《lecroy AP034-OM-E-05 电路图.pdf》由会员分享,可在线阅读,更多相关《lecroy AP034-OM-E-05 电路图.pdf(12页珍藏版)》请在收音机爱好者资料库上搜索。
1、AP034-OM-E Rev DISSUED: January 2000? ? ? CLEANING The exterior of the probe and cable should be cleaned only using a soft cloth moistened with water or isopropyl alcohol. Abrasive agents, strong detergents, or other solvents may damage the probe. Always ensure that the input receptacles are free of
2、 debris before connecting accessories. CALIBRATION INTERVAL To guarantee accurate performance, you should have the probe calibrated every 12 months. Avoid exposing the probe to extreme mechanical shock and excessive bending of the cable because these may alter the calibration. SERVICE STRATEGY The A
3、P034 circuits utilize fine pitch surface mount devices; it is, therefore, impractical to attempt repair in the field. Defective probes must be returned to a LeCroy service facility for diagnosis and exchange. A defective probe under warranty will be replaced with a factory refurbished probe. A probe
4、 that is not under warranty can be exchanged for a factory refurbished probe. A modest fee is charged for this service. The defective probe must be returned in order to receive credit for the probe core. TROUBLESHOOTING If the probe is not operating properly, the problem may be the way in which it i
5、s used. Before assuming the probe is defective, perform the following troubleshooting procedures. A. Trace Off Scale This is typically caused by improper offset setting, or by an input signal that exceeds the probe differential or common mode range. Perform the following: 1. Remove the input signal
6、from the probe, return the offset to zero, and Autobalance the probe. Does the trace return to approximately the center of the graticule? If not, proceed to step 7. ! RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? 2. Set the oscilloscope calibrator to output a 100-mV, 1-kHz square w
7、ave. Using the flex lead set, connect the probe + input to the calibrator output signal, leave the input open. Set the oscilloscope to 50 mV/div and 500 s/div. Is the displayed waveform a 100-mV, 1-kHz square wave with the correct polarity? 3. Repeat step 2 with the input connected to the calibrator
8、 and the + input left open. In this case, the displayed waveform should be inverted. 4. Connect both the + input and input to the calibrator output. Is the trace approximately a flat line near zero volts? 5. If steps 1 to 4 give the correct results, the problem is likely a result of the input signal
9、 exceeding the differential or common mode range. 6. Connect both the + input and input to one of the two input signals. If the trace is off scale, the input signal is probably exceeding the common mode range. Repeat with the other input signal. 7. Is a ProBus Power Supply Overload error message dis
10、played? If so, remove all other ProBus accessories from the oscilloscope. Is the message still displayed? If so, remove the AP034. Is the message still displayed? If so, the oscilloscope should be returned for service. 8. If, after removing the AP034, the Power Supply Overload error message is not d
11、isplayed, the problem may be either the probe or the oscilloscope. Repeat the test with a different ProBus accessory. If the message does not return with a different accessory, the AP034 may be defective and should be returned for service. B. Incorrect Frequency Response Possible causes are a defect
12、ive probe or oscilloscope, poor connections, or poor grounding. Try the following: 1. Verify that the BW limiting of the oscilloscope is off. RadioFans.CN 收音机爱 好者资料库 AP034-OM-E Rev DISSUED: January 2000? ? 2. Connect the probe to another oscilloscope. If the probe now measures properly, the problem
13、may be in the oscilloscope. 3. If the probe behaves as if it is ac-coupled at high frequency, check for an open input connection. 4. Poor frequency or transient response and AC gain errors may result when one of the two input connections is open. 5. Excessive “ring” and other transient problems can
14、result from excessive lead length. To test this, shorten the input leads to less than 1 cm. If the transient response changes significantly, the lead parasitics are the cause. C. DC Errors Incorrect DC gain requires recalibration or factory repair. This can be determined by completing the gain check
15、s in the Performance Verification Procedure. 1. Extremely high source resistance will result in DC gain errors. Check the probe accuracy with the oscilloscope calibrator signal. 2. Verify that the probe is not being overdriven into clipping for its current gain setting. 3. Excessive offset can resul
16、t from large changes in ambient temperature. Remove the input signal from the probe and repeat the Autobalance cycle. With the Offset set to zero, did the trace return to the center of the graticule? D. Poor Common Mode Rejection Use the 1-kHz calibrator signal from the oscilloscope to check common
