HP-130A-Manual 电路图.pdf

上传人:cc518 文档编号:205264 上传时间:2025-03-08 格式:PDF 页数:66 大小:2.52MB
下载 相关 举报
HP-130A-Manual 电路图.pdf_第1页
第1页 / 共66页
HP-130A-Manual 电路图.pdf_第2页
第2页 / 共66页
HP-130A-Manual 电路图.pdf_第3页
第3页 / 共66页
亲,该文档总共66页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

《HP-130A-Manual 电路图.pdf》由会员分享,可在线阅读,更多相关《HP-130A-Manual 电路图.pdf(66页珍藏版)》请在收音机爱好者资料库上搜索。

1、HP Archive This vintage Hewlett Packard document was reproduced with permission from Agilent Technologies preserved and distributed by Please visit us on the web ! On-line curator: Tony Gerbic RadioFans.CN 收音机爱 好者资料库 OPER.TINGNDSERVICING MNU.L ).,- OSCILLOSCOPE Serial125 HEWLETTPACKARDCOMPANY 275Pav

2、eMillR_dPolo.Ita,California,U.S. RadioFans.CN 收音机爱 好者资料库 PRELIMINARYOPERATINGINSTRUCTIONS FOR I OSCILLOSCOPE Serial125 Copyright1956byHewlett-PackardCompany Theinformationcontainedinthisbooklet isintendedfortheoperationandmain- tenanceofHewlett-Packardequipmentand isnottobeusedotherwiseorreproduced

3、withoutthewrittenconsentoftheHewlett- PackardCompany. HEWLETT-PACKARDCOMPANY 275PAGEMILLROAD,PALO1LTO,CALIFORNIA,U.S.A. -_._- l30AOOa RadioFans.CN 收音机爱 好者资料库 .:.(lBSrTliJHAl1ISUTtHHI WHENIVEAT.SENSITIVITYI ISTUItNEDfROM11111.TOMAX, ADJUSTVERTICALPOSITIOIII-_ SETTO(OIRCT-COlIPlfDIIIPJDfOR DCSICU.LS.S

4、ETTO(ITYCOUPl.ll 11l1lJT)FORAtSICIIAlS, FOCUS INTENSITY /CAL. VERNIER HeR,Z SENSITIVITY ,.,).-, ADjUSTSlEEPTIIlE/CIlBETlEEN STEPSOFISWEEPT1MEICMISElECTOR VERNIER SWEEPTIME/eM ;_.=-,-_.JSWEEPMODE TIIIGGIt-FIlEII:ItUNINPUT VERNIER VRT.SENSITIVITY INPUT SELECTTlIlE/CIlOFIIIHRUlSlEEP (:!:5)IITHIVERNltAI

5、SETrol2CAlI REMOYESHORTlIICSAltFOR-t-it(5)1 BALANCEOINPUT.5;L?r;i;.2: ADJUSTTOSTUTSIHIATDESIRED-J.00L-_,1-_ YOlTACEOFSYIICSICIlAlOR,IFlIfSSARr.-:7L=:t!:=:1,:=_ AOJBIlllESlEEP11TH_SWEEPMODEl III ADJUSTVERTICALSfIISIT1YIHBElIEE.-1Itt STEPSOFtIERilCAI:$ENS!TMTjJSElECTOR SELECTDESIREDYRTlCAlPUT;-l SENSI

6、TIVITY(t5)11THLYERHIERISElIDIleALI SUTOICALITOCllECKSEIlSITlYITY VERtiCALINPUT.-, BALANCEDIIiPurCAliBEUSED011 1-20NY/eNRAIIGES. snTOTRICCERSlEEPFROMElfHERPOslTlvE-1-;,-t ORUCUlVSLOPEOfSlitSICNAl. I , ADJUSTTODSIAIISTArlOIlARI POSITION1KENlHORlz.SENSITIVITYI ISTURNEDfROMIIINTOMAX ADJUSTHORIZONTAlPOSI

