1、HPARCHIVE This vintage Hewlett Packard document was preserved and distributed by www. Please visit us on the web! Scanned by on-line curator: Jack Hudler * For FREE Distribution Only * RadioFans.CN 收音机爱 好者资料库 OPERATING AND SERVICE MANUAL TIME INTERVAL CALIBRATOR P a HEWLETT PACKARD RadioFans.CN 收音机爱
2、 好者资料库 506-59992A TIME INTERVAL CALIBRATOR SERIAL PREFIX: 2526A This manual applies to Serial Prefix 2 5 2 6 4 unless accompanied by a Manual Change Sheet indicating otherwise. HEWLElT PACKARD OPERATING AND SERVICE MANUAL Q Copyright HEWLETT-PACKARD COMPANY 1985 5301 STEVENS CREEK BOULEVARD. SANTA C
3、LARA. CALIFORNIA 95051 MANUAL PART NUMBER 59992-91094 Printed OCTOBER 1985 www. RadioFans.CN 收音机爱 好者资料库 www. . CERTIFICATION HewIm-hckard Company certifies that this product met its published sFificatiohs at the time of shipment from the faCtOfy. Hewlett-Padcard further certifies that its calibratio
4、n measurements are traceable to the United States National Bureau of Standah, to the extent allowed by the Bureaus calibration facility, and to the calibration facilities of other International Standards Organization members. WARRANTY This Hewlett-Padcard instrument produa i s warranted against defe
5、cts in material and workmanship for a period of one year from date of shipment. During the warranty period, Hewlett-Packard Company will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility
6、designated by HP. Buyer shall prepay shipping charges to HP and H P shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to HP from another country. HP warrants that its software and firmware designated by H
7、P for use with an instrument will execute its programming instructions when properly installed on that instrument. HP does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error free. LIMITATION OF WARRANTY The foregoing warranty shall not apply to
8、defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse,operation outsideof the environmental specifications for the product, or improper site preparation or maintenance. NO OTHER WARRANTY IS E X P R E S S E D OR
9、 IMPLIED. HP SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. EXCLUSIVE REMEDIES T H E REMEDIES PROVIDED HEREIN A R E BUYERS SOLE AND EXCLUSIVE REMEDIES. HP SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT,SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGE
10、S, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY. ASSISTANCE Product maintenance agreements and other customer assistance agreements are awilable for Hewlett- Packard products. for any assistance, Contad your nearest Hewlett-hckard Sales and Service Office. Addresses are provided at the
11、 back of this manual. Table of Contents Section I GENERAL lNFORMATlON Page l.l.IN?RODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1. 2.
12、GENERAL DESCRIPTION 1 . 1 . A 1 . 3 . SPECIFICATIONS 1 2 . . 1 . 4 . EQUIPMENT SUPPLIED 1-2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 5 . RECOMMENDED TEST EQUIPMENT 1 . 3 Section 2 INSTALLATION Page 2- 1 . IN?RODUCTZON . . . . . . . . . . . . . . . . . . . . . .
13、 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 2-2 PREPARATIONFORUSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2- 3 Power Requ
14、irements 2-1 2-4.Power-On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2- 5 . LOCAL OPERATION/REMOTE OPERATION 2-2 . . . . . . . . . . . . . . . . . . . . . . . . .
15、. . . . . . . . . . 2-6 . Setting the HP J06-59992A HP-IB Address 2-2 . 2- 7 . HP-IB Interconnections 2-3 Section 3 OPERATING INFORMATION Page 3-1.INTRODUcIION . 3-1 3-LADJUSTMENTS . 3-1 3-3 . OPERATION VERIFICAON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
16、. 3-1 3-4 . AUTOMATED CALIBRATION ROUTINE -USING HP 5370AIB COUNTER . . . . . . . . . . . . . 3-3 3.5 . AUTOMATED CALIBRATION ROWTINE .USING HP 5 363B PROBES . . . . . . . . . . . . . . . . 3-6 3.6 . TROUBLESHOOTING HINTS AND GENERAL COMMENTS . . . . . . . . . . . . . . . . . . . . . . . . 3-9 3.7 .
