1、 Part No. Z1-001-802, IB001697 Mar. 2004 OPERATION MANUAL WITHSTANDING VOLTAGE TESTER TOS5052 DANGER This Tester generates high voltage. Any incorrect handling may cause death. Read Chapter 2 “WARNINGS AND CAUTIONS FOR OPERATION THE TESTER” in this manual to prevent accident. Keep this manual near t

2、he tester for easy access of the operator. RadioFans.CN 收音机爱 好者资料库 Use of Operation Manual Please read through and understand this Operation Manual before operating the product. After reading, always keep the manual nearby so that you may refer to it as needed. When moving the product to another loc

3、ation, be sure to bring the manual as well. If you find any incorrectly arranged or missing pages in this manual, they will be replaced. If the manual it gets lost or soiled, a new Operation Manual can be purchased. In either case, please contact your Kikusui agent, and provide the Part No. given on

4、 cover. This manual has been prepared with the utmost care; however, if you have any questions, or note any errors or omissions, please contact your Kikusui distributor/agent. Reproduction and reprinting of this operation manual, whole or partially, without our permission is prohibited. Both unit sp

5、ecifications and manual contents are subject to change without notice. Copyright1998-2004 Kikusui Electronics Corporation Printed in Japan. RadioFans.CN 收音机爱 好者资料库 RadioFans.CN 收音机爱 好者资料库 TOS5052Introductionxiii 7. A large color display The TOS5052 has a large color VACUUM FLUORESCENT DISPLAY. It is

6、 a wide viewing angle type of display with high intensity, and clearly indicates information in clearly readable large letters and in color annunciators. The indicated information includes test conditions, instrument status, readback current, result of pass/fail judgement, etc., assisting you to con

7、duct your tests accurately and efficiently. 8. Analog/digital voltmeters The TOS5052 is furnished with an analog voltmeter (5% full-scale) and a digital voltmeter (1.5% full-scale). The analog voltmeter serves as an output indicator and the digital voltmeter as a high-precision output voltmeter for

8、accurate and prompt testing. (Neither analog nor digital voltmeters can be used to measure any external voltages that are present outside the TOS5052. An attempt to apply an external voltage to their output terminal would cause fatal damage to the TOS5052.) 9. A digital ammeter The TOS5052 has a dig

9、ital ammeter to measure the current that flows through the DUT (device under test). 10. A window comparator for pass/fail judgement The TOS5052 has a window comparator for pass/fail judgment with reference to both upper (U) and lower (L) criteria (cutoff current). The comparator generates a FAIL sig

10、nal when the measured current that flows through the DUT is greater than the preset upper limit criterion or even when it is less than the preset lower limit criterion. The L FAIL detection function contributes to improve the test reliability by detecting open-circuiting or imperfect contacting of t

11、he test leadwires. Separately for each of U type and L type of fail, the TOS5052 indicates a fail annunciator message on its display and delivers a fail event signal, allowing you to immediately find out the type of the fail. You can preset the upper limit and lower limit currents (cutoff currents)

12、mutually independently of each other (0.1mA to 110mA, AC). 11. A digital timer The timer allows you to preset the period during which the test voltage is to be applied to the DUT. The preset range is 0.3 to 999 seconds. When the timer function is ON, the preset period is decremented and the timer in

13、dicates the remaining period; when it is OFF, time is incremented and the timer indicates the elapsed period. 12. Remote control provision The tester has provisions for remote start/stop control operation. That is, it has a 5-pin DIN connector (for the optional Remote Control Box or High Voltage Tes

14、t Probe) on its front panel and a 14-pin Amphenol connector on its rear panel. The remote control function, together with the status signal function, will help you conduct efficient automatic labor-saving tests. TOS5052SETUP1-5 Secure adequate space around the power plug. Do not insert the power plu

15、g to an outlet where accessibility to the plug is poor. And, do not place objects near the outlet that would result in poor accessibility to the plug. 1.3 Checking AC Line Voltage The AC line requirements of the TOS5052 are as follows. Voltage tolerance range : 90 V to 110 VAC Allowable frequency ra

16、nge : 45 Hz to 65 Hz The TOS5052 might not only malfunction but also develop a mechanical breakdown at AC line voltage outside the above range. Make sure that the TOS5052 is run within the specified AC line voltage range. Do not use the AC power cable that comes with the product as a AC power cable

17、for other equipment. The TOS5052 is available in three models with different AC line requirements. Contact your Kikusui agent. 1. 104 to 125 VAC (110/120VAC model) 2. 194 to 236 VAC (220VAC model) 3. 207 to 250 VAC (230/240VAC model) WARNING The AC line voltage must be adjusted only by the Kikusui-

18、qualified service engineer. Before turning on the TOS5052, make sure that the AC line voltage is set up correctly. The AC line voltage setting (labeled SETTING SUPPLY) is indicated on the rear panel. No mark indicates the default 90/110 VAC setting. The AC line voltage of the TOS5052 is set at the f