7、TION. SHTOIINT.SWEEPITOOPERATEINTERNALSWEEP. SElECTOESIRtDHORIZONTALINPUTSENSITIVm (S4lWITHIItORLZ.!lNSITIYITY1SETTOl.cALI SETTOICALITOtHEe,SfNSIlIVITY SETTOf(OIRfCHnJPlDINPUllfill DCSleNAlS;SETTOlCAPM:ITY-COOPLfD I.PUT)fOllAtSICULS. ADJUSTHORtlOHAlSEItSITIVITTmlnN STEPSOFIHoRlz,sENSITlVlTYISELECTOR

8、 (CJf+-iI-SWEEPOR5YNCHRONIZIMCINPUTS. TOSYNCSCOPE.SETlKOAIZ.SENStTlVITVITOliNT.SWEEP! BAlANCEDINPUTCANBEUSEDONl-Z0MY/GilRANGES. 11;tt-+-REIlOVfSHORTINGBARFOR BALANCEDINPUT INTENSITY VERNIER HOA1ZSENSITIVITY ., - , :, r 10 0 ADJUSTSHPTlMlCMBET,M STPSorISWEEPTlMEltMISlCTOR VERNIER SWEEPTIME/eM IFIMTRl

9、lAlSP sn11112CAl.1 IANIER SENSITIVITY 1:- S 2.I ?.OO.,;:t:;:=_I-SELECTCHARACTEROfSYNCSIGNALOSIRED. _1:!:!:!1=L=:STTOIPRE-SET!FOR8STSYNCSTABILITY ORTOlFREE-RUN!ASOSIRD. I 130ASPECIFICATIONS _SWEEPGENERATOR INTERNALSWEEP: SYNCHRONIZATION: SYNCCONTROL: 21calibratedrangesprovideSweepspeedsfrom1I-l sec/c

10、mto5sec/cm,accuratetowithin50/. Averniercontrolprovidescontinuousadjustmentof sweepspeedbetweencalibratedrangesandextends slowestsweeptoapproximately15sec/ern. Internal,fromlinepowerorverticalinputsignalhav- ing1/2Crnormaredeflection. External,eithercapacitivelyordirectlycoupledfrom signalsof1/2volt

11、peak-to-peakormore. Sweepctriggtredfromeitherapositive_ora nllllflJv.e-_t:lgrfQtriggeringvoltagelevelof externalsyncsignas1Sontinuouslyadjustablefrom -30to+30volts. Switchposition,Preset,automaticallyprovidesop- timumsyncstabilityformajorityofuses. -VERTICALANDHORIZONTALAMPLIFIERS - BANDWIDTH: SENSI

12、TIVITY: COMMONMODEREJECTION: (Balancedinputonly) INPUTIMPEDANCE SingleEnded: Balancedinput(onI,2.10, and2.0mv/cmrangesonly): OUTPUT: TYPE: DEFLECTIONPLATE CONNECTIONS: INTENSITYMODULATION: REPLACEABILITY: POWERREQUIREMENTS: SIZE: WEIGHT: Directcurrentto300kilocycles. 14calibratedrangesprovidesensiti

13、vitiesfrom1milli- vOlt/cmto20volts/ern,accuracy50/. Averniercon- trolprovidescontinuousadjustmentbetweencalibrated rangesandextendstheminimumsensitivitytoapproxi- mately50volts/ern. Rejectionatleast40db.Commonmodesignalmust notexceed1.5volts. 1megohmshuntedbyapproximately50I-ll-lf. Zmegohmsshuntedby

14、approximately25IJ4J.f. CALIBRATOR FixedamplitudelOOO-cyclesquarewaveautomatically fedtoverticalandhorizontalamplifierinputswhen SwitchedtoCAL. CATHODERAYTUBE 5AQPmono-accelerator,flatface.availablewithPl. P7orPllscreen.3000-voltacceleratingpotential. Screwterminalsinrearaccessreceivewiresforcon- nec