17、 HP-IB PROGRAM LISTINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9 Section 4 THEORY OF OPERATION Page 4.1.INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 4. 2 . TIME INTERVAL CAL
18、IBRATOR . 4-1 4- 3 . A l VHF Switch Control Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 4.4 . A2 VHF Switch Board . 4-1 .4. 5. Theory of Operation . 4-1 4.6 . HP-IB COMMAND OPERATION . 4-1 4.7 . INPUT SIGNAL CONVERSION . . . . . . . . . . . . . . .
19、 . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2 4.8 . CALIBRATION WITH AN OFFSET VOLTAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 4.9 . EXAMPLE OF A NON-ZERO TRIGGER LEVEL MEASUREMENT . . . . . . . . . . . . . . . . . 4-4 . . . . . . . . . . . . . . . . . . . . . .
20、. . . . . . . . . . . . . . . . . . . 4- 1 0 Functional Block Diagram Theory 4-5 4- 1 1 . HANDSHAKE LOGIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 4.12 . BUSLOGIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21、. . . . . . . . . . . . 4-5 4- 1 3 . DECODE LOGIC . 4-5 www . Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4- 1 4 . REMOTE/LOCAL LOGIC 4-7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4- 1 5 . LOCAL LOCKOUT LOGIC 4-7 4- 16 . REL
22、AY SELECT LOGIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7 Section 5 REPLACEABLE PARTS Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5- 1 . INTRODUCTION 5-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
23、 . . . . . . . . . . 5-2 PARTS ORDERING 5-1 5- 3 . COMPONENT LOCATORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9 www . SAFETY CONSIDERATIONS GENERAL This product and related documentation must be reviewed for familiarization with safety markings and instruction
24、s before operation. This product is a Safety Class I instrument (provided with a protective earth terminal). BEFORE APPLYING POWER Verify that the product is set to match the available line voltage and the correct fuse is installed. Refer to Section Il, Installation. SAFETY EARTH GROUND An uninterru
25、ptible safety earth ground must be provided from the mains power source to the product input wiring terminals or supplied power cable. SAFETY SYMBOLS Instruction manual symbol; the The WARNING sign denotes a product will be marked with this hazard. It calls attention to a symbol when it is necessary
26、 for the procedure, practice, or the like, user to refer to the instruction which, if not correctly performed or adhered to, manual. could result in personal injury. Do not proceed beyond a WARNING sign until the indicated con- Indicates hazardous voltages. ditions are fully understood and met. The
27、CAUTION sign denotes a Indicates terminal is Onnected hazard. It calls anention to an /j7 OR to chassis when such connection is not apparent. operating procedure, practice, or the like, which, if not correctly performed or Alternating current. - , , , Directcurrent. adhered to, could result in damag
28、e to or destruc- tion of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated con- ditions are fully understood and met. SAFETY INFORMATION Any interruption of the protective grounding conductor r inside or outside the instrument1 or discon- necting the protective ear
29、th terminal will cause a potential shock hazard that could result in personal injury. (Grounding one conductor of a two conductor outlet is not sufficient protection. I Whenever it is likely that the protection has been impaired, the instrument must be made inoperative and be secured against any uni
30、ntended operation. If this instrument is to be energized via an autotransformer (for voltage reduction) make sure the common terminal is connected to the earthed pole terminal (neutral) of the power source. Instructions for adjustments while covers are removed and for servicing are for use by servic
31、e-trained personnel only. To avoid dangerous electric shock, do not perform such adjustments or servicing unless qualified to do so. For continued protection against fire, replace the line fusels I only with 250V fuse1 s) of the same current rating and type (for example, normal blow, time delay). Do
32、 not use repaired fuses or short circuited fuseholders. 4-1. INTRODUCTION GENERAL INFORMATION This manual contains operating and sewice information for the 306- 59992A Time Interval Calibrator. Included is the following. * general description of the calibrator *+ specifications * equipment supplied
33、and recommended test equipment H* installation information * operating information * automated procedure for determining calibration constants * HP-IB program listings * theory of operation * replaceable parts lists - component locators -* schematic diagrams rn r;7 1-2. GENERAL DESCRIPTION The 306-5
