1、 Fluke 433, 434, 435 Three Phase Power Quality Analyzer Service Manual 4822 872 05396 November 2006, Rev. 1, 05/2007 2006, 2007 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 Limited Warranty for the more

2、 accurate model Fluke 435 you must use Fluke 5700A. The tests mainly verify accuracy and offset of voltage and current inputs. Tests indicated with (*) are optional. They can be done to check if the Analyzers hardware (analog input channels) is OK. Bear in mind that the outcome of these checks is ra

3、ther influenced by hardware sanity than by Calibration Adjustment. Accuracy of all other Analyzer functions is linked to input accuracy and is embedded in the software. This link is tested extensively for each new software release. This testing is done with the Fluke 6100A Electrical Power Standard

4、and 6101A Auxiliary units. This setup allows multiphase operation and direct testing of measuring functions such as flicker, harmonics, interharmonics, fluctating harmonics, dips the test pattern as shown in Figure 3-1 may be not visible or hardly visible. Observe the display closely, and verify tha

5、t the display shows no abnormalities, as for example very light pixels or lines. Figure 3-1. Display Pixel Test Pattern 7. Press F2 . The test pattern is removed; the Analyzer shows Contrast (CL 0100): 8. Press F2 again to do the next step Contrast (CL 0110): 9. Press F3 CALIBRATE The Analyzer shows

6、 the display test pattern shown in Figure 3-1, at default contrast. Observe the display closely, and verify that the display shows no abnormalities. Also verify that the contrast of the upper left and upper right square of the test pattern is equal. 10. Press F2 . The test pattern is removed; the An

7、alyzer shows Contrast (CL 0110): 11. Press F2 again to do the next step Contrast (CL 0120): 12. Press F3 CALIBRATE The Analyzer shows a light display; the test pattern as shown in Figure 3-1 may not be visible or hardly visible. Observe the display closely, and verify that the display shows no abnor

8、malities. 13. Turn the Analyzer OFF and ON to exit the calibration menu and to return to the normal operating mode. If the maximum, minimum, or default display contrast is not OK, then you can set these items without performing a complete calibration adjustment; refer to Section 4 for detailed infor

9、mation. Performance Verification 3.7. Verification of Current Inputs 3 3-7 3.7. Verification of Current Inputs 3.7.1 Preparation Proceed as follows: Reset the Analyzer. Press SETUP . Set the Analyzer to 3-phase WYE, 60 Hz, 120 V. Set Clamp for Phase and Neutral to 1 mV/A, 400 A, 1:1. Set the Analyze

10、r in DEMO mode: press SETUP, press F3 - FUNCTION PREF, and select DEMO ON with F1. Demo mode gives increased sensitivity at the voltage inputs. Input voltage must not exceed 2 V rms in DEMO mode! 3.7.2 Accuracy Proceed as follows: 1. Must be checked for all phases A/L1, B/L2, C/L3, and N(eutral). 2.

11、 Set the Calibrator to 1 V, 60 Hz and STBY. 3. Connect the LO output (black) with the Ground (green) input of the Analyzer. Connect the HI output (red) with all current (BNC) inputs and the A/L1, B/L2, C/L3, and N voltage banana inputs of the Analyzer. Refer to Figure 3.2. A/L1B/L2C/L3NGND A/L1B/L2C

12、/L3N CALIBRATOR POWER ADAPTER BC 430 ANALYZER INPUT BLOCK LO HI NORMAL PM 9093PM 9093PM 9093 PM 9082RED PM 9092 Figure 3-2. verification of Current Inputs 4. Set the Calibrator to OPR (Operate, indicated by a green LED in the OPR key or on the display). 5. On the Analyzer press MENU, select Volts/Am

13、ps/Hertz, press F5 OK to enter the function. Fluke 433, 434, 435 Service Manual 3-8 6. Fluke 433/434: check for a current readout A rms between 985 . 1015 A (tolerance 10 + 5 A = 15 A). Fluke 435: check for a readout between 990 . 1010 A (tolerance 5 + 5 A = 10 A). (Voltage readout V rms must be app

14、rox. 0 V at A/L1, B/L2, C/L3 and 172 V at N inputs, frequency readout between 59.88 . 60.02 Hz, Analyzer in DEMO mode). 7. Set the Calibrator to 0 Hz, 0 V and then to OPR. 8. On the Analyzer check for a current readout A rms between 0 . 5 A. 3.7.2 Bandwidth check of current channels (*) Is an option

