1、PN 600616 August 1995 Rev.7, 7/08 1995-2008 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies. Instruction Sheet Model 5500A/COIL 50-Turn Current Coil Introduction The 5500A/COIL is a 50-Turn Current Coil used as a tool for calibrati

2、ng clamp-type current meters that operate by two different principlesas current transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. How- ever, by using the 50 turns of the 5500A/COIL in conjunction wit

3、h a current source calibrator one can ef- fectively multiply the current of the current source calibrator by a factor of 50 to support the calibration and verification of these clamp-type current meters. PPWARNING UNFUSED. Exceeding current and voltage ratings can cause burn or fire hazard. Using th

4、e Coil Clamp-type current meters operate as current transformers, with differing degrees of magnetic coupling between primary and secondary that vary from meter to meter. The position of the clamp meter with re- spect to the cable also affects the magnetic coupling between primary and secondary of t

5、he current trans- former, which causes variation in reading of the current meter. This is important to under- stand in order to make the most accurate and repeatable measurements. The base of the 5500A/COIL was designed so the current clamp can be centered carefully on the coil, minimizing operator

6、error for best repeatabil- ity. Calibration accuracy to specifications is guaranteed only when proper clamp alignment is made. The clamp current meter should be centered as much as possible on the base dur- ing calibration and verification. If the clamp- type current meter has alignment marks, the a

7、lignment marks should align the clamp with the center bundled wire of the 50-Turn 5500A/COIL. (See Figure 1). Figure 1. Positioning the Clamp RadioFans.CN 收音机爱 好者资料库 Specifications Number of Turns 50 Maximum Current 11A rms, continuous 20.5A rms, 2 minutes Maximum Duty Cycle Derating 11A, 2 minutes

8、ON, 8 minutes OFF Maximum Voltage 7 V dc or 7 V rms ac Minimum Inner Diameter of Clamp Jaws 2.54 cm 1” Specification for Clamp Meter/Coil interaction Specification for 5500A/COIL Coil Specifications clamp/coil interaction Calibrator Output Effective Current Output Toroidal-type Clamps such as the Fl

9、uke 80i-600 and 80I-1000 other-type of Clamps such as the Fluke 80i-KW, -400, 80i-410, -500, -1010 and 30-series Magnitude Frequency Amp-turns (% of output + Amps) (% of output + Amps) 0.2 - 0.33 A DC 10 - 16.4999 0.25 0.002 0.50 0.02 0.33 - 2.9999 A DC 16.5 - 149.999 0.25 0.015 0.5 0.14 3.0 - 20.5

10、A DC 150 - 1025 0.25 0.05 0.5 0.5 0.2 - 0.33 A 45 Hz to 65 Hz 10 - 16.4999 0.28 0.003 0.56 0.03 0.33 - 2.9999 A 45 Hz to 65 Hz 16.5 - 149.999 0.28 0.025 0.56 0.25 3.0 - 20.5 A 45 Hz to 65 Hz 150 - 1025 0.28 0.09 0.56 0.9 0.2 - 0.33 A 65 Hz to 440 Hz 10 - 16.4999 0.79 0.003 1.00 0.03 0.33 - 2.9999 A

11、65 Hz to 440 Hz 16.5 - 149.999 0.79 0.027 1.00 0.25 3.0 - 20.5 * A 65 Hz to 440 Hz 150 - 1025 0.79 0.1 1.00 0.9 0.2 - 20.5 A Above 440 Hz 10 - 1025 Usable but unspecified *Note: Some calibrators may not have adequate compliance voltage to drive coil in this range. Calculating Total Specification The

12、 total specification of the effective current that the clamp-type meter measures is a function of the clamp/coil interaction and of the current calibrator. To determine the total specification use the following formula: source 2 coil 2 total USS+= Example: Assume we are driving the coil with the Flu

13、ke 5500A at 4A, 60 Hz (the clamp meter will see an effective current of 200A, 60 Hz) and we are calibrating a toroidal-type clamp meter. The Calibrators 1 year specification at 4 Amps is (0.06% + 2 mA), so the effective current in the coil bundle will have a speci- fication of (0.06% + 0.1A). Next,

14、we find the total specification of the calibrator and the coil as a per- centage of the output: Specification of effective calibrator current in coil bundle = (0.06% + 0.1A) = 0.11% Specification due to clamp Meter/Coil Interaction = (0.28% + 0.09A) = 0.325% The RSS of these two specifications determines the total specification of the clamp/source combination: %343. 0%11. 0%325. 0 22 total =+=S RadioFans.CN 收音机爱 好者资料库