15、tiontoplates. Terminalsinsiderearaccesstoreceive+20pulsefor blankingCRTtraceofnormalintensity. CRTbezelremoveswith150twistforreplacementof graticu1eorCRTfromthefrontpanel. CRTbezelprovidesfirmmountingforstandardoscil- loscopecameras. 115/230VAClO%.50/400cycles;approximately175 watts. 9-3/4widex15-1/

16、4highx20deep. A. .-._- INTERNALSWEEP-INTERNALSYNCHRONlZAnON 9J-. 8 2 1 JUMPER . I , / (, 110 n+_+_-7 H-+-4 1.PlaceverticalsignalintoINPUT.Youcan useabalancedinputonthe1to20mv/cm rangesoftheVERTSENSITIVITYcontrol (2)byremovingthegroundjumperonthe INPUTterminal. 2.AdjustVERTSENSITIVITYfordesired VOLTS

17、/CMdeflection. 3.InmostcasessetforACcoupling.See paragraph2-4. 4.PlaceexternalsynchronizingsignalaCross horizontalINPUT. 5.SwitchSYNCselectortoEXTACorDCas desired.Seeparagraph2-4. 6.SetSWEEPMODEtoPRESET. 7.SwitchHORIZSENSITIVITYtoSWEEP. 8.AdjustTRIGGERLEVELtoZERO. 9.AdjustTRIGGERSLOPEtotriggersweep

18、onpositiveornegativeslopeofinputwave asdesired. 10.SelectdesiredsweepspeedwithSWEEP TIMEelMswitch. 2-2 EXTERNALHORIZONTAL INPUT 1.PlaceexternalhorizontalsignaltoINPUT. Balancedinputmaybeusedon1-20mv/cm rangesofHORIZ.SENSITIVITYcontrolby removinggroundjumperonINPUT. 2.Selectdesiredhorizontalsensitivi

19、ty.This controlisahorizontalinputattenuatorwhen notintheINTSWEEPposition,andalliI. ternal-sweepcontrolsarede.energized. 3.Selectdesiredinputcoupling.Paragraph 2-4. 4.Adjusthorizontalpositionofsignal. 2-3 REMOVING CRTBEZEL LOCK RELEASE BUTTON BEZEL TheCRTbe.,;elcontainsthefilterandgrati- culeshields;

20、ithasalsobeendesignedtoac- ceptoscilloscopecameraequipmentwithout adaptersofanykind.Tochangefiltersor cathoderaytubesthebe.,;elmustberemoved. 2-4 Toremovebezel: 1.Depresslockreleasebutton,andtwistbezel counterclockwiseabout15degrees. 2.Pullbezelstraightfrompanelasshown. .-.,.- ALIGNINGSCOPETRACEWITH

21、 GRATICULE 2-5 VERTICAL AMPLIFIER I HORIZONTAL AMPLIFIER +88VOLT REGULATED POWERSUPPLY +600AND-150VOLT REGULATEDPOWER SUPPLIES,TRANSFORMER ANDALLRECTIFIERS (BELOW) DEFLECTIONPLATE TERMINALBOARD +300AND +135VOLT REGULATED POWERSUPPLIES REGULATEDR-F HIGHVOLTAGE POWERSUPPLY (INSIDE) SWEEPGENERATOR, SYN

22、CCIRCUITAND CALIBRATOR (BELOW) 4-0 SECTIONIV MAINTENANCE 4-1INTRODUCTORY Thissectio1containsinstructionsforadjustingandservicingthe BOAOSCilloscope.TheBOAisconstructedsothateachofthe majorcircuitsectionsisphysicallylocatedonasingleetched circuitboard_exceptforportionsofthepowersupplieswhich utilizet