17、mode rejection. With both the + input and input connected to the calibrator signal, a flat line at zero volts should be seen on the graticule. 1. Check the probe with the plug-on attenuator installed and removed. If excessive common mode signal appears only when the attenuator is present, the attenu
18、ator may need to be rematched to the probe. Use the procedure listed in this section to match the attenuator. 2. If the common mode signal appears when the probe is connected to the test circuit, but not when it is attached to RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? the calib
19、rator, the problem may be caused by large mismatches in the source impedance. Try connecting both inputs to one of the input signals in the test circuit, then the other. If the common mode signal disappears, try probing lower impedance points within the circuit. RETURNING A DEFECTIVE PROBE The proce
20、dure for returning a defective probe to be exchanged, is as follows: Contact your local LeCroy sales representative to find out where to return the product. All returned products should be identified by model and serial number. You should describe the defect or failure, and provide your name and con
21、tact number. In the case of products returned to the factory, a Return Authorization Number (RAN) should be used. The RAN can be established by contacting your nearest LeCroy office, or the New York Customer Care Center. Return shipments should be made prepaid. LeCroy cannot accept COD or Collect Re
22、turn shipments. We recommend air- freighting. It is important that the RAN be clearly shown on the outside of the shipping package for prompt redirection to the appropriate LeCroy department. 1. Contact your local LeCroy sales or service representative to obtain a Return Authorization Number. 2. Rem
23、ove all accessories, including all removable cables, from the probe. 3. Pack the probe in its case, surrounded by the original packing material (or equivalent) and box. 4. Label the case with a tag containing: The RAN Name and address of owner Instrument model and serial number Description of the fa
24、ilure mode 5. Package the probe case in a cardboard shipping box with adequate padding to avoid damage in transit. AP034-OM-E Rev DISSUED: January 2000? ? 6. Mark the outside of the box with the shipping address given to you by the LeCroy representative; be sure to add the following: ATTN: FRAGILE 7
25、. Insure the item for the replacement cost of the probe. 8. Ship the package to the appropriate address. REPLACEMENT PARTS The probe connection accessories, and other common parts can be ordered through the regional customer care centers. Refer to Figure 7 for LeCroy Part numbers. Defective probes c
26、an be replaced on an exchange basis. The replacement exchange probe will have been factory repaired, inspected and calibrated to the same standards as a new product. In order to obtain an exchange probe, you must return the defective probe. The returned probe should be sent back to the regional cust
27、omer care center without any accessories, manual, or case. ?ISSUED: January 2000AP034-OM-E Rev D ? Figure 7. Replceable Parts 13 14 12 11 2 5 8 6 7 4 9 10 1 3 AP034-OM-E Rev DISSUED: January 2000? ? ReferenceLeCroy Part NumberDescriptionRevQty. 1529-304-001Plastic probe tip housing TopA1 2377-000-00
28、7LeCroy AP034 Probe Tip LabelA1 3529-304-003Plastic probe tip housing EndA1 4529-304-002Plastic probe tip housing BottomA1 5555-001-000#2 Allen Screw, BlackA1 6554-425-003Screw, M 2.5x6, self tappingA4 7709-3XX-P53ProBus End CapA1 870A-P03-303Termination BoxA1 9337-000-018LeCroy Termination Box Labe
29、l with cutouts A1 10416-090-004Grey Switch CapA4 11377-000-008LeCroy Termination Box Rear Label A1 12PK033Accessory Kit, AP033/34 without AC Coupler and Attenuator A1 13AP03X-FLEX-LEADFlex LeadB1 14AP03X-OFFSET-PINOffset Pin, package of 10B- -AP034-PROBE-FRUAP034 Probe only ExchangeA- -AP034-DA10Plu
30、g-On Attenuator, 10, AP034A1 -AP034-DA20Plug-On Attenuator, 20, AP034A1 -AP03X-AC-COUPLERPlug-On AC Coupler, AP033/AP034 B1 -AP034-OM-EInstruction Manual, AP034, EnglishD1 ?ISSUED: January 2000AP034-OM-E Rev D ? MATCHING PROCEDURE FOR 10 PLUG-ON EXTERNAL ATTENUATOR The 10 and 20 Plug-on attenuators
31、provided as standard accessories with the AP034 are calibrated to match the specific probe they are shipped with. Individual probes will have small variations in parasitic capacitance within the input circuits. To obtain maximum common mode rejection performance, the attenuators are calibrated to ma
32、tch a specific probe during the manufacturing process. In order to preserve the maximum Common Mode Rejection, do not interchange external attenuators between probes. The Plug-on AC coupling adapter is not matched to a specific probe and, therefore, does not need to be matched when interchanged. If
33、the 10 or 20 Plug-on attenuators become accidentally mixed between probes, you can use the procedure listed below to restore the compensation match. This adjustment does not affect any of the parameters in the warranted specifications. Therefore, the required test equipment does not need to be calib
