1、 Second Edition January 2008 Audio Analyzers UP 300/UP 350 10 Hz to 80 kHz The new product family from Rohde can also be superimposed on the sinewave signals Audio Analyzers UP 300/UP 35012 5 Diverse high-quality test signals The generators in the Audio Analyzers UP 300 and UP 350 set new stan- dard

2、s in the lower price segment. By providing a wide variety of sinewave signals, two-tone and multitone signals, bursts and noise, the instruments offer the ideal test signals for measurements in the lab, in service and in production, as well as in uni- versity education. Because the audio analyzers h

3、ave inherent distortion of less than 90 dB, even high-end audio devices can be measured. High-end measurement characteristics The audio analyzers offer bandwidths of up to 80 kHz, enabling the user to perform measurements even on broadband audio equipment. The UP 350 is capable of sampling rates up

4、to 192 kHz unprecedented in this class of instruments. Extensive selection of filters The UP 300 and UP 350 contain a wide variety of weighting, third-octave and octave filters. As many as three filters can be combined. Powerful FFT analysis The FFT analysis capability of the UP 300/350 also sets ne

5、w standards in this class of instruments. This capability supports up to 16 k points and provides numer- ous window functions, which enables it to display the spectral composition of sig- nals up to 80 kHz in bandwidth. 6Audio Analyzers UP 300/UP 350 The new instrument family equipped for the future

6、 Versatile applications Desktop use Portable for mobile use Integration into 19-inch racks 7Audio Analyzers UP 300/UP 350 USB interfaces The USB host interface links the instruments to the PC world. The bus ensures high data transmission rates at low cost. Other peripherals (e.g. printers) can be ad

7、dressed via another USB interface. Identical housing All instruments based on the Family 300 concept have an almost identical “face”, a 5.4-inch TFT display, front- panel control elements, protective guards and a handle that can be adjusted to different positions. Only the connectors on the front an

8、d rear panel vary depending on the instrument type. If the protective guards and the handle are removed, the UP 300/350 can be installed in a 19-inch rack. Owing to their slim design, two instruments of the Family 300 can be placed next to each other. In addition to its analog interfaces, the UP 350

9、 has digital BNC interfaces in consumer and professional format on its rear panel. UP 300 UP 350 Audio Analyzers UP 300/UP 3508 Analyzer Analog audio inputs Digital audio inputs (model UP 350 only) Frequency rangeDC/10 Hz to 80 kHz Frequency response (referenced to 1 kHz) 10 Hz to 20 Hz 20 Hz to 22

10、kHz 22 kHz to 40 kHz 40 kHz to 80 kHz 0.1 dB 0.05 dB 0.1 dB 0.25 dB BNC connectors2 channels, floating or grounded (from serial no 100050), selectable AC/DC coupling, channel 1 on front panel, channel 2 on rear panel Maximum input voltagerms, sinewave33 V Measurement rangesin steps of 6 dB390 mV to

11、50 V (max. input 33 V) Input impedanceinner/outer conductor to ground100 k Crosstalk attenuationfrequency 100 dB Common-mode rejectionat 50 Hz, Vin 3 V at 1 kHz, Vin 3 V at 16 kHz, Vin 80 dB 75 dB 60 dB Generator outputeach input channel switchable to the other generator output channel BNC connector

12、unbalanced, grounded, on rear panel Impedance75 Input level (Vpp)100 mV to 5 V Optical inputTOSLINK Channels1, 2, or both Audio bits16 to 24 Sampling rate32 kHz, 44.1 kHz, 48 kHz, 96 kHz, 192 kHz Formatprofessional and consumer Specifications Important: We continously refine our products. Please che

13、ck our homepage www.up300.rohde- for new applications and features. Specifications apply under the following conditions: specified environmental conditions met, calibration cycle adhered to and total calibration performed. 9Audio Analyzers UP 300/UP 350 Measurement functions RMS value, wideband Erro

14、r limitsmeasurement speed AUTO, at 1 kHz sine, AC coupling additional error with measurement speed AUTO FAST additional error with DC coupling 0.1 dB 0.1 dB 0.1 % of measurement range Integration timeAUTO FAST/AUTO VALUE 5 ms/50 ms, at least 1 cycle 1 ms to 10 s Noisewith A filter, 600 source impeda

15、nce with CCIR unweighting filter, 600 source impedance 2 V 4 V Filtersweighting filters and predefined octave and third-octave filters; up to 3 filters can be combined RMS value, selective Error limits0.2 dB Bandwidth (0.1 dB)fixed bandwidth filters3 Hz, 10 Hz, 30 Hz, 100 Hz or 300 Hz Bandwidth (3 d

16、B)relative bandwidth absolute bandwidth 1 %, 3 %, 1/12 octave, 1/3 octave, value (10 Hz minimum) 10 Hz to fmax/5 Selectivity100 dB Frequency settingfixed through entered value or automatic to input signal Peak value Measurementpos. peak, neg. peak, peak-to-peak, absolute peak Error limitsat 1 kHz0.2