23、hreeseparatechassis.Thematerialinthissectionis dividedasthecircuitsections,eachsectionhasacompletesetof adjustmentinstructions_andattherearofthemanual_asche- maticandavoltage-resistancediagram.Thematerialinthis sectionisasfollows: 4-2RemovingtheCabinet 4-3Connectingfor230-VoltPowerLines 4-4Servicing

24、EtchedCircuits 4-5TubeReplacementChart 4-6IsolatingTroublestoMajorSections 4-7AdjustingtheLow-VoltagePowerSupplies 4-8AdjustingtheRFHigh-VoltagePowerSupply 4-9ReplacingandAdjustingtheCRT 4-10CheckingtheCalibrator 4.11AdjustingtheVerticalAmplifier 4.12AdjustingtheHorizontalAmplifier 4.13AdjustingtheP

25、resetSensitivityoftheSweep Generator 4-14AdjustingtheSweepGenerator 4.2REMOVINGTHECABINET The130Achassisandpanelareremovedfromthecabinetbyre- movingthetworetainerscrewsontherearofthecabinetand slidingthechassisforwardoutofthecabinet. 4-1 SERVICINGPRINTEDCIRCUITBOARDS CAUTION-, Whenservicingprintedci

26、rcuitsDONOTpushorpull wiresinsuchawayastoraisetheprintedwiringfroTIl theboard. Whensolderingleads,use50wattironorSTIlaller. Applyheatsparinglytotheleadsontheparttobere- placed,nottotheprintedwiringontheboard. Beforeinstallingnewparts,cleanholestoreceive newpartwithoutforcing.Havenewleadstinnedand if

27、necessaryfluxedtoreceivesolderquicklywitha TIliniTIluTIlofheatandwithoutresidue. CDAPPLY HEATSPARINGLYTOLEADOFPARTTOBEREPLACED. REMOVEPARTFROMCARDASIRONHEATSTHELEAD. CD USINGA SMALLAWL,CAREFULLYCLEAN INSIDEOFHOLELEFTBYOLDPART. I.,tta!(la!IIII141 CDBENDCLEANLEADSONNEWPART ANDCAREFULLYINSERTTHROUGH HO

28、LESONBOARD. CDHOLDPARTAGAINSTBOARD ANDSOLDERLEADS. DiagraTIlshowinghowtoreplacepartsTIlounted onprintedcircuitboards 4-3CONNECTINGFOR230-VOLTPOWERLINES TheBOAisnormallyshippedfromthefactorywiththedualpri- marywindingsofthepowertransformerconnectedinparallelfor useonll5-volta-clines.Thewindingscaneas

29、ilybereconnected inseriesforuseon230-voltpowerifdesired.Toreconnectthe primarywindingsofT2foruseon230voltsdisconnectthejumpers whichjointerminalsAltoA3andA2toA4.Connectajumper betweenterminalsA 2andA3.Replacethe2-ampslow-blowfuse witha1-1/4-ampslow-blowfuseandtheBOAcannowbeoperated from230-voltlines

30、withnochangeinoperation. 4-4SERVICINGETCHEDCIRCUITS Figure4-2illustrateshowtoreplaceelectricalpartsonetched circuits. Whenservicingetchedcircuits,DONOTpushorpullwiresin suchawayastoraisethewiringfromtheboard. Whensolderingleadsontheetchedboard,usea50wattironor smaller.Applyheatsparinglytotheleadsont

31、heparttobere- placed,nottothewiringontheboard. Beforeinstallingnewparts,cleanholetoreceivenewpartwith- outforcing.Havenewleadstinnedandifnecessaryfluxedtore- ceivesolderquicklywithaminimumofheatandwithoutresidue. 4-5TUBEREPLACEMENTCHART ThefollowingchartlistsallthetubesintheBOAwiththeirfunc- tionsan

32、dadjustmentsrequiredwhenreplacingtubes.Theheaters ofsomeofthetubesinthel30Aareoperatedinseriesfroma regulatedd-cvoltageobtainedfromtheLow-VoltagePowerSupply. ThisseriesheaterstringisshowninonepieceontheLow-Volt- agePowerSupplyschematicdiagram;theactualconnections througheachcircuitareshownindividual