15、al test. Proceed as follows: 1. Set the Calibrator to 1 V, 60 Hz and OPR. 2. The current channels must be checked one by one. Use the Calibrators edit field function (5500A: FIELD EDIT key, 5700A: AMPL/FREQ key) to adjust a readout of 1000 A for the current channel to be checked. 3. Increase the fre

16、quency to 3 kHz. Check for a readout of 945 A or more. Table 3-1. Bandwidth Check of Current Channels Current Channel to be verified Readout at 60 Hz (adjust Calibrator with EDIT FIELD) Readout at 3 kHz Fluke 433/434/435 A/L1 1000 A 945 A or more B/L2 1000 A 945 A or more C/L3 1000 A 945 A or more N

17、 1000 A 945 A or more 4. Set the Calibrator to STBY. 5. Disconnect the leads from the current inputs. 6. Switch the Analyzers DEMO mode to OFF. 3.8 Voltage Inputs 3.8.1 Introduction WARNING Dangerous voltages will be present on the calibration source and connecting cables during the following steps.

18、 Ensure that the Calibrator is in standby mode before making any connection between the Calibrator and the Analyzer. Performance Verification 3.8 Voltage Inputs 3 3-9 Proceed as follows: 1. To check the A/L1, B/L2, C/L3 inputs, connect the N input to Ground (See Figure 3- 3). N now will give zero re

19、ading. A/L1B/L2C/L3NGND A/L1B/L2C/L3 N CALIBRATOR POWER ADAPTER BC 430 ANALYZER INPUT BLOCK LO HI NORMAL Figure 3-3. Checking the A/L1, B/L2, and C/L3 voltage inputs 2. To check the N input, connect the N input with the adjacent C/L3 input. A/L1, B/L2, C/L3 now will give zero reading. Refer to Figur

20、e 3-4. A/L1B/L2C/L3NGND A/L1B/L2C/L3 N CALIBRATOR POWER ADAPTER BC 430 ANALYZER INPUT BLOCK LO HI NORMAL Figure 3-4. Checking the N (Neutral) voltage input Fluke 433, 434, 435 Service Manual 3-10 3.8.2 Verification of voltage inputs in 120 V range Accuracy Check of Channel A/L1, B/L2, C/L3: 1. Set t

21、he Analyzer to 120 V, 60 Hz nominal (SETUP menu, arrow and ENTER keys). Press MENU, select Volts/Amps/Hertz, press F5 - OK. 2. To check the A/L1, B/L2, C/L3 inputs: connect the N input to Ground (See Figure 3- 3). 3. Set the Calibrator to 60 Hz, 60 V and then to OPR. 4. Fluke 433/434: check for a vo

22、ltage readout V rms between 59.4 . 60.6 V. Fluke 435: check for a voltage readout V rms between 59.88 . 60.12 V. 5. Check the 120 V range according to the table below. Table 3-2. Accuracy Check of Voltage Channels A/L1, B/L2, C/L3 Set Calibrator to Readout at Voltage Channels 60 V, 60 Hz, OPR Fluke

23、433/434: 59.4 . 60.6 V Fluke 435: 59.88 . 60.12 V 120 V, 60 Hz, OPR Fluke 433/434: 119.4 . 120.6 V Fluke 435: 119.88 . 120.12 V 240 V, 60 Hz, OPR Fluke 433/434: 239.4 . 240.6 V Fluke 435: 239.88 . 240.12 V 6. Set the Calibrator to STBY. Optional Test. Bandwidth Check of Channel A/L1, B/L2, C/L3 (*):

24、 7. Set the Calibrator to 120 V, 60 Hz and OPR. 8. Now check the voltage channels one by one. Use the Calibrators field edit function (5500A: FIELD EDIT key, 5700A: AMPL/FREQ key) to adjust the Calibrator to an Analyzer readout of 120.0 V / 120.00 V. 9. Increase the frequency to 3 kHz. Fluke 433/434

25、: check for a readout of 113.8 V or more. Fluke 435: check for a readout of 114.00 V or more. Check the channels according to the table below. Table 3-3. Bandwidth Check of Voltage Channels A/L1, B/L2, C/L3 Voltage Channel to be verified Readout at 60 Hz (adjust Calibrator with EDIT FIELD) Readout a

26、t 3 kHz A/L1 Fluke 433/434: 120.0 V Fluke 435: 120.00 V Fluke 433/434: 113.8 V or more Fluke 435: 114.00 V or more B/L2 Fluke 433/434: 120.0 V Fluke 435: 120.00 V Fluke 433/434: 113.8 V or more Fluke 435: 114.00 V or more C/L3 Fluke 433/434: 120.0 V Fluke 435: 120.00 V Fluke 433/434: 113.8 V or more

