《leader_lbo-516_delayed_time-base_100_mhz_oscilloscope_calibration_and_maintenance 电路图.pdf》由会员分享,可在线阅读,更多相关《leader_lbo-516_delayed_time-base_100_mhz_oscilloscope_calibration_and_maintenance 电路图.pdf(51页珍藏版)》请在收音机爱好者资料库上搜索。
1、 ENJOY YOUR BOOKS PLEASE VISIT OUR STORE FOR EVEN MORE GREAT STUFF! WWW.EVERYTHING4LESSSTORE.COM COPYRIGHT NOTICE ALL MATERIALS INCLUDING CD/DVD AND PDF FILES ARE COPYRIGHTED WWW.EVERYTHING4LESSSTORE.COM VON WALTHOUR PRODUCTIONS AND MAY NOT BE REPRODUCED, COPIED OR RESOLD UNDER ANY CIRCUMSTANCES. YO
2、U MAY HOWEVER MAKE A COPY FOR YOUR OWN PERSONAL BACKUP. MATERIALS ARE FOR PERSONAL USE ONLY. IF YOU PURCHASED THIS FROM ANYWHERE BUT FROM US PLEASE NOTIFY US IMMEDIATELY SO THAT WE MAY CHECK IF YOU PURCHASED FROM AN AUTHORIZED RESELLER SO WE CAN LET YOU KNOW IF YOU NEED TO RETURN FOR FULL REFUND FRO
3、M AN UNAUTHORIZED SELLER. THANKS AGAIN AND PLEASE TAKE THE TIME TO VISIT OUR STORE. ATTENTION! EVERYTHING ON SALE NOW! RadioFans.CN 收音机爱 好者资料库 RadioFans.CN 收音机爱 好者资料库 WARNING! THE SERVICING INSTRUCTIONS CONTAINED IN THIS MANUAL ARE FOR USE BY QUALIFIED PERSONNEL ONLY. TO AVOID ELECTRIC SHOCK, DO NOT
4、 PERFORM ANY SERVICING OTHER THAN THAT CONTAINED IN THE OPERATING INSTRUCTIONS UNLESS YOU ARE QUALIFIED TO DO SO. RadioFans.CN 收音机爱 好者资料库 TABLE OF CONTENTS (Continued) 3-4 CALIBRATION OUTPUT ADJUSTMENT .30 3-5 A TIMEBASE ALIGNMENT .30 3-5-1 Slow Sweep Time .30 3-5-2 Sweep Length .30 3-5-3 Fast Sweep
5、 Time .30 3-5-4 Sweep Start Point .30 3-5-5 Timebase Accuracy Check .30 3-6 A TIMEBASE X10 MAGNIFIER ADJUSTMENT .30 3-6-1 Magnifier Positioning .30 3-6-2 Magnifier Speed Accuracy .30 3-7 B TIMEBASE ALIGNMENT .30 3-7-1 Slow Sweep Time .30 3-7-2 Length Adjustment .31 3-7-3 Sweep Start Point .31 3-7-4
6、Fast Sweep Time .31 3-7-5 Timebase Accuracy Check .31 3-7-6 Start Points Alignment .31 3-8 VERTICAL AMPLIFIERS .31 3-8-1 DC Balance Adjustment .31 3-8-2 Attenuator Step Balance .30 3-8-3 X1 AC Gain Compensation .31 3-8-4 X10 AC Gain Compensation .31 3-8-5 Gain Calibration .31 3-8-6 CH-2 Invert Balan
7、ce Adjustment .32 3-8-7 CH-1/CH-2 Input Capacitance Adjustment .32 3-8-8 CH-1/CH-2 Input Attenuator Compensation .32 3-8-9 CH-3 Direct Input Capacitance Adjustment .32 3-8-10 CH-3 Attenuator Compensation .32 3-8-11 CH-3 Attenuator Input Capacitance Adjustment .32 3-8-12 CH-3 Gain Adjustment .32 3-8-
8、13 CH-1 Output Level Adjustment .32 3-8-14 CH-1/CH-2 HF Pulse Response Alignment .32 3-8-15 CH-1/CH-2 Frequency Response Check .33 3-8-16 CH-1 Output Pulse Response .33 3-8-17 CH-1/CH-2 X10 Magnifier Bandwidth Check .33 3-8-18 Vertical Position Control Centering .33 3-8-19 ADD Balance Adjustment .33
9、 3-9 TRIGGER CIRCUITRY ADJUSTMENTS .33 3-9-1 Trigger Balance and Centering Alignments .33 3-9-2 Trigger Balance Adjustments for Multitrace Modes .33 3-9-3 Preset Trigger Adjustment .34 3-9-4 CH-3 Pulse Response Adjustment .34 3-9-5 Internal Trigger-Pulse Response Adjustment .34 3-10 X-Y MODE ADJUSTM
10、ENTS .34 3-10-1 Gain Adjustment .34 3-10-2 Balance Adjustment .34 4. REPLACEMENT PARTS LIST 37 5. BLOCK DIAGRAM, SCHEMATICS AND P.C. BOARDS 51 3. MAINTENANCE AND CALIBRATION 3-0. The following test equipment is required to perform the calibration/maintenance procedures described in this section. The
