《KIKUSUI TOS5050 Operation 电路图.pdf》由会员分享,可在线阅读,更多相关《KIKUSUI TOS5050 Operation 电路图.pdf(139页珍藏版)》请在收音机爱好者资料库上搜索。
1、WARNINGS Against HIGH VOLTAGE DANGER This Tester generates high voltage. Any incorrect handling may cause death. Read Chapter 3 WARNINGS AND CAUTIONS FOR OPERATING THE TESTER in this manual to prevent accident. This manual should be placed within the reach of the operator so that he or she may read
2、it whenever necessary. OPERATION MANUAL Part No. Z1-000-632 IB000598, MAR/2004 TOS5000 SERIES WITHSTANDING VOLTAGE TESTER TOS5101 TOS5051 TOS5050 RadioFans.CN 收音机爱 好者资料库 All or any parts of this manual may not be reproduced in any forms, without express written permission of Kikusui Electronics Corp
3、oration. The contents of this manual, including the specifications of the instrument, are subject to change without notice. 1993-2004 Copyright Kikusui Electronics Corporation. Printed in Japan. All rights reserved. RadioFans.CN 收音机爱 好者资料库 RadioFans.CN 收音机爱 好者资料库 4.1.1 Turning ON/OFF the AC Input Po
4、wer .4-3 4.1.2 Start/Stop of Test .4-5 4.1.3 Test Voltage .4-9 4.1.4 Cutoff Currents .4-12 4.1.5 Test Time .4-15 4.1.6 Others .4-17 4.2 Display Items.4-18 4.2.1 Test Voltage Display.4-18 4.2.2 Readout for Cutoff Current and Measured Current .4-21 4.2.3 Test Time Display.4-23 4.2.4 READY Message.4-24
5、 4.2.5 TEST Message .4-24 4.2.6 PASS Message .4-24 4.2.7 FAIL Messages .4-24 4.2.8 PROTECTION Message.4-25 4.2.9 Others .4-25 4.3 Description of Rear Panel Items.4-26 5. PREPARATIVE TEST PROCEDURES.5-1 5.1 Initial Setup.5-2 5.1.1 Initial Setup of Switches and Controls.5-2 5.1.2 Initial Test Setup Da
6、ta.5-2 5.1.3 Procedure for the Initial Test Setup Data.5-3 5.2 Checking the Tester Operation.5-4 5.3 Checkout Before Starting Test Operation.5-6 6. TEST PROCEDURES.6-1 6.1 AC Withstanding Voltage Test Procedure.6-2 6.1.1 Selecting an AC Test Voltage Range.6-2 6.1.2 Setting the Upper Cutoff Current.6
7、-3 6.1.3 Setting the Lower Cutoff Current .6-4 6.1.4 Setting the Test Time .6-5 6.1.5 Setting the Test Voltage .6-6 6.1.6 Connecting the DUT .6-7 6.1.7 Executing a Test.6-7 6.2 DC Withstanding Voltage Test Procedure.6-12 6.2.1 Selecting a DC Test Voltage Range.6-12 6.2.2 Setting the Upper Cutoff Cur
8、rent.6-13 6.2.3 Setting the Lower Cutoff Current .6-14 6.2.4 Setting the Test Time .6-16 6.2.5 Setting the Test Voltage .6-17 6.2.6 Connecting the DUT .6-17 6.2.7 Executing a Test.6-18 6.3 Remote Control.6-22 6.3.1 Remote Control through REMOTE CONTROL Connector.6-23 6.3.2 Remote Control through SIG
9、NAL I/O connector.6-24 6.3.3 Interlock Function .6-26 VII 6.4 Output Signals.6-27 6.4.1 Output Signals of SIGNAL I/O Connector.6-27 6.4.2 Output Signal of STATUS SIGNAL OUTPUT receptacle .6-30 6.5 Settings for Special Test Modes.6-31 6.5.1 DOUBLE ACTION Switch .6-31 6.5.2 PASS HOLD Switch.6-31 6.5.3
10、 MOMENTARY Switch .6-32 6.5.4 FAIL MODE Switch.6-32 7. OPERATING PRINCIPLE.7-1 7.1 Block Diagrams.7-2 7.2 Componential Circuits and Devices.7-3 7.3 Zero-turn-on Switch.7-4 7.4 Delay Time for Pass/Fail Judgement in DC Mode.7-5 7.5 Automatic Discharge Function.7-5 8. MAINTENANCE.8-1 8.1 Cleaning the O
11、uter Surfaces.8-2 8.2 Checking the Cord and Leadwires.8-2 8.2.1 Checking the AC Power Cable .8-2 8.2.2 Checking the HV Test Leadwires .8-2 8.2.3 Overhaul.8-3 8.3 Calibration.8-3 9. SPECIFICATIONS.9-1 9.1 TOS5050.9-2 9.2 TOS5101 and TOS5051.9-9 9.3 Overall Dimensions.9-18 9.3.1 Dimensions of Models T
12、OS5050 and TOS5051 .9-18 9.3.2 Dimensions of Model TOS5101 .9-18 10. OPTIONS.10-1 10.1 Model RC01-TOS/RC02-TOS Remote Control Box.10-2 10.2 Model HP01A-TOS/HP02A-TOS High Voltage Test Probe.10-3 10.3 Model PL01-TOS Warning Light Unit.10-4 10.4 Model BZ01-TOS Buzzer Unit.10-4 10.5 High Voltage Test L
13、eadwires.10-4 INDEX.I-1 VIII Chapter 2 GENERAL PRELIMINARY NOTES AND PRECAUTIONS 2-2 2.1 Checking for Shipping Damages and Accessories When you receive the Tester, inspect it for any obvious damages that may have occurred during shipment. Immediately report any damages to the carrier and your Kikusu
14、i agent. You should also find the standard accessories as shown below accompany the Tester in the same package. TOS5101TOS5050/TOS5051 AC power cableHV Test Leadwires 14-pin Amphenol plug (assembly type)DANGER and WEIGHT sticker TL01-TOS TL03-TOS (TOS5101 only) WEIGHTkg (TOS5101 only) Do not use the