27、 Fluke 435: 114.00 V or more 10. Set the Calibrator to STBY. Performance Verification 3.8 Voltage Inputs 3 3-11 Accuracy Check of Channel N (Neutral): 11. Connect the N input with the C/L3 input (See Figure 3-4). A/L1, B/L2, C/L3 now give zero reading. 12. Set the Calibrator to 60 Hz / 60 V and then

28、 to OPR. 13. Fluke 433/434: check for a voltage readout V rms between 59.4 . 60.6 V. Fluke 435: check for a voltage readout V rms between 59.88 . 60.12 V. 14. Check the 120 V range according to the table below. Table 3-4. Accuracy Check of Voltage Channel N (Neutral) Set Calibrator to Readout at Vol

29、tage Channel N 60 V, 60 Hz, OPR Fluke 433/434: 59.4 . 60.6 V Fluke 435: 59.88 . 60.12 V 120 V, 60 Hz, OPR Fluke 433/434: 119.4 . 120.6 V Fluke 435: 119.88 . 120.12 V 240 V, 60 Hz, OPR Fluke 433/434: 239.4 . 240.6 V Fluke 435: 239.88 . 240.12 V 15. Set the Calibrator to STBY. Optional test. Bandwidth

30、 Check of Channel N (*): 16. Set the Calibrator to 120 V, 60 Hz and OPR. 17. Now check the voltage channel N. Use the Calibrators field edit function (5500A: FIELD EDIT key, 5700A: AMPL/FREQ key) to adjust the Calibrator to an Analyzer readout of 120.0 V / 120.00 V. 18. Increase the frequency to 3 k

31、Hz. Fluke 433/434: check for a readout of 113.8 V or more. Fluke 435: check for a readout of 114.00 V or more. 19. Set the Calibrator to STBY. 3.8.3 Verification of voltage inputs in 230 V range Accuracy Check of Channel A/L1, B/L2, C/L3: 1. Set the Analyzer to 230 V, 50 Hz nominal (SETUP menu, arro

32、w and ENTER keys). Press MENU, select Volts/Amps/Hertz, press F5 - OK. 2. To check the A/L1, B/L2, C/L3 inputs: connect the N input to Ground (See Figure 3- 3). 3. Set the Calibrator to 50 Hz, 115 V and then to OPR. 4. Fluke 433/434: check for a voltage readout V rms between 113.8 . 116.2 V. Fluke 4

33、35: check for a voltage readout V rms between 114.77 . 115.23 V. 5. Check the 230 V range according to the table below. Fluke 433, 434, 435 Service Manual 3-12 Table 3-5. Accuracy Check of Voltage Channels A/L1, B/L2, C/L3 Set Calibrator to Readout at Voltage Channels 115 V, 50 Hz, OPR Fluke 433/434

34、: 113.8 . 116.2 V Fluke 435: 114.77 . 115.23 V 230 V, 50 Hz, OPR Fluke 433/434: 228.8 . 231.2 V Fluke 435: 229.77 . 230.23 V 460 V, 50 Hz, OPR Fluke 433/434: 458.8 . 461.2 V Fluke 435: 459.77 . 460.23 6. Set the Calibrator to STBY. Optional Test. Bandwidth Check of Channel A/L1, B/L2, C/L3 (*): Note

35、: for Fluke 435 you must use the Fluke 5700A Calibrator. This instrument is able to generate 219.99999 V as a maximum at 3 kHz. Therefor a 50 Hz reference voltage of 215 V must be adjusted. 7. Set the Calibrator to 230 V (5700A: 215 V), 50 Hz and OPR 8. Now check the voltage channels one by one. Use

36、 the Calibrators field edit function (5500A: FIELD EDIT key, 5700A: AMPL/FREQ key) to adjust the Calibrator to an Analyzer readout of 230.0 V / 215.00 V 9. Increase the frequency to 3 kHz. Fluke 433/434: check for a readout of 218.3 V or more. Fluke 435: check for a readout of 204.25 V or more. Chec

37、k the channels according to the table below. Table 3-6. Bandwidth Check of Voltage Channels A/L1, B/L2, C/L3 Voltage Channel to be verified Readout at 50 Hz (adjust Calibrator with EDIT FIELD) Readout at 3 kHz A/L1 Fluke 433/434: 230.0 V Fluke 435: 215.00 V Fluke 433/434: 218.3 V or more Fluke 435:

38、204.25 V or more B/L2 Fluke 433/434: 230.0 V Fluke 435: 215.00 V Fluke 433/434: 218.3 V or more Fluke 435: 204.25 V or more C/L3 Fluke 433/434: 230.0 V Fluke 435: 215.00 V Fluke 433/434: 218.3 V or more Fluke 435: 204.25 V or more 10. Set the Calibrator to STBY. Accuracy Check of Channel N (Neutral)

39、: 11. Connect the N input with the C/L3 input (See Figure 3-4). A/L1, B/L2, C/L3 now give zero reading. 12. Set the Calibrator to 50 Hz / 115 V and then to OPR. 13. Fluke 433/434: check for a voltage readout V rms between 113.8 . 116.2 V. Fluke 435: check for a voltage readout V rms between 114.77 .