33、lyontheschematicof thecircuit.Whenreplacingtubesduringtroubleshooting,ifone oftheseriestubesispulled,itwillturnoffallothertubesinthe string. 4-2 DESIGNATION VI V2 V3 V4 V5 V6 V7 V8 V9 VIO VIi VI2 V13 Vl4 VI5 Vl6 Vl7 VI8 Vl9 V20 V21 V22 V23 V24 V25-V26 V27 V28-V29 V30 V31 V3Z V33 V34 SR2 TUBE TYPE 12

34、AU7 6211 12AT7 6BQ7 12AU7 6211 12AT7 6BQ7 6BQ7A 12AT7 6BQ7A 6BC7 6AW8 12AX7 12AU7 6BQ7 6AQ5 lX2B lXlB 6BQ7A 5AQP 12B4 6BH6 6BW4 12B4 6AW8 12B4 6BH6 6X4 IZB4 6BH6 5651 212-122 TUBEREPLACEMENTCHART FUNCTION VERT.AMPL. PhaseInverterAmplifier DifferentialAmplifier DifferentiaiAmplifier DifferentialAmpli

35、fier HORIZ.AMPL. PhaseInverterAmplifier DifferentialAmplifier DilferentialAmplifier D,fferentialAmplifier SWEEPGENERATOR TriggerAmplifier TriggerGenerator SweepStart-StopTrigger a.DiodeClamp b.IntegratorSwitch c.IntegratorSwitch a.IntegratorCathodeFollower b.FeedbackIntegrator a.RetriggeringHoldOff

36、b.RetriggeringBiasControl a.GateOutCathodeFollower b.Clamp CalibratorMultivibrator HIGHVOLTAGEPOWERSUPPLY Oscillator HighVoltageRectifier HighVoltageRectifier ControlTube CRT bOO-voltSeriesRegulator 600-voltControlTube 300-voltRectifier 300-voltSeriesRegulators a.300-voltControlTube b.I35-voltCathod

37、eFollower +88-voltSeriesRegulators +8a-voltControltube -150-voltRectifier -150.voltSeriesRegulator -150-voltControlTube ReferenceTube +88-voltRectifier 4-3 CHECKOR ADJUSTMENTSNECESSARY AdjustBalGainandC23(para.4.11) AdjustBalGainandC23(para.4-11) AdjustGain,C23andC21(para.4-11) AdjustGain,C23andC21(

38、para.4-11) AdjustBal,GainandC59(para.4-12) AdjustBal,GainandC59(para.4-12) AdjustGain,R96,C49, C53,C55, C57,C59(para.4-12) AdjustGain.R96,C49.C53,C55, C57,C59(para.4-12) Noadjustmentrequired Noadjustmentrequired AdjustPreset,SweepLength (paras.4.13and4-14) Noadjustmentrequired CheckSweepCalib.(para.

39、4-14) AdjustPreset(para.4-13) Noadjustmentrequired Noadjustmentrequired-Check Symmetry Noadjustmentrequired Noadjustmentrequired Noadjustmentrequired ReadjustHighVoltage(para.4-8) AdjustgainofVERT.andHORIZ. amplifiers(para.4-11and4-12) Noadjustmentrequired Noadjustnoentrequired Noadjustmentrequired

40、Noadjustmentrequired NoadjustmentreqUired AdjustL.V.Supplyifnecessary (para.4.7) Adjust.L.V.Supplyifnecessary (para.4-7) AdjustL.V.Supplyifnecessary (para.4-7) AdjustL.V.Supplyifnecessary (para.4-7) AdjustL.V.SupplyifneCeSsary (para.4-7) AdjustL.V.Supplyifnecessary (para.4-7) AdjustL.V.Supplyifneces