17、 dB Interval20 ms to 10 s Filtersweighting filters and predefined octave and third-octave filters; up to 3 filters can be combined Quasi-peak Measurementin accordance with CCIR 468-4 Error limitsanalyzer bandwidth 22 kHzin accordance with CCIR 468-4 Noisewith CCIR weighting filter, 600 source impeda

18、nce12 V Filtersweighting filters and predefined octave and third-octave filters; up to 3 filters can be combined DC voltage Voltage range0 V to 33 V Error limits (1 % of measured value + 0.5 % of measure- ment range) Total harmonic distortion (THD) Fundamental20 Hz to 20 kHz Frequency tuningfixed th

19、rough entered value, auto-tuning to input signal Weighted harmonicsup to 80 kHzany combination of d2 to d9 Error limitsharmonics 50 kHz harmonics 80 kHz 0.7 dB 1 dB Inherent distortionfundamental 1 kHz fundamental 20 Hz to 5 kHz fundamental 5 kHz to 15 kHz fundamental 15 kHz to 20 kHz 100 dB 90 dB 8

20、5 dB 80 dB Spectrumbargraph showing signal and distortion Audio Analyzers UP300/UP35010 Measurement functions THD+N and SINAD Fundamental20 Hz to 20 kHz Frequency tuningfixed through entered value, auto-tuning to input signal Bandwidthweighting filters and predefined octave and third-octave filters;

21、 up to 3 filters can be combined Error limitsbandwidth 22 kHz bandwidth 80 kHz 0.8 dB 1.4 dB Inherent distortionbandwidth 20 Hz to 22 kHz, fundamental 1 kHz bandwidth 20 Hz to 22 kHz, fundamental 20 Hz to 5 kHz bandwidth 20 Hz to 80 kHz, fundamental 20 Hz to 20 kHz 95 dB + 4 V 90 dB + 4 V 80 dB + 8

22、V Spectrumpost-FFT of filtered signal Difference frequency distortion (DFD) Measurement methodin accordance with IEC 268-3 or IEC 118 Frequency rangedifference frequency center frequency 80 Hz to 2 kHz 200 Hz to 80 kHz Error limitsfcenter 20 kHz0.5 dB Inherent distortionDFD d2, fcenter 20 kHz DFD d3

23、, 5 kHz fcenter 20 kHz 105 dB 90 dB Spectrumbargraph showing signal and distortion Modulation distortion (MOD DIST) Frequency rangelower frequency upper frequency 30 Hz to 2.7 kHz 8 flower to 20 kHz Error limits0.5 dB Inherent distortionflower = 60 Hz, 4 kHz fupper 15 kHz flower = 60 Hz, 15 kHz fupp

24、er 4 V 85 dB 80 dB 75 dB Spectrumbargraph showing signal and distortion Frequency Frequency range20 Hz to 80 kHz Error limitsmeasurement time 10 s measurement time 1 s 10 ppm 100 ppm Phase Frequency rangeanalyzer bandwidth 22 kHz analyzer bandwidth 80 kHz 20 Hz to 22 kHz 80 Hz to 80 kHz Error limits

25、f 20 kHz, both channels with same range1 Polarity test Measurementpolarity of unsymmetrical input signal Displaypositive/negative Audio Analyzers UP300/UP35011 FFT analyzer Frequency rangeDC to 80 kHz FFT size1 k, 2 k, 4 k, 8 k, 16 k points Window functionsrectangular, Hann, Blackman-Harris, Rife- V

26、incent 1 to 3, Hamming, flat top, Kaiser (b = 12) Resolution16 k points, bandwidth 22 kHz2.93 Hz Averagingexponential or normal1 to 256 BNC connectors2 channels, electronic, floating (max. 0.2 V peak referenced to ground) or grounded, short-circuit- proof, max. current 120 mA with external feed chan

27、nel 1 on front panel, channel 2 on rear panel Voltage rangesine, open-circuit0.1 mV to 7.5 V (Vrms) Source impedance27 Crosstalk attenuationf 100 dB Load impedance200 Common-mode rejectionat 1 kHz50 dB BNC connectorunbalanced, transformer coupling, on rear panel Impedance75 , short-circuit-proof Out

28、put level (Vpp)into 75 0.5 V Optical outputTOSLINK Channels1, 2, or both Audio bits16 to 24 Sampling rate32 kHz, 44.1 kHz, 48 kHz, 96 kHz, 192 kHz Formatprofessional and consumer Weighting filtersA weighting C message CCITT CCIR unweighted CCIR 1k weighted CCIR 2k weighted deemphasis 50/15, 50, 75,

29、J. 17 IEC/IEEE tuner Third-octave and octave filters Filters For all analog and digital analyzers. Up to three filters can be combined. All filters are digital filters with a coefficient accuracy of 32 bit floating point. Analog audio outputs Digital audio outputs (model UP 350 only) 12Audio Analyze