34、9992A is a special instrument used to increase the accuracy of precision time interval measurements. When employed to determine the sys- tematic errors of a particular measurement set-up, the calibrator can provide for absolute accuracy on the order of +/- 100 ps when making time interval measuremen
35、ts with the Hewlett-Packard 5370A/B Time Interval Counter. The calibrator can also be used to improve the accuracy of the HP 53630 Time Interval Probes when used in conjunction with the HP 5370A/B counter. The instrument itself is a linear, passive device which takes a stable input pulse and produce
36、s two output pulses that are effectively in phase or 180 degrees out of phase. This provides two identical and coherent signals to the input of a measurement instrument such,as a counter. Any time difference measured by the counter in a calibration set-up is a measure of the errors as- sociated with
37、 the measurement channels electrical path lengths and the trig- ger level timing error of the counter. The resuitant computed calibration constant is then subtracted from the measurements that follow, thereby removing the effect of systematic enon from the measurement results. Page 1-1 www. HP J06-5
38、9992A Time Interval Calibrator General Information 1-3. SPECIFICATIONS The instrument specifications for the 106- 59992A Time Interval Calibrator are listed in Table I - 1. Table 1 - 1. J06-59992A Specifications. - Bandwidth: 1 MHz to 200 MHz square wave; 100 kHz to 200 MHz sine wave. Input Voltage:
39、 SV peak-to-peak, max. Effective Edge Repeatability J06- 59992A Time Interval Calibrator AC Power Cord Documentation: 306- 59992A Operating and Service Manual Product Note 53708-2, Better Than 100 ps Accuracy in HP 5370% Time Interval Measurements Through Bias Error Reduction. HP Pub- lication #02-5
40、952-7770. Page 1-2 www. HP 106-59992A Time Interval Calibrator Cenenl Information 1-5. RECOMMENDED TEST EQUIPMENT For Local Operation: Pulse Generator - a ignal source capable of simulating the signal to bc measured. HP 5370A/B Time Interval Counter Oscilloscope - dual channel, general purpose DC Vo
41、ltage Source - (Optional; used to provide offset voltage when not using Time Interval Probes) HP 5363B Time Interval Robes (Optional) The procedures to follow to determine calibration constants using Local Operation are contained in Product Note 5370B-2, Better Than 100 ps Accuracy in HP 5 3708 Time
42、 Interval Measurements Through Bias Error Reduction. For Remote Opention: All equipment listed under Local Operation above, PLUS HP Series 200 Controller such as HP 9 8 36, 9 826,9 8 17) or 9 8 16f HP-IB Interconnection Cables - HP 108 3 3A/B/C/D l These controlIers require a separate disc drive suc
43、h as the HP 9 122 for 3 1/2 microfloppy discs or the HP 9 125 for 5 1/4 floppy discs. Page 1-3 2-1. INTRODUCTION INSTALLATION This section presents the information needed to prepare the HP J06-59992A Time Interval Calibrator for use. Information to prepare for remote opera- tion using HP-IB (Hewlett
44、-Packard Interface Bus, HPs mplementation of IEEE Std. 4 8 8- 19 7 8) is also included. Ma 1 2-2. PREPARATION FOR USE 2-3. Power Requirements The J06- 59992A Time Interval Calibrator operates from either 1 1 S or 2 30 volts, 48 to 66 Hz Before applying power, the voltage selector switch mounted on t
45、he rear panel of the J06-59992A must be set to the correct position (1 15 or 230) using a small screwdriver, and the correct fuse must be installed. This instrument uses a .2A 2 50V time-delay (TD) fuse for 1 1 5 volt operation or a .lA 2 50V time-delay (TD) fuse for 230 volt operation. Refer to Tab
46、le 5-3 in Replaceable Parts for the HP Part Numbers of these fuses. BEFORE PLUGGING THIS INSTRUMENT INTO THE AC POWER LINES, be sure the correct voltage and fuse have been selected. This is a Safety Class I product (i.e., provided with a protective earth terminal). An uninterruptible safety earth gr
47、ound must be provided from the Mains power source to the product input wiring terminals, power cord, or supplied power cord set. When- ever it is likely that the protection has been impaired, the in- strument must be made inoperative and be secured against any unintended operation. If this instrument is to be energized via an external autotransformer for voltage reduction, make sure that the common terminal is connected to the earthed pole of the power source. Page 2-1 HP J06-59992A Time Interval Calibrator Installation 2-4.