40、 115.23 V. Performance Verification 3.8 Voltage Inputs 3 3-13 14. Check the 230 V range according to the table below. Table 3-7. Accuracy Check of Voltage Channel N (Neutral) Set Calibrator to Readout at Voltage Channel N 115 V, 50 Hz, OPR Fluke 433/434: 113.8 . 116.2 V Fluke 435: 114.77 . 115.23 V

41、230 V, 50 Hz, OPR Fluke 433/434: 228.8 . 231.2 V Fluke 435: 229.77 . 230.23 V 460 V, 50 Hz, OPR Fluke 433/434: 458.8 . 461.2 V Fluke 435: 459.77 . 460.23 V 15. Set the Calibrator to STBY. Optional Test. Bandwidth Check of Channel N (*): Note: for Fluke 435 you must use the Fluke 5700A Calibrator. Th

42、is instrument is able to generate 219.99999 V as a maximum at 3 kHz. Therefor a 50 Hz reference voltage of 215 V must be adjusted. 16. Set the Calibrator to 230 V (5700A: 215 V), 50 Hz and OPR. 17. Now check the voltage channel N. Use the Calibrators field edit function (5500A: FIELD EDIT key, 5700A

43、: AMPL/FREQ key) to adjust the Calibrator to an Analyzer readout of 230.0 V / 215.00 V. 18. Increase the frequency to 3 kHz. Fluke 433/434: check for a readout of 218.3 V or more. Fluke 435: check for a readout of 204.25 V or more. 19. Set the Calibrator to STBY. 3.8.4 Verification of voltage inputs

44、 in 400 V range Accuracy Check of Channel A/L1, B/L2, C/L3: 1. Set the Analyzer to 400 V, 50 Hz nominal (SETUP menu, arrow and ENTER keys). Press MENU, select Volts/Amps/Hertz, press F5 - OK. 2. To check the A/L1, B/L2, C/L3 inputs: connect the N input to Ground (See Figure 3- 3). 3. Set the Calibra

45、tor to 50 Hz, 200 V and then to OPR. 4. Fluke 433/434: check for a voltage readout V rms between 198.0 . 202.0 V. Fluke 435: check for a voltage readout V rms between 199.60 . 200.40 V. 5. Check the 400 V range according to the table below. Fluke 433, 434, 435 Service Manual 3-14 Table 3-8. Accuracy

46、 Check of Voltage Channels A/L1, B/L2, C/L3 Set Calibrator to Readout at Voltage Channels 200 V, 50 Hz, OPR Fluke 433/434: 198.0 . 202.0 V Fluke 435: 199.60 . 200.40 V 400 V, 50 Hz, OPR Fluke 433/434: 398.0 . 402.0 V Fluke 435: 399.60 . 400.40 V 800 V, 50 Hz, OPR Fluke 433/434: 798.0 . 802.0 V Fluke

47、 435: 799.20 . 800.80 V 6. Set the Calibrator to STBY. Optional Test. Bandwidth Check of Channel A/L1, B/L2, C/L3 (*): Note: for Fluke 435 you must use the Fluke 5700A Calibrator. This instrument is able to generate 219.99999 V as a maximum at 3 kHz. Therefor a 50 Hz reference voltage of 215 V must

48、be adjusted. 7. Set the Calibrator to 400 V (5700A: 215 V), 50 Hz and OPR. 8. Now check the voltage channels one by one. Use the Calibrators field edit function (5500A: FIELD EDIT key, 5700A: AMPL/FREQ key) to adjust the Calibrator to an Analyzer readout of 400.0 V / 215.00 V. 9. Increase the frequency to 3 kHz. Fluke 433/434: check for a readout of 379.8 V or more. Fluke 435: check for a readout of 204.25 V or more. Check the channels according to the table below. Table 3-9. Bandwidth Check of Voltage Channels A/L1, B/L2, C/L3 Voltage Channel to be verified Readout at 5