1、BOONTON ELECTRONICS a Wireless Telecom Group Company 4400A/4500A MANUAL ADDENDUM Applies to Instruction Manual Model 4500A RF Peak Power Meter /Analyzer, Model 4400A RF Peak Power Meter, Part Number 98404700A, revised 09/97. Effective for Control Software Revision 20020511 and later. 1.0 NEW FEATURE
2、S 1.1 Peak Sensor Temperature Compensation When used with a peak sensor that contains a valid temperature compensation table, the model 4400A and 4500A Peak Power Meters can provide temperature compensated power measurements. The default mode for temperature compensation is active. The Chan n Calibr
3、ationTemp Comp menu box will be visible with “Sensor Tbl” displayed. To turn off temperature compensation press the menu button and “Off” will be displayed. The “Off” setting is volatile and not preserved through power cycles or major mode changes. If the sensor in use does not have a temperature co
4、mpensation table the Temp Comp menu box will not appear at all. The SpclChan n Sensor report will include the message “Sensor has Temperature Compensation Table” when appropriate. The SpclServicingConfiguration report shows TC system status if either channel has a sensor with a valid table. The form
5、at of this message if present is: TC System Status 1:err code # 2:err code # The error codes are: 0 No error. 145 TC # of Temps Err Table parameter error 146 TC # of Powers Err Table parameter error 147 TC Interp Err Table interpolation error 148 TC Expand Err Table expansion error 149 TC Extend Err
6、 Table extension error 150 TC Chksum Err Table read checksum error 151 TC Table Length Error 152 TC Temp Value Err A temperature value is out of range 98404706A 1 RadioFans.CN 收音机爱 好者资料库 153 TC Temp Non-mono A temperature array is non-monotonic 154 TC Power Value Err A power value is out of range 15
7、5 TC Power Non-mono A power array is non-monotonic 156 TC Corr Value Err A correction value is out of range Use CH1 or CH2 to specify to which channel the following temperature compensation related GPIB commands apply: TCON Turn on temperature compensation if available. If not, ignore. TCOFF Turn of
8、f temperature compensation. This is a volatile setting if compensation is available. 1.2 Sensor Auto-calibration File Retention Sensor auto-calibration files are now saved by channel as .AC1 and .AC2 files in non- volatile (flash) memory. This avoids the need to perform auto-cal every time a sensor
9、is removed and replaced by a different one, provided the sensors are known to the channel(s) involved. When a sensor is plugged-in a search is made to find an existing auto-cal file. If one is found, it is installed. If not, the “Needs Auto-cal” message will appear. When auto-cal is performed the ex
10、isting file is overwritten with the new result. If no previous file exists, one is created. Sensor filenames have the form SEN#.AC n, where # is the serial number and n is the channel number. The file directory system is expanded to display auto-calibration files in flash memory as well as the previ
11、ous files on the floppy disk. The UtilityDiskFlash Disk path lists sensor auto-cal files by channel. The Select File menu contains a sequence number which refers to the position of the file in the list. The selected file is shown in RED and may be deleted by pressing the menu button next to the “Del
12、ete” box. Deletion of files must be confirmed or cancelled. There are no GPIB operations on the file directory. 1.3 Color *.bmp File of Display The Hardcopy section now includes a color *.bmp file of the current display that can be saved to the floppy disk, sent to the COM1 port or the GPIB. To sele
13、ct this feature set the UtilHardcopyDevice menu to “Plotter”. Then select UtilHardcopyModel “.BMP”. Choose the Plot Label, Output Port and File Select number if the output is the floppy disk. Graph & Text is not applicable. Note that the “IEEE-488” output selection applies to the listen only (lon) G
14、PIB addressing mode only. For controller directed GPIB output 98404706A 2 RadioFans.CN 收音机爱 好者资料库 see below. Press the PLOT key to send the file to the selected output. The .bmp extension is added to the floppy disk file directory to allow viewing the filenames saved on disk. The GPIB commands for c
15、ontroller directed return of the .bmp file contents are: Send the sequence PLOTTER PLOT.BMP to select the bit-map mode. Send TKPLOT to set the talk mode that returns the file when addressed to talk. The GPIB commands to send the plot file contents to an output other than the GPIB are: Send the seque
16、nce PLOTTER PLOT.BMP to select the bit-map mode. Send PLOTSER1, PLOTCOM1, PLOTLPT1 or PLOTDISK to select the output. Use PLOT488 only with the front panel PLOT key manually to send in the talk only (ton) mode to a listen only (lon) device. For PLOTDISK send FILENO # to select a filename. Send PLOT t