19、actory or by a qualified service engineer. If the AC line voltage range is altered, a mark is placed on the left of the voltage label. FUSE 250V STANDARD90V-110V 104V-125V 194V-236V 207V-250V SETTING SUPPLY 15A SLOW 6.3A SLOW Table 1-2 SETTING SUPPLY Label (90/110 VAC Model) 3-4OPERATING PROCEDURETO

20、S5052 2) Initial Setup Procedure 1.-Make sure that the POWER switch is in the OFF position. 2.-Connect the AC input power cable (supplied as an accessory) to the AC LINE connector on the rear panel. 3.-Connect other end of the AC input power cable to an AC line outlet of the correct voltage. 4.-Keep

21、ing the SHIFT key pressed, press the POWER switch. The Vacuum Fluorescent Display will start illuminating, indicating that power has been turned ON. (The DC illuminator turns on but it has nothing to do with the operation of the TOS5052.) 5.-Release both SHIFT key and POWER switch. 6.-Within several

22、 tens seconds from the above, the Version Number and Model Number of the TOS5052 will appear on the display screen. The three digits on the left hand side indicate the version number; those on the light hand side are of the three least-significant digits of the model number. 7.-The initial test setu

23、p data will appear on the screen. If no messages at all appear on the screen when more than 60 seconds has elapsed from the above, repeat the procedure all over again from its beginning. 3.3 Checking the Tester Operation The TOS5052 will generate no output unless you reset protection using the inter

24、lock function for the SIGNAL I/O connector on the rear panel. Run your TOS5052 tentatively using the attached 14-pin Amphenol plug. WARNING Pins 9 and 14 of the 14-pin Amphenol plug are short-circuited to reset protection. Before making a practical withstanding voltage test, reset the interlock func

25、tion. For details, see the section entitled Interlock Function. CAUTION This operation check initializes the setup data. Any existing setup values are lost. TOS5052OPERATING PRINCIPLE5-3 Voltmeters The TOS5052 is furnished with a digital voltmeter (1.5% full-scale) and an analog voltmeter (5% full-s

26、cale). (a)Digital voltmeter This digital voltmeter divides the output voltage with high-precision high- voltage resistors and supplies the output to the AC/DC converter circuit where DC voltage is generated. This DC voltage is sent to the A/D converter and indicate the display during the test. (b)An

27、alog voltmeter The analog voltmeter rectifies the current that flows through the divider resistors to drive the DC ammeter. Current detector circuit The current detector circuit converts the current flowing over the output terminal to develop a voltage across the reference resistor. This voltage is

28、converted in the AC/ DC converter circuit to a DC voltage. The digital ammeter takes in this voltage using its A/D converter and indicate it on the display. The ammeter range and resolution are determined by the UPPER value. Upper cutoff current0.1-7.5mA7.6-30mA31-110mA Resolution4A16A64A i10mA.mA.m

29、A.mA 10mAi31mA.mA.mA 31mAi100mA.mA 100mAimA Current Display Format Table 5-1 Relationship between the Ammeter Ranges/Resolutions and Upper cutoff current Pass/fail judgment circuit The DC voltage from the current detector circuit is compared against the reference voltage that corresponds to the pred

30、efined upper or lower cutoff current. When the current proves to be greater than the upper cutoff current, the judgment circuit flags a withstanding voltage error and shuts off the test voltage. It also signals a broken high-voltage test leadwire or poor connection condition and shuts off the test v

31、oltage when the current proves to be smaller than the lower cutoff current. Control circuit The control circuit contains a 8-bit single-chip CPU that provides voltage and current measurement control, pass/fail judgment, test time and test voltage rise time management, and sequence control. It saves

32、the test voltage, cutoff current values, and test time in EEPROM before the CPU is reset following the detection of the POWER switch off condition. 7-2SPECIFICATIONSTOS5052 7.1 Specifications Item Output block Output voltage range0.50 kV to 5.00 kVAC (100 mA output possible range) Voltage setting ra

33、nge0.00 to 2.95 kV/0.00 to 5.45 kV, 2 ranges (3-digit digital setting) Setting accuracy(2% of setting + 2 digits) at 0.20 kV or higher with no load Resolution10V Maximum rated output (*1)500VA (5kV/100mA) Transformer capacity500VA Sine wave Short-circuit current200 mA to 300 mA (when the output volt

34、age is 1.1 kV or more) Distortion factor Frequency50 or 60 Hz selectable (0.5% of setting, except during voltage rise) Voltage regulation9% or less (maximum rated load to no load) Output typePWM switching Output voltmeter AnalogScale5kV f.s Accuracy5% f.s IndicationMean-value response/rms-value Scal

35、e2.5kV/ 5kV f.s DigitalAccuracy ResponseMean-value response/rms-value (400 ms response time) Hold function Ammeter DigitalMeasuring range0.00 to 110mA Accuracy ResponseMean-value response/rms-value (400 ms response time) Hold function Specification The voltage measured at the end of test is held dur

