (continued) Multipliers MA mega, or unit x 1,000,000) K kilo, or unit x 1,000) M mil

6、li, or unit 1,000; (except MOHM and MHZ) U micro, or unit 1,000,000) Examples: MOHM or MAOHM Megohms MHZ or MAHZ Megahertz MV millivolts MAV Megavolts MA millliamperes Multiple Commands To combine commands in one statement, use a semicolon (;). For example, instead of using these two statements: 20

7、PRINT 3,OUT 100 MV ! Select 100 mV dc 30 PRINT 3,OPER ! Activate the output You can accomplish the same with this single line: 20 PRINT 3,OUT 100 MV ; OPER Terminators Terminators sent by the 5720A: LF with EOI. Terminators received by the 5720A: LF or any character with EOI Responses The responses

8、described in the command table are correct for IEEE-488 remote control, and for serial remote control in COMPUTER mode. In serial remote control in TERMINAL mode, responses contain more descriptive text intended for an operator using a terminal interactively. 5 Command Summary by Function Error Mode

9、 Commands ADJOUT? Returns adjusted output magnitude and frequency. INCR Increments or decrements the output MULT Multiplies reference by a value and establishes the product as a new reference. NEWREF Establishes a new reference. OFFSET Sets and enables or disables an offset. OFFSET? Returns the valu

10、e of the present offset. OLDREF Sets the output to the present reference. OUT_ERR? Returns the UUT error computed by the 5720A. REFOUT? Returns the value of the present reference. SCALE Activates and deactivates scaling. SCALE? Returns scaling information. SCAL_ERR? Returns the scale error value if

11、scaling is active. Instrument Configuration Commands BTYPE Selects an amplifier type for voltage or current boost. BTYPE? Returns the amplifier types set for voltage and current boost. CLOCK Sets the clock/calendar. CLOCK? Returns the setting of the clock/calendar. CUR_POST Selects active binding po

12、st for current output. CUR_POST? Returns active binding post for current output. EXTGUARD Opens and closes an internal connection. between GUARD and LO. EXTSENSE Opens and closes an internal connection between SENSE and OUTPUT. FORMAT Restores calibration constants and other parameters to factory de

13、faults. LIMIT Sets positive and negative output limits. LIMIT? Returns the programmed positive and negative output limits. 6 Command Summary by Function (continued) *LRN? Returns a list of commands that can duplicate the present state. RANGE? Returns the present output range. RANGELCK Locks the pres

14、ent range, or selects autoranging. Output Commands BOOST Activates or deactivates an auxiliary amplifier. DBMOUT? Returns the output amplitude and frequency, but in dBm if ac V. OPER Activates the 5720A output if it is in standby. OUT Sets the output and establishes a new reference for error mode. O

15、UT? Returns the output amplitude and frequency. PHASE Sets the phase shift of the phase output signal. PHASE? Returns the phase shift of the phase output signal. PHASELCK Activates or deactivates phase locking to an external signal. PHASESFT Activates or deactivates variable phase output. RCOMP Acti

16、vates or deactivates two-wire compensation circuitry. STBY Puts the 5720A in standby. VOUT? Returns the output amplitude in volts if the active units are dBm. WBAND Activates or deactivates the Wideband AC Module (Option 5700A-03). General Functions *CLS Clear; clears status registers, any service r

17、equest, and flags. ECHO? Echoes a string to the remote port. GAL Go to Alternate Language; enters Fluke 5100B or 5200A emulation mode. *OPC Sets bit 0 in the ESR to 1 when pending operations are complete. 7 Command Summary by Function (continued) *OPC? Returns a 1 when all pending remote operations

18、are complete. *PUD Stores a string of user-selected characters in non-volatile memory. *PUD? Returns the contents of the PUD memory. RPT_STR Loads the user report string. RPT_STR? Returns the user report string. *RST Resets the 5720A to its default power-up state. UNCERT? Returns the absolute uncert

19、ainty of the output. *WAI Halts program execution until pending operations are complete. RS-232C Interface Parameter Setting Commands SP_EOF Sets the End-of-File (EOF) string. SP_EOF? Returns the End-of-File (EOF) string. SP_SET Sets RS-232C interface parameters. SP_SET? Returns RS-232C interface pa

20、rameters. SP_TIME Sets the time-out period. SP_TIME? Returns the time-out period. 5720A Calibration, Testing, and Diagnostics Commands CAL_ADJ Does the internal portion of calibration. CAL_CHK Starts a calibration check. CAL_CLST? Returns a group of calibration constant names and their values. CAL_C

21、ONF Sets the specification confidence level to 95% or 99%. CAL_CONF? Returns the active calibration confidence level. CAL_CONST? Returns the value of a particular constant. CAL_DATE? Returns the date of the most recent calibration. CAL_DAYS? Returns number of days since last calibration recall. CAL_