11、 performance specifications given are the minimum necessary to accurately calibrate the oscilloscope. Description Minimum Specifications 1. Digital Voltmeter DC Voltage Range: 0-200 VDC Accuracy: +0.5% 2. High Voltage DC Voltage Range: 0-2000 VDC Voltmeter Accuracy: -+ 1% 3. Frequency Counter Freque
12、ncy Range: 100 Hz - 200 MHz Overall Accuracy: +0.001% 4. Time Mark Generator Marker Outputs: 2nS - 0.5 S with calibrated + 5% offset adjust Accuracy: -+0.1% 5. Sine Wave Generator Frequency Range: lkHz - 200 kHz Output Level: 15 mV - 5 Vp.p Accuracy: -+.3 dB, 1 MHz - 200 MHz as frequency is changed
13、6. Test Oscilloscope 100 MHz Bandwidth 7. Square Wave Frequency: 1 kHz - 100kHz, -+1% Generator Output Level: I mV - 1.0 Vp-p, -+1% Rise Time: 3 nS 8. Amplitude Calibrator Output: 1 kHz square wave Frequency Accuracy: -+0.25% Output Level: 2 mV - 20 Vp-p 9. Capacitance Meter Range: 0-50 pF Accuracy:
14、 - 1% 3-1. INITIAL SETUP CONDITIONS POWER switch .Pushed in A INTEN control . Centered* B INTEN control . Centered* ILLUM control . Clockwise FOCUS control . Centered* AC/GND/DC switches . DC VOLTS/DIV switches . 20mV VARIABLE controls . Clockwise X10 MAG switches Pushed in Vertical POSITION control
15、s . Centered*. PULL QUAD control . Pushed in PULL TRIPLE control . Pushed in V MODE switches . CH-1 CH-2 INV switch . Out DLY TIME MULT control . Counterclockwise. COUPLING switch . AC SOURCE switch .CH-1 LEVEL control . 0 HOLDOFF control . Detented counter clockwise and pulled out A/B TRACE SEP con
16、trol . Centered SLOPE switch . Out HORIZ DISPLAY switches. . A START switch . Out SWEEP MODE switches . AUTO A TIME/DIV switch . .5 mS B TIME/DIV switch . .1 mS A VARIABLE control . Clockwise Horizontal POSITION control . Centered* * Adjusted afterwards for best viewing. Allow 30 minutes warmup befo
17、re making any adjust- ments. Remove the top and bottom covers to gain access to test points and internal adjustments. 3-2 POWER SUPPLY CHECK AND ADJUSTMENT 3-2-1 -8 Volt Adjustment Connect a digital voltmeter s positive lead to the scope chassis (GND), and the voltmeter s negative lead to TP-5 lo- c
18、ated on PCB T-3153. Adjust VR-1 for -8.0 V. 3-2-2 Power Supply Check Check the voltages listed below by moving the voltmeter s negative lead to chassis ground, and applying the positive lead, in turn, to each of the associated test points on PCB T-3153. Nominal Test Point Voltage Tolerance TP-I +100
19、 +98 VDC to 102VDC TP-2 + 50 49 51 TP-3 + 12 11.75 12.25 TP-4 + 8 7.85 8.15 TP-6 + 5 4.8 5.2 TP-7 + 19 17.5 20.5 3-2-3 High-Voltage Adjustment Turn off the unit under test. Connect the positive lead of a HV voltmeter to chassis ground, and its negative lead to TP-8. Turn the scope on and allow a 2 m
20、inute warmup. Adjust VR-2 on PCB T-3162 for a reading of - 1950 volts. Connect an X10 probe to the test oscilloscope and hold its tip close to the face of the CRT under test. The high-voltage ripple displayed on the test oscilloscope should be less than 0.1Vp-p. 3-3. CRT CONTROL ADJUSTMENTS Make sur
