《FLUKE 187_189_Calibration Supplement 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 187_189_Calibration Supplement 电路图.pdf(2页珍藏版)》请在收音机爱好者资料库上搜索。
1、Manual Supplement 2003-2007 Fluke Corporation. All rights reserved. Printed in U.S.A. xx Manual Title: 187/189 Calibration Supplement Issue: 2 Part Number: 1584337 Issue Date: 4/07 Print Date: November 2000 Page Count: 1 Revision/Date: 1, 6/02 This supplement contains information necessary to ensure
2、 the accuracy of the above manual. Enter the corrections in the manual if either one of the following conditions exist: 1. The revision letter stamped on the indicated PCA is equal to or higher than that given with each change. 2. No revision letter is indicated at the beginning of the change. Radio
3、Fans.CN 收音机爱 好者资料库 187/189 Calibration Manual Supplement 4/07 1 Change #1 On page 17, Table 3, replace the entry for 350.0C Input with: Input Lower Limits Upper Limits 350.0 C 345.5 C 354.5 C Change #2 On page 17, under Testing Temperature, replace steps 2, 4, and 5 with the following and add step a
4、nd 8. 2. Set the 5520A for K-type thermocouple and an output of -10.0 C. 4. Adjust the temperature offset, using the up and down arrow keys, until the offset on the Meter secondary display reads 0.0. 5. Enter the offset value by pressing the yellow button, then the backlight button. 8. If the Meter fails any of the Temperature Accuracy tests, perform the Temperature Calibration Adjustment procedure. RadioFans.CN 收音机爱 好者资料库