22、INTV Sets the calibration interval. CAL_INTV? Returns the calibration interval. 8 Command Summary by Function (continued) CAL_PR Prints a calibration report out the serial interface. CAL_REF Performs a calibration procedure using a 1, 10 k, or 10V standard. CAL_RNG Starts a self calibration procedur

23、e of a range. CAL_RPT? Returns a formatted calibration report. CAL_SHIFT? Returns the shift of a range due to calibration. CAL_SLST? Returns the shifts of a group of ranges. CAL_STORE Stores new calibration constants in nonvolatile memory. CAL_TEMP Sets the temperature for calibration. CAL_TEMP? Ret

24、urns the calibration temperature last specified. CAL_WBFLAT Performs Wideband AC Module (Option 5700A- 03) flatness calibration. CAL_WBGAIN Performs Wideband AC Module (Option 5700A- 03) gain calibration. CAL_ZERO Performs internal zeros calibration. DIAG Runs self diagnostics. DIAGFLT Sets the 5720

25、A response to faults in remote diagnostics. DIAGFLT? Returns the 5720A response to faults in remote diagnostics. OHMSREF? Returns a calculated resistance reference. STOP_PR Stops printing a report. *TST? Performs self tests and returns the results. Serial Only Remote Commands LOCAL Places the 5720A

26、into the local state. LOCKOUT Places the 5720A into the local lockout state. REMOTE Places the 5720A into remote state. Status Commands *ESE Loads the Event Status Enable register. *ESE? Reads the Event Status Enable register. 9 Command Summary by Function (continued) *ESR? Reads and clears the Even

27、t Status Register. ETIME? Returns the time power has been on. EXPLAIN? Returning a brief description of a fault code. FAULT? Returns the most recent fault code from the fault queue. *IDN? Returns 5720A identification information. ISCE Loads Instrument Status Change Enable Register. ISCE? Reads Instr

28、ument Status Change Enable Register. ISCR? Reads and clears the Instrument Status Change Register. ISR? Reads and clears the Instrument Status Register. ONTIME? Returns the time elapsed since power-up. *OPT? Returns a list of installed modules and attached amplifiers. *SRE Loads the Service Request

29、Enable Register. *SRE? Reads the Service Request Enable Register. STATE? Returns the long term state of the 5720A. *STB? Returns the status byte. 10 Commands Command Parameters Response ADJOUT? None 1. (Float) Amplitude 2. (String) Units 3. (Float) Frequency BOOST ON or OFF None BTYPE VB5725, VB5205

30、, VB5215, IB5725, or IB5220 None BTYPE? None 1. (String) VB 2. (String) IB CAL_ADJ None None CAL_CHK None None CAL_CLST? CAL, CHECK, or PREV 1. (String) List (See manual.) CAL_CONF 99 or 95 None CAL_CONF? None CONF99 or CONF95 CAL_CONST? 1. CAL, CHECK PREV 2. Name of constant (Float) Value of the co

31、nstant CAL_DATE? B5725, CAL, WBFLAT, WBGAIN, or ZERO (Integer) Date as MMDDYY 11 Commands (continued) Command Parameters Response CAL_DAYS? B5725, CAL, WBFLAT, WBGAIN, or ZERO (Integer) # of elapsed days since calibration CAL_INTV 1, 90, 180, or 365 None CAL_INTV? None (Integer) 1,90, 180, or 365 CA

32、L_PR CAL, CHECK, or RAW None CAL_REF Value and units of external standard None CAL_RNG 1. range identifier, NULL or STORE 2. Calibration magnitude (Only if 1 is range id) 3. Frequency (Optional, and only if 1 is range id) None CAL_RPT? CAL, CHECK, or RAW 1. (String) Formatted report CAL_SHIFT? 1. CA

33、L or CHECK 2. Range identifier 1. (String) Range info. (See manual.) CAL_SLST? CAL or CHECK 1. (String) Info. for all ranges. (See manual.) CAL_STORE None None CAL_TEMP Temperature in C None CAL_TEMP? B5725, CAL, WBFLAT, WBGAIN, or CHECK (Float) Temperature in C. 12 Commands (continued) Command Para

34、meters Response CAL_WBFLAT START, NULL, or STORE None CAL_WBGAIN PGAIN, NGAIN, or STORE None CAL_ZERO None None CLOCK 1. 24-hour time as HHMMSS 2. Date as MMDDYY None CLOCK? None 1. (Integer) Time as HHMMSS 2. (Integer) Date as MMDDYY *CLS None None CUR_POST NORMAL, AUX, or IB5725 None CUR_POST? Non

35、e (String) NORMAL, AUX, or IB5725 DBMOUT? None 1. (Float) Output value 2. (String) Units 3. (Float)Frequency DIAG ALL, D5700, DV5725, DI5725, CONT, or ABORT None 13 Commands (continued) Command Parameters Response DIAGFLT HALT, ABORT, or CONT None DIAGFLT? None (String) HALT, ABORT, or CONT ECHO? An

