1、AP034-OM-E Rev DISSUED: January 2000? ? ? The following specifications are valid for model AP034 probes after the probe has reached operating temperature. This takes 20 minutes with power applied, in an environment with stable ambient temperature. The probe must be operating within the environmental
2、 conditions listed in the General Characteristics section, and must have been calibrated within the past 12 months in an ambient temperature of 23 5 C. NOMINAL CHARACTERISTICS Nominal characteristics describe parameters and attributes that are guaranteed by design, but do not have associated toleran
3、ces. Input Configuration:True Differential (+ and Inputs); with shield Ground connector. Effective Gain1:X1, 101, 201 Input coupling:DC. AC Coupling obtained by installing AC Coupling Adapter. Differential Mode Range:400 mV(1 Attenuation) 4 V(10 Attenuation1) 8 V(100 Attenuation1) Common Mode Range:
4、16 V(1 Attenuation) 42 V(10 Attenuation1) 42 V(20 Attenuation1) Maximum Input Voltage: 42 V either input from ground RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? WARRANTED ELECTRICAL CHARACTERISTICS Warranted characteristics are parameters with guaranteed performance. Unless other
5、wise noted, tests are provided in the Performance Verification Procedure for all warranted specifications. LF Gain Accuracy:2% into 50.0 load2, measured at 1 kHz Common Mode Rejection Ratio3:(probe head grounded, DC coupled without attenuator) 70 Hz 80 dB 1 MHz 40 dB 100 MHz 25 dB Notes: 1 10 and 20
6、 obtained with external plug-on attenuators 2 Output impedance is 50 , intended to drive 50 . Add uncertainty of termination impedance to accuracy. 3 LeCroy measures CMRR with a fixture that connects the probe tip ground to the signal source ground. This method is necessary to obtain a reproducible
7、CMRR measurement. Often, users leave the probe tip ungrounded when measuring high frequency signals. Not grounding the probe tip can actually improve CMRR by allowing some of the common mode signal to be impressed across the entire length of the probe cable instead of from probe tip to probe ground.
8、 The CMRR improvement obtained without grounding the probe tip depends on proximity to probe cable ground; therefore, it is nonreproducible. LeCroy has chosen to use a reproducible method of measurement, rather than obtain a more optimistic measurement. RadioFans.CN 收音机爱 好者资料库 AP034-OM-E Rev DISSUED
9、: January 2000? ? TYPICAL ELECTRICAL CHARACTERISTICS Typical characteristics are parameters with no guaranteed performance. Tests for typical characteristics are not provided in the Performance Verification Procedure. Bandwidth, probe only (3 dB):DC to 1 GHz Output Zero: 3 mV within 30 minutes of au
10、tobalance Residual Autobalance Offset: 100 V referred to input with X1 effective gain Differential Offset Range:1.6 V(1 Attenuation) 16 V(10 Attenuation) 42 V(20 Attenuation) CMRR:See Figure 13. Input Resistance (each side to ground):1 M? Input Capacitance (between inputs): 0.85 pF (See Figure 1.) I
11、nput Capacitance (each side to ground): 1.5 pF (See Figure 1.) Noise (referred to input, 10 to 1000 MHz):35 nV/Hz (1 Attenuation) 350 nV/Hz (10 Attenuation) 700 nV/Hz (20 Attenuation) Output Impedance:50 ? Harmonic Distortion 3rd order distortion:60 dB below fundamental (200 mVp-p output, at 100 MHz
12、) 3rd order intercept:+20 dBm (at 100 MHz at output) AC Coupling LF Cutoff (3 dB):16 Hz (using plug-on AC coupler) RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? Figure 13. Typical CMRR Graphs AP034-OM-E Rev DISSUED: January 2000? ? GENERAL CHARACTERISTICS Temperature:0 to 50 C (ope
13、rating) 40 to 75 C (storage) Input Connectors:compatible with 0.025 in. (0.635 mm) square pins 0.036 in. (0.91 mm) maximum diameter (for round pins) Power Requirements:powered from oscilloscope through ProBus interface or with ADPPS power supply Dimensions: Control Housing: Length:3.625 in.(9.2 cm)
14、Width:1.50 in.(3.8 cm) Height:1.00 in.(2.5 cm) Head:Length:4.0 in.(10.1 cm) w/o Attenuator or AC Coupler Width:2.25 in.(2.25 cm) Height:0.625 in.(1.6 cm) Cable:Length:42 in.(106 cm) Diameter:0.275 in.(7.0 mm) Weight:Probe only:6.4 oz(0.18 kg) Shipping:2 lbs, 8.4 oz(1.15 kg) ?ISSUED: January 2000AP03
15、4-OM-E Rev D ? COMPLIANCE AND CERTIFICATIONS EC Declaration of Conformity:Conforms to EMC Directive 89/335/EEC for electromagnetic emission and immunity requirements. EN 55011:1997:The probe has been tested to verify compliance with this standard, Class B for Conducted and Radiated Emissions. NE 500
16、82-1:1997:The probe has been tested to verify compliance with this standard for ESD, Radiated Immunity, EFT/Burst Immunity, Fast Surge Immunity, Conducted Immunity, and Voltage Sags however, probe offset can only be controlled through the buttons on the probe body. Also, the scale factor will be displayed incorrectly in some modes. If required, contact your local LeCroy representative for information on upgrading the software in your oscilloscope. # # # ?ISSUED: January 2000AP034-OM-E Rev D ? BLANK PAGE