21、e the controls are set according to the initial setup conditions in Paragraph 3-1 before starting the following adjustments. 3-3-1 Intensity Range Adjustment. Center the CH- 1 trace on the CRT with the vertical POSITION control. Set the A INTEN control knob mark to an approximate 45 angle as shown i
22、n Figure 3-1. Adjust VR-I on PCB T-3162 until the trace becomes just barely visible. 3-3-2 Astigmatism Adjustment Connect a sine wave generator to the CH-1 input connec- tor. Set generator frequency and output to produce five or six sine waves. Set output level and POSITION controls for a centered d
23、isplay 6 cm high. (Peaks of the sine waves just touching the graticule lines I cm above and below the bottom and topmost graticule lines.) Adjust A INTENS and FOCUS for a medium-bright, sharp display. Adjust VR- 1 on PCB T- 3157 for optimum overall sharpness. 29 3-4. CALIBRATION OUTPUT ADJUSTMENT Co
24、nnect the test oscilloscope to the CAL connector on the fron panel of the LBO-516. Adjust VR-1 on PCB T-3287 for a CAL output level of 0.5 Vp-p +-1%. The duty cycle of the square wave should be 45-55%. Connect the CAL output to a frequency counter; the fre- quency should be 1000 Hz + - 10%. 3-5. A T
25、IMEBASE ALIGNMENT 3.5-1 Slow Sweep Time Return the oscilloscope to the initial setup conditions. (See Paragraph 3-1). Adjust the time-mark generator for an output of .5 mS and connect it to the CH-1 input connector. Using the horizontal POSITION control, align the first mark with the leftmost vertic
26、al graticule line. Adjust VR-22 on PCB T-3158 to align each subsequent mark with a major vertical graticule line. 3.5-2 Sweep Length With the same conditions as in the previous paragraph, adjust the horizontal POSITION control to align the third mark with the leftmost vertical graticule line. (See F
27、igure 3-2.) Adjust VR-1 on PCB T-3158 so the 13th mark is fully dis- played on the CRT screen. 3.5-3 Fast Sweep Time Return the oscilloscope to the initial setup conditions. (See Paragraph 3-1). Set the A TIME/DIV switch to 0.5 PS. Set the time-mark generator to 0.5/.PS and connect it to the CH- 1 i
28、nput connector. Using the horizontal POSITION control, align the first mark with the leftmost vertical graticule line. (See Figure 3-2.) Adjust VC-22 on PCB T-3158 to align each subsequent mark with a major vertical graticule line. 3.5-4 Sweep Start Point Return the oscilloscope to the initial setup
29、 conditions. (See Paragraph 3-1). Set the A TIME/DIV switch to 0.1 mS. With the horizontal POSITION control, adjust the trace startpoint to the first minor division (0,2 major division). Change the A T1ME/DIV switch to 50/aS and adjust VC-1 on PCB T-3158 so the sweep starts at the leftmost vertical
30、graticule line. 3.5-5 Timebase Accuracy Check Return the oscilloscope to the initial setup conditions. (See Paragraph 3-1). Set the A TIME/DIV switch to 0.5 S. Set the time-mark generator to 0.5 S and connect it to the CH- 1 input connector. Align the first mark with the left-most vertical graticule line. Adjust the generator so that each subsequent mark is aligned with a major vertical graticule line. Repeat for each A TIME/DIV switch setting from.2 S to 0.2PS, verify