36、ing the PASS or FAIL interval. (5% of upper cutoff current + 20 A) when the measured current does not change within the digital ammeters response time. The current measured at the end of test is held during the PASS interval. Output voltage waveform (*2) Output voltage monitoring function Output is

37、shut off and protection is effected when the output voltage exceeds the set value plus 200V. kV blinks when the output voltage falls below the set voltage minus 100V. Output voltage of 0.5 kV or higher: 2% or less (under no load or resistive load) (1.5% f.s) when the measured voltage does not change

38、 within the digital voltmeters response time. TOS5052SPECIFICATIONS7-5 ItemSpecification REMOTE5-pin DIN connector on the front panel Start/stop control is achieved remotely using options. Remote controller: RC01-TOS, RC02-TOS High-voltage test probe: HP01A-TOS, HP02A-TOS (Valid only when the test v

39、oltage is 4 kVAC.) SIGNAL I/O(*4)14-pin Amphenol connector on the rear panel Output signal control pin Nos. 2 to 8 Indicate the state of the TOS5052. Input signal control pin Nos. 9 to 12 No.Signal NameDescription 1N.CNot used 2H.V ON Output type 3TEST ( 4.5 to 30 VDC) 4PASS 5U FAIL 400 mA (TOTAL) 6

40、L FAIL Isolation type 7READYHeld on in the standby state. 8PROTECTION 11 V to 15 V 10 RR STARTStart signal input pin (low active) 0 V to 4 V 11 RR STOPStop signal input pin (low active)-5 mA max. Input time width5 ms min. 12 RR ENABLEIsolation type 13 ISOL COMCircuit common terminal 14 ISOL COMCircu

41、it common terminal Shut off the output in conjunction with an external device via the INTERLOCK function. Turns on and off the output using a make contact such as a relay or switch. 30 VDC Remote control enable signal input pin (low active) Held on when a current greater than the upper cutoff curren

42、t is detected and FAIL is judged. When the interlock signal is input, the TOS5052 enters the PROTECTION state and cannot start a test. Held on when the PROTECTION function is active. Photocoupler (30 VDC/30 VAC rms max.) Maximum output current High level input voltage Low level input current Photoco

43、upler (30 VDC/30 VAC rms max.) 9INTERLOCK Low level input voltage Set and held on for approx. 0.2 s when PASS is judged. Stays on in the PASS HOLD mode. Held on when a current smaller than the lower cutoff current is detected and FAIL is judged. Held on while a high voltage is being applied across t

44、he output terminals during the test. Held on during the test (except when the voltage is rising). Open collector output Approx. 1.1 V (at 25 C) Output saturation voltage Output withstanding voltage 1234567 1413 1211 1098 SIGNAL I/O READY L FAIL U FAIL PASS TEST H. V ON N. C PROTECTION INTERLOCK RR S

45、TART RR STOP RR ENABLE ISOL COM ISOL COM 7-6SPECIFICATIONSTOS5052 ItemSpecification STATUS SIGNAL OUTPUT 1. H.V ON 5. L FAILOutput type 2. TEST 6. READY 3. PASS7. PROTECTIONOutput voltageApprox. 100 V AC 4. U FAIL 8. POWER ONLeak current1 mA or less. 0.3A Isolation Triode AC switch output 100 VAC is

46、 generated when the TOS505 enters the selected state (100 VAC is generated even the AC line input range was changed). 1200 V AC, 1 second The following modes are selected using the DIP switches on the rear panel (logical sum is taken when two or more modes are selected): Maximum output current *4 SI

47、GNAL I/O The circuit common pins are shared by the signal output and input circuits. The input terminals are pulled up to +15V with resistors. To make the input terminals open is equivalent to that the high level signals are applied to the terminals. TOS5052SPECIFICATIONS7-7 Item TEST MODE 1. DOUBLE

48、 ACTION 2. PASS HOLD The TOS5052 retains the PASS state. 3. MOMENTARY 4. FAIL MODE 5. ON: 60 Hz/OFF: 50 Hz Switches the test voltage frequency. ON: 60 Hz OFF: 50 Hz Environment Temperature5 C to 35 C Humidity20 % to 80 % RH (non condensing) Temperature0 C to 40 C Humidity20 % to 80 % RH (non condens

49、ing) Temperature -20 C to 70 Humidity90 % RH or less (non condensing) Power requirement Allowable voltage range90 V to 110 VThe following power voltage options are factory options: (104 V to 125 V)(194 V to 236 V)(207 V to 250 V) No load time150 VA or less (READY) 1000 VA max. Allowable frequency range45 Hz to 65 Hz Insulation resistance30 M min. (500 VDC), between AC line and chassis Withstanding voltage1390 V AC (2 seconds), between AC line and chas