34、rated. Note The AP033 and AP034 Active Differential Probes have different input capacitance. The 10 Plug-on attenuator supplied with model AP033 cannot be properly adjusted for use with model AP034. Make sure that the attenuator is marked “AP034” before attempting this procedure. EQUIPMENT REQUIRED
35、Test Oscilloscope The oscilloscope must support ProBus. Otherwise use a non- ProBus oscilloscope and ADPPS power supply. Signal Source Low frequency square wave: frequency from 50 Hz to 5 kHz, amplitude from 1 V to 10 V. The output waveform must have a square corner and flat top with minimum oversho
36、ot suitable for adjusting compensation. The generator should have trigger output, or use a BNC Tee connector and separate BNC cable AP034-OM-E Rev DISSUED: January 2000? ? from the output to provide the trigger signal for the test oscilloscope. Interconnect Cable This is for connecting the output of
37、 the signal source to the probe. A BNC cable and a pair of small alligator clips or “lead grabber” adapter (Pomona #3788) may be used. 0.025 in. (0.635 mm) Square Pins (3 required). The pins from the header supplied in the probe accessory kit are suitable. Tools Flat bladed screwdriver, 0.040 in. (1
38、 mm) wide Adjustment Tool: 0.025 in. (0.635 mm) square head Note You can fabricate the Adjustment Tool by flattening the end of a 0.025 in. (0.635 mm) square pin with a file. Then, using heat from a soldering iron, insert the pin into a short length of rigid plastic tubing to serve as a handle. PROC
39、EDURE 1. Attach the AP034 to the test oscilloscope. If the test oscilloscope is not equipped with ProBus, use the ADPPS to provide power for the AP034. 2. Attach the 10 or 20 Attenuator Adapter to the AP034 probe tip. 3. Insert 0.025 in. (0.635 mm) square pins into the +, , and input connectors of t
40、he Plug-on Attenuator. 4. Attach the interconnect cable to the output of the signal source. 5. Attach the Trigger Out signal from the signal source to the External Trigger Input of the test oscilloscope. If the signal source does not have a separate Trigger Out signal, use a BNC Tee connector in the
41、 output. Run one cable to the External Trigger Input of the test oscilloscope. Connect the other to the probe inputs. ?ISSUED: January 2000AP034-OM-E Rev D ? 6. Using the alligator clips on the end of the interconnect cable, connect the signal source ground to the square pin on the Attenuator Adapte
42、rs “ input. Attach the signal output to the square pin on the Attenuator Adapters “+” input. 7. Turn on the test oscilloscope. The EFFECTIVE GAIN indicator for 10 or 20 should be lit. (NOTE: If the X1 EFFECTIVE GAIN indicator is lit instead of the 10 or 20 indicator, make sure that the plug-on adapt
43、er installed on the probe tip is the Attenuator, and not the AC Coupler.) 8. Set the test oscilloscope Volts/Div to 200 mV (for LeCroy oscilloscopes with ProBus) or 20 mV/Div when using the ADPPS with an oscilloscope without scale factor correction; 5 s/Div; AUTO trigger mode; Trigger source: Extern
44、al. Set the Bandwidth Limiting to 2030 MHz and Average the waveform 1:31 to reduce noise. 9. Turn on the signal source. Set the output frequency to approximately 5 kHz. Set the output amplitude to approximately 1 V for matching the 10 Attenuator Adapter, or 2 V for matching the 20 Attenuator Adapter
45、. 10. Adjust the test oscilloscope trigger level for a stable trace. If necessary, use the probe offset to position the waveform to show the square corner of the test signal. NOTE: Do not use the oscilloscope offset or position controls to reposition the trace. Many of the signal generators used for
46、 compensation calibration only have square corners on one of the two edges of the output waveform. (Rising or falling edge, but not both.) Be sure to display the correct edge for this step. 11. Using the square adjustment tool, adjust the +LF Comp (See Figure 8) to achieve the best square corner and
47、 flat top of the displayed waveform. Note that the added capacitance of the adjustment tool may change the compensation of the waveform when it is inserted. The correct adjustment is achieved when the best corner is displayed with the adjustment tool removed. AP034-OM-E Rev DISSUED: January 2000? ?
48、Figure 8. Attenuator Adjustment Locations 12. Move the connections on the interconnect cable so the signal generator ground is connected to the square pin inserted in the ground connector of the Attenuator Adapter. Connect the signal generator output to BOTH the + input and input square pins in the
49、Attenuator Adapter. 13. Set the signal generator frequency to about 50 Hz, and the output amplitude to about 10 V. 14. Set the test oscilloscope Volts/Div to 20 mV (for LeCroy oscilloscopes with ProBus) or 2 mV/Div when using the ADPPS with an oscilloscope without scale factor correction; and set the time scale to 2 ms/Div. I