15、 AC power cable attached to this product for the AC power cable of other instruments. Chapter 2 GENERAL PRELIMINARY NOTES AND PRECAUTIONS 2-6 2.2.3 Environmental Conditions Be sure to observe the following precautions when installing the tester. Do not use the tester in a flammable atmosphere. To pr
16、event explosion or fire, do not use the tester near alcohol, thinner, or other combustible materials, or in an atmosphere containing such vapors. Avoid locations where the tester is exposed to high temperatures or direct sunlight. Do not locate the tester near a heater or in areas subject to drastic
17、 temperature changes. Operating temperature range:0 C to +40 C Storage temperature range:-20 C to +70 C Avoid humid environments. Do not locate the tester in a high-humidity environment-near a boiler, humidifier, or water supply. Operating humidity range:20 % to 80 % RH (no dew condensation permitte
18、d) Storage humidity range:80 % RH or less (no dew condensation permitted) Condensation may occur even within the operating humidity range. In that case, do not start using the tester until the location is completely dry. Do not place the tester in a corrosive atmosphere. Do not install the tester in
19、 a corrosive atmosphere or one containing sulfuric acid mist or the like. This may cause corrosion of various conductors and imperfect contact with connectors, leading to malfunction and failure, or in the worst case, a fire. Do not locate the tester in a dusty environment. Dirt and dust in the test
20、er may cause electrical shock or fire. Do not use the tester where ventilation is poor. Prepare sufficient space around the tester to allow for air flow. Do not place the tester on a tilted surface or in a location subject to vibrations. If placed on a non-level surface or in a location subject to v
21、ibration, the tester may fall, resulting in damage and injury. Do not use the tester in locations affected by strong magnetic or electric fields. Operation in a location subject to magnetic or electric fields may cause the tester to malfunction, resulting in electrical shock or fire. Chapter 2 GENER
22、AL PRELIMINARY NOTES AND PRECAUTIONS 2-7 Do not use the tester in locations near a sensitive measuring instrument or receiver. Operation in a location subject, may cause such equipment may be affected by noise generated by the tester. At a test voltage exceeding 3 kV, corona discharge may be generat
23、ed to produce substantial amounts of RF broadband emissions between grips on the test leadwire. To minimize this effect, secure a sufficient distance between alligator clips. In addition, keep the alligator clips and test leadwire away from the surfaces of conductors (particularly sharp metal ends).
24、 Secure adequate space around the power plug. Do not insert the power plug to an outlet where accessibility to the plug is poor. And, do not place objects near the outlet that would result in poor accessibility to the plug. Chapter 2 GENERAL PRELIMINARY NOTES AND PRECAUTIONS 2-8 2.2.4 Notes for Movi
25、ng the Tester When moving the tester to the installation site or otherwise transporting it, take the following precautions: Before moving the tester, turn off the power switch. Transporting the tester with its POWER switch on can lead to electric shock and damage. When moving the tester, Disconnect
26、all wires from it. Moving the tester without disconnecting the cables may result in breakage of the wire or injury due to the tester tipping over. For transportation, use the special packing material for the tester. Transport the tester in its original package to prevent vibration and falls, which m
27、ay damage the tester. If you require packing material, contact Kikusui distributor/agent. Have two or more people move the tester. The TOS5101 weighs approximately 21 kg. Two persons are required to carry the TOS5101. Take extreme care when moving the TOS5101. Chapter 3 WARNINGS AND CAUTIONS FOR OPERATING THE TESTER 3-3 3.2 Warnings and Cautions when Operating the Tester 3.2.1 Connecting the LOW Test Leadwire For the connection method of