36、y string (String) Same string *ESE Decimal equiv. of 16- bit word (0- 255 only) None *ESE? None (Integer) Decimal equiv. of byte *ESR? None (Integer) Decimal equiv. of byte ETIME? None (Integer) minutes powered on EXPLAIN? Fault code (String) Explanation text EXTGUARD ON or OFF None EXTSENSE ON or O

37、FF None FAULT? None (Integer) The fault code FORMAT ALL, B5725, CAL, RNG, or SETUP None GAL L5100 None *IDN? None (String, string, string, string) Fluke, model #, serial #, main CPU s/w rev.+ inguard CPU s/w rev.+ BOOST CPU s/w rev.) 14 Commands (continued) Command Parameters Response INCR step size

38、 None ISCE Decimal equivalent of 16-bit word None ISCE? None Decimal equivalent of 16-bit word ISCR? None Decimal equivalent of 16-bit word ISR? None Decimal equivalent of 16-bit word LIMIT 1. Positive limit with unit 2. Negative limit with unit None LIMIT? None (Float, float, float, float) 1. Posit

39、ive voltage limit 2. Negative voltage limit 3. Positive current limit 4. Negative current limit LOCAL None None LOCKOUT None None *LRN? None (String) Re-configuration command MULT Multiplier expressed as a floating point number None NEWREF None None OFFSET ON or OFF None OFFSET? None 1. (Float) Offs

40、et value 2. (Float) Offset units 15 Commands (continued) Command Parameters Response OHMSREF? CAL, CHECK, or PREV (Float) Resistance reference OLDREF None None ONTIME? None (Integer) Time unit is powered on (minutes) *OPC None None *OPC? None (Integer) 1 OPER None None *OPT? None (Strings separated

41、by commas) OUT Either one of the following is required. 1. (Optional) Amplitude or 2. (Optional) Frequency None OUT? None 1. (Float) Amplitude 2. (String) Units 3. (Float) Frequency OUT_ERR? None 1. (Float) UUT error magnitude 2. (String) Units for above PHASE phase in degrees None PHASE? None (Inte

42、ger) Phase in degrees PHASELCK ON or OFF None PHASESFT ON or OFF None 16 Commands (continued) Command Parameters Response *PUD (See manual for parameter.) None *PUD? None (See manual for response.) RANGE? None (String) Name of range RANGELCK ON or OFF None RCOMP ON or OFF None REFOUT? None 1. (Float

43、) Output 2. (String) Units 3. (String) Frequency REMOTE None None RPT_STR String of up to 40 char. None RPT_STR? None (String) User report string *RST None None SCALE ON or OFF None SCALE? None 1. (Float) Nominal full-scale 2. (Float) Actual full-scale 3. (String) Units for above two SCAL_ERR? None

44、1. (Float) Scale error 2. (String) Units for scale error 17 Commands (continued) Command Parameters Response SP_EOF 1. First ASCII code 2. Second ASCII code None SP_EOF? None 1. (Integer) First ASCII code 2. (Integer) Second ASCII code SP_SET 1. 110, 300, 600,1200, 2400, 4800, (9600), or 19200 2. (T

45、ERM) or COMP 3. (XON), RTS, or NOSTALL 4. DBIT7 or (DBIT8) 5. (SBIT1) or SBIT2 6. PNONE, (PEVEN), PODD, or PIGNORE 7. CR, LF, or (CRLF) None 18 Commands (continued) Command Parameters Response SP_SET? None 1. (Integer) Baud rate 2. (String) TERM or COMP 3. (String) XON, RTS, or NOSTALL 4. (String) D

46、BIT7 or DBIT8 5. (String) SBIT1 or SBIT2 6. (String) PNONE, PEVEN, PODD, or PIGNORE 7. (String) CR, LF, or CRLF SP_TIME 0 - 30 None SP_TIME? None (Integer) 0 - 30, seconds *SRE Decimal equiv. of byte None *SRE? None Decimal equivalent of byte STATE? None (See manual for response.) *STB? None Decimal

47、 equivalent of byte STBY None None STOP_PR None None *TST? None (Integer) 0, pass; 1, fail 19 Commands (continued) Command Parameters Response UNCERT? None 1. (Float) 5720A uncertainty 2. (String) Units for response 1 3. (Integer) Cal interval in days VOUT? None 1. (Float) amplitude 2. (String) Units 3. (Float) Frequency *WAI None None WBAND ON or OFF None 20 Operating State Transitions From To Use 1722A GPIB Command Serial Command Local Remote MLA + REN REMOTE REMOTE Local/ Lockout LLO + PEN LOCKOUT LOCKOUT Remote Local GTL, or “GO TO LOCAL CONTROL” softkey LOCAL LO