17、o simulate pressing the PLOT key to send the file to the selected output. 1.4 External Trigger Input Calibration A provision to zero and calibrate the external trigger inputs has been added to provide better accuracy for voltage measurements made with the trigger inputs. The following procedure is u
18、sed to calibrate each external trigger input: Set TimeTimebase to “5 ms/Div” Set TrigTrig Mode to “Auto” Select the external trigger input in the TrigTrig Source menu corresponding to the selected measurement channel (CH1 to 1EXT or CH2 to 2EXT). Set the Chan#ExtensionsDisplay to “Trig” (Trigger Vie
19、w Mode) Set the Chan#Vert Scale to 1.00 V/Div Set the SpclServicingCal Mode “On” With Cal Mode “On” two boxes labeled Ext Trig Zero and Ext Trig Cal will appear in the Chan#Extensions menu. Ext Trig Zero will have a bright “Start” label. With no input to the selected external trigger input, press th
20、e menu button for Ext Trig Zero “Start”. The input will be zeroed and the Ext Trig Cal “Start” label will be bright. With +3.00 volts applied to the selected external trigger input, press the menu button for Ext Trig Cal “Start”. The input will be calibrated for 3 divisions of deflection at 1 V/Div.
21、 Set the SpclServicingCal Mode “Off” The results of the calibration are stored in non-volatile memory with file extension .TRV. Absent a file, default data is supplied automatically and simulates the existing software. External trigger level calibration is not available on the GPIB. 98404706A 3 Radi
22、oFans.CN 收音机爱 好者资料库 1.5 UNDIM Command. A GPIB command, UNDIM, is added to reset the screen saver without touching the panel or re-loading the color table. This is helpful in remotely controlled applications. 2.0 Changes 2.1 Instrument Setup Save/Recall change. The instruments setup save and recall s
23、ystem has been modified to save a binary file instead of an ASCII file. The binary file is smaller and more comprehensive and is identical to the internal save/recall format. The new file has the extension .ISU. For customers with existing .INS files the ability to read an .INS file is still present
24、, but new features will not be available using this method. Existing files should be converted by reading the .INS files and saving them as .ISU. The file directory is modified to display .ISU files. The GPIB commands are not affected. 2.2 GPIB command *OPT? change. An installed hardware options lis
25、t has been appended to the *OPT? GPIB command format previously used. For example, a single channel instrument with Option 04 hardware installed and a sensor plugged-in returns: 1,1,0,0,4 2.3 Configuration report change. Installed options are now identified in the Configuration Report. For example:
26、The SpclServicingConfiguration Report for Option 04 installed reports: Opt 04 Trigger Delay by Events installed 2.4 Sensor Temperature Reporting change. Sensor auto-cal temperature and current temperature readings have been moved from the UtilityReport to the SpclCh 1 SensorReport and SpclCh 2 Senso
27、rReport. 98404706A 4 3.0 Corrections 1. Remove glitches that occur when in Triggered mode (as opposed to Auto) and certain commands are executed. Also, measurement traces can now be moved and re-scaled when Waiting for Trigger on the slow time bases. 2. Calculate the auto-measure parameter “OFF TIME
28、”. This function has always been enabled but there was no calculation method included. 3. Correct an overflow error in the cal table expansion that overwrites the first position of the channel 2 table when channel 1 is expanded. This may cause the channel 2 PDF to not appear. 4. GPIB command TKBMEAS
29、 now reports the sign of Pk/Avg ratio correctly. 5. Marker math mode changes now occur immediately even in wait for trigger. 6. When both markers are in trigger view mode the marker math functions MK1-MK2, MK2-MK1, MAX-MIN and MIN-MAX are computed as voltage difference and appear in the middle windo
30、w with voltage difference units. The PK/AVG mode is not recognized in trigger view mode but is not an error. This correction also appears in the parameters of GPIB commands TKMEAS, TKBMEAS and TKUNITS when appropriate. 7. The trigger pointer is now removed when the direct set of a statistical mode o
31、ccurs. 8. Restore the legacy GPIB command MKDELTA to set the marker math to power difference in the linear units mode only. This command was deleted by mistake in the “A” series. 9. Update the RUN/STOP message in the recall stored setup function to avoid out of sync messages. 10. Change the linear mode reference lines to track vertical offset in “divisions” rather than watts, which is incorrect. Revised 20020513 98404706A 5