30、rs UP 300/UP 350 Sine Frequency range2 Hz to 80 kHz Frequency error10 ppm Level errorat 1 kHz0.1 dB Frequency response (ref. to 1 kHz)20 Hz to 20 kHz0.05 dB Inherent distortion THD+Nmeasurement bandwidth 20 Hz to 22 kHz90 dB Sweep parametersfrequency, level MOD DISTfor measuring modulation distortio

31、n Frequency rangelower frequency upper frequency 30 Hz to 2700 Hz 8 flower to 39.95 kHz Level ratio (LF:UF)selectablefrom 10:1 to 1:1 Error limits0.5 dB Inherent distortionat 60 Hz, 7 kHz, level ratio 4:1 other settings; fupper 20 kHz 90 dB 84 dB DFDfor measuring difference frequency distortion Freq

32、uency rangedifference frequency center frequency 80 Hz to 2 kHz 200 Hz to 39.95 kHz Error limits0.5 dB Inherent distortionDFD d2, 7 kHz fcenter 20 kHz DFD d3, 7 kHz fcenter 20 kHz 105 dB 90 dB Multisine Frequency range2.4 Hz to 80 kHz Minimum frequency spacingbandwidth 22 kHz2.4 Hz Dynamic rangerefe

33、renced to peak value100 dB Characteristics1 to 17 spectral lines, level, start phase and frequency selectable for each line Sine burst Burst time1 signal period up to 60 s Interval timeburst time up to 60 s Low levelzero to burst level, absolute or relative to burst Noise DistributionGaussian, trian

34、gular, rectangular Polarity test signal SINE2 BURST signal1.2 kHz ON-TIME1 cycle INTERVAL2 cycles Signals 13Audio Analyzers UP 300/UP 350 Measurement functionsSweep RMS Sweep THD(N) wideband or selective Generator signalsine Sweep modefrequency and/or level Sweep spacinglinear, logarithmic Sweep ste

35、ppingsingle, continuous Sweep pointsX-Axis Z-Axis (Freq. automatically detected no error, sequence errors, preamble errors Clock rate measurementerror limits50 ppm Interfaces USB hostprinter; USB stickA plug, protocol version 1.1 USB devicedevice-specific command set, remote control via Windows driv

36、er (Windows XP/2000) B plug, protocol version 1.1 Connector for external monitor (VGA)15-pin D-Sub female Keyboard connectorPS/2 female Display Type5.4“ active TFT color display Resolution320 240 pixels Max. refresh rate10 pictures/s, nominal Power supply Input voltage rangeautoranging100 V to 240 V

37、 (AC), 50 Hz to 60 Hz Power consumption120 VA Ambient conditions Operating temperature rangemeets EN 60068-2-1/2+5 C to +45 C Storage temperature range20 C to +70 C Relative humiditymeets EN 60068-2-3 (non-condensing)95 % at +40 C Mechanical resistance Sinusoidal vibrationmeets EN 60068-2-6, EN 6101

38、0-1 and MIL-T-28800D class 5 5 Hz to 150 Hz, max. 2 g at 55 Hz, 55 Hz to 150 Hz: 0.5 g constant Random vibrationmeets EN 60068-2-6410 Hz to 500 Hz: 1.9 g Shockmeets EN 60068-2-27 and MIL-STD-810shock spectrum Electromagnetic compatibilitymeets EN 55011 class B and EN 61326 (EMC Directive of EU (89/3

39、36/EEC) EMI field strength10 V/m SafetyEN 61010-1/IEC 61010-1, UL 3111-1; CSA C22.2 No. 1010.1 Dimensions (W H D)219 mm 147 mm 350 mm (8.62 in 5.79 in 13.78 in) Weight9 kg (19.84 lb) Digital audio protocol (model UP 350 only) General specifications Audio Analyzers UP300/UP35015 DesignationTypeOrder

40、No. Audio Analyzer (analog interfaces)UP 3001147.2494.03 Audio Analyzer (analog and digital interfaces)UP 3501147.2507.03 Rack AdapterZZA-3001147.1281.00 Carrying CaseZZK-3001147.2542.02 Accessories supplied with the UP 300/350 Operating manual (German/English), USB interface cable for connection to

41、 PC, power cable Ordering information Certified Quality System ISO 9001 DQS REG. NO 1954 QM Certified Environmental System ISO 14001 DQS REG. NO 1954 UM www.rohde- Europe: +49 1805 12 4242, customersupportrohde- USA and Canada: 1-888-837-8772, customer.supportrsa.rohde- Asia: +65 65 130 488, customersupport.asiarohde- is a registered trademark of Rohde & Schwarz GmbH & Co. KG Trade names are trademarks of the owners Printed in Germany (ed) PD 5213.5674.32 UP 300/UP 350 Version 02.00 January 2008 Data without tolerance limits is not binding Subject to change