41、sary (para.4.7) 4-6ISOLATINGTROUBLETOMAJORSECTIONS Inanycaseoftroubleshooting.attemptoperationofthevarious sectionsoftheoscilloscopeanddeterminewhichmajorsection containsthecircuitfailure.Proceedinthefollowingmanner. a.ChecktheLow-VoltagePowerSupplyvoltagesasdescribed inparagraph4-7. b.ChecktheHigh-

42、VoltagePowerSupplyvoltagesasdescrib- edinparagraph4-8.TurnbothVERT.SENSITIVITYand theHORIZ.SENSITIVITYinputselectorstooff.andcenter thepositioncontrolstoobtainaspot.Notetheactionof theintensityandfocuscontrols.Ifthespotdoesnotappear, measurethedeflectionplatevoltages. c.ChecktheVerticalAmplifierbysw

43、itchingtoCAL,andnote theresultantdeflectionwhichshouldbeexactly6centi- meters. d.ChecktheHorizontalAmplifierinthesamemannerasthe VerticalAmplifier. e.ChecktheSweepGenerator.WiththeSWEEPMODEcon- trol.attemptfree-running;thensettostopthefree-run- ning. Whenadefectivesectionisfound.seethatallthetubesar

44、e lighted;checkbyreplacementorwithatubechecker;thenmeas- urevoltages.RefertotheVoltageResistanceDiagramsandthe TubeReplacementChartforassistancewithchangingtubesand measuringvoltagesandresistancesineachcircuit. Thetwosidesofthedirect-coupled,push-pullcircuitsarenorm- allybalancedandcausethespottobes

45、tationaryinthecenterof thescopescreen.Afaultineithersideusuallyunbalancesthe circuitandcausesthespottomoveoffthescreen.Tobringthe spotback.shorttogetherthecontrolgrids(ortheplates)ofthe twosidesofonestagestartingatthelaststage.Thiseliminates signalsofalltypes,d-cunbalance.jitter,etc.whichoriginate p

46、riortotheshortedpoints.Ifshortingthetw0halvesofastage togetherdoesnotbringthespotonthescreenandholditmotion- less,asubsequentcircuitisfaulty.Bycontinuingthisprocess throughtheamplifiertroublecanbeisolatedtoasmallcircuit area. 4-7ADJUSTINGTHELOW-VOLTAGEPOWERSUPPLIES Thecomplete.low-voltagepowersupply

47、providesfive.separate- ly-regulatedoutputvoltages:-150vdc.+88vdc,+135vdc.+300 vdcand+600vdc.Thelow-voltagesupplieshaveonlyone 4-4 R268 LOWVOLTAGEADJ.-_-l.-L R210 ASTIGMATISM-1 6,3VACFORCRT-+-+-I C137 (COMMONTERM.) GROUND C37A MEASURE+300VOLTS C137C MEASURE+88VOLTS C137B MEASURE+135VOLTS C140 (COMMON

48、TERM.) MEASURE -150VOLTS CI38 MEASUREOUTPUT FROMSR2 CI34 MEASUREOUTPUT- FROMV24 C131 MEASUREOUTPUTJ FROMSRI C133 MEASURE+600VOLTS-iI:-.LJ,.fI ONPOS, TERM. +300VOLTSONNEG.TERM. C140A MEASUREOUTPUT FROMV31 C140B GROUND FIGURE403POWERSUPPLYMEASUREMENTPOINTS adjustmentwhichsetsthe+88-voltfilamentsupply.

49、Thereare noseparateadjustmentsforthe-150-volt,+135-volt,+300-volt and+600-voltregulatedsupplies.Alltheremainingvoltagesde- penduponcorrectadjustmentofthe+88-voltsupplyandalso uponhavingthecorrectoutputvoltagefromthe-ISO-voltsupply. Allregulatedvoltagescanbeconvenientlymeasuredontherear, left-handcornerofthe130A,seeFig.4-3.